    Spectrum of the line-narrowed ArF laser (FWHM 0.376 pm).
    Experimental setup for wavelength calibration of the ArF laser. HCL, hollow cathode lamp; HV, high voltage; PC personal computer; HR, high-reflection mirror.
    Wavelength with the position of the laser spot on the LDA.
    Waveform of iron OG signal recorded by an oscilloscope.
    OG signal amplitude with discharge current.
    OG signal amplitude at a different laser power and discharge current.
    OG signal amplitude recorded in the wavelength range of 193.38500–193.39500 nm.
    • Table 1. Key Parameters of the Wavelength Measurement

      View table
      View in Article

      Table 1. Key Parameters of the Wavelength Measurement

      ParametersValues
      Diffraction order of the grating k90
      Distance between groves d (mm)1/112.96
      Diffraction angle of the grating θ (deg.)79
      Pixel length (μm)25
      Focal length f (mm)200
    Tools

    Get Citation

    Copy Citation Text

    Zhijun Yuan, Haibo Zhang, Ren Ye, Jun Zhou, Yunrong Wei, Qihong Lou, "Wavelength calibration of narrowband ArF laser with iron hollow cathode lamp," Chin. Opt. Lett. 15, 071402 (2017)

    Download Citation

    EndNote(RIS)BibTexPlain Text
    Save article for my favorites
    Paper Information

    Category: Lasers and Laser Optics

    Received: Dec. 5, 2016

    Accepted: Mar. 30, 2017

    Published Online: Jul. 20, 2018

    The Author Email: Jun Zhou (junzhousd@siom.ac.cn)

    DOI:10.3788/COL201715.071402

    Topics