Acta Optica Sinica, Volume. 40, Issue 20, 2018001(2020)

Wide-Field Error Correction Method for Parallel Differential Confocal Axial Measurement

Yiqing Ye1, Dingrong Yi1、*, Wei Jiang1, Linghua Kong2、**, and Caihong Huang1
Author Affiliations
  • 1College of Mechanical Engineering and Automation, Huaqiao University, Xiamen, Fujian 361021, China
  • 2School of Mechanical and Automotive Engineering, Fujian University of Technology, Fuzhou, Fujian 350118, China
  • show less
    Figures & Tables(20)
    Schematic of parallel differential confocal axial measurement method
    Confocal axial light intensity differential response curves
    Schematic of calibration method of axial measurement curve
    Experimental platform for parallel differential confocal axial measurement
    Measured axial measurement curve
    Coin sample. (a) Actual picture and partially enlarged picture of coin sample; (b) confocal picture of coin sample in upper defocusing plane; (c) confocal picture of coin sample in lower defocusing plane
    Topography of coin stripes before correction. (a) Three-dimensional view; (b) top view
    Light intensity differential response simulation curve and axial measurement simulation curve with different peak values
    Light spot array images. (a) Light spot array image in focal plane; (b) enlarged view of solid line frame; (c) enlarged view of dotted line frame
    Axial light intensity response curves of each point in one-dimensional line
    Distribution diagram of light intensity peak value in global field of view
    Light intensity differential response curves of each point in one-dimensional line
    Topography of coin stripes after correction. (a) Three-dimensional view; (b) top view
    Three-dimensional topography of coin stripes measured using commercial microscope. (a) Three-dimensional topography of coin stripes; (b) sectional height profile curve of coin stripes
    Three-dimensional image inpainting figure of coin stripes. (a) Before correction; (b) after correction
    Sectional height profile curves of coin stripes before and after correction
    Step standard sample and enlarged picture of measurement position
    Sectional height profile curves of step standard sample before and after correction
    • Table 1. Height measurement results of standard sample step before and after correction

      View table

      Table 1. Height measurement results of standard sample step before and after correction

      StepnumberStep height aftercorrection /μmStep height beforecorrection /μmStepnumberStep height aftercorrection /μmStep height beforecorrection /μm
      14.7171.702114.8403.282
      24.7112.016124.6713.131
      34.6162.332134.7202.918
      44.7542.730144.7282.570
      54.8242.784154.7402.444
      64.5912.948164.9992.204
      74.6123.052174.9321.711
      84.8623.188184.6631.314
      94.8123.264194.6370.962
      104.8183.276204.6010.892
    • Table 2. Height error analysis of standard sample step before and after correction

      View table

      Table 2. Height error analysis of standard sample step before and after correction

      ConditionRoot mean square /μmMean absolute error /μmRelative error /%
      After correction4.7440.0891.89
      Before correction2.5502.30348.60
    Tools

    Get Citation

    Copy Citation Text

    Yiqing Ye, Dingrong Yi, Wei Jiang, Linghua Kong, Caihong Huang. Wide-Field Error Correction Method for Parallel Differential Confocal Axial Measurement[J]. Acta Optica Sinica, 2020, 40(20): 2018001

    Download Citation

    EndNote(RIS)BibTexPlain Text
    Save article for my favorites
    Paper Information

    Category: Microscopy

    Received: May. 25, 2020

    Accepted: Jul. 3, 2020

    Published Online: Sep. 19, 2020

    The Author Email: Yi Dingrong (yidr@hqu.edu.cn), Kong Linghua (klh@fjut.edu.cn)

    DOI:10.3788/AOS202040.2018001

    Topics