Different from a grazing incidence system[
Chinese Optics Letters, Volume. 15, Issue 4, 043401(2017)
13.5 nm Schwarzschild microscope with high magnification and high resolution
A Schwarzschild microscope with a numerical aperture of 0.2 and a magnification of 130 in a 100 μm field of view (FOV) is designed and is working at 13.5 nm. Meanwhile, a CCD is used as a detector with a pixel size of
Different from a grazing incidence system[
The imaging microscope that will be researched in this Letter should not only focus on eliminating spherical aberrations and comas, but also consider the effects on the Schwarzschild system of magnification, numerical aperture (NA), and FOV. The FOV is inversely proportional to the magnification, so they have rarely been considered at the same time. In 2000, Artioukov
Many different detectors were used on the abovementioned experiments. At present, a CCD as a detector in a Schwarzschild microscope became mainstream and has several remarkable advantages, such as high sensitivity, wide dynamic ranges, fast response speed, and real-time imaging, without reprocessing. But due to the limited pixel size of a CCD, the imaging resolution was rarely up to 200 nm by using a CCD as a detector in past experiments. In this Letter, in order to make the image of the mesh edge resolution better than before in the laboratory, we use a PIXIS-XO optical CCD as a detector with an imaging area of
Sign up for Chinese Optics Letters TOC Get the latest issue of Advanced Photonics delivered right to you!Sign up now
The optical structure of a Schwarzschild microscope is shown in Fig.
Figure 1.Optical structure of the Schwarzschild microscope.
The spherical mirror imaging formulas are
If we want to obtain a 200 nm mesh edge image by the CCD, which can distinguish two pixels in most instances, so one pixel of the CCD in the object must be up to 100 nm, the magnification
Meanwhile, the object resolution
This requires the image spatial frequency to be better than 39 lp/mm at modulation transfer function (MTF)
The optical diffraction-limited resolution is decided by the NA, which is calculated by
The greater the NA, the better the resolution, so we set NA at 0.2. From Eqs. (
|
The image MTF curve can be obtained using the Zemax optical software and is shown in Fig.
Figure 2.Image MTF curve of the Schwarzschild objective.
The image spatial resolution of this system is 78 lp/mm (
Because the CCD imaging area is
Figures
Figure 3.Object resolutions along with the change of object distance at different FOVs: (a) 1 to 1200 nm resolution; (b) zoomed-in view of the 0 to 450 nm resolutions in (a).
The abscissa 0 point in Figs.
Figure
Figure 4.Object resolutions along with the change of image distance at different FOVs.
The abscissa 0 point in Fig.
There are three errors that influence the imaging quality of the Schwarzschild objective: one is the figure error of the mirrors and the others are mirror-adjustment errors (centering error and position error on the axes). To achieve a high object resolution, the RMS value of the figure error of the mirrors must be small enough. We bought the spherical mirrors from Winlight company[
Figure 5.MTF of different errors.
From Fig.
By the same formula, the object resolution with figure error
Considering these errors, the ultimate object resolution of the Schwarzschild objective can be calculated by
This experiment is completed in a vacuum chamber, the 13.5 nm EUV light comes from a Cu cylindrical target that is irradiated by an Nd:YAG laser with 800 mJ of a single laser pulse energy and 10 ns pulse width. The peak reflectivity of the Mo/Si multilayer can achieve 65% at 13.5 nm, so the total transmission of the Schwarzschild objective is about 42%. A Zr filter of 0.4 μm thickness is used for letting 13.5 nm light pass away and eliminating visible light. The best object resolution needs an accurate object distance, which is hard to find indeed. Meanwhile, the image distance is a fixed value in the vacuum chamber. We use a high-precision 1-axis nanopositioning system by Physik Instrumente (PI) (GmbH & Co. KG) to maintain the precision of the object distance. The moving range of the nanopositioning system with a 0.01 μm precision is 100 μm. Then we put a freestanding gold grid on it as an object. The experimental schematic is shown in Fig.
Figure 6.Schematic of the source and the Schwarzschild imaging system.
In order to guarantee enough EUV light intensity on the CCD, we use the least number of reflectors possible in the light path and the grid is put as close as possible to Cu target. The real precision of nanopositioning with vibration error is 0.5 μm, so we find the best object distance through the change of nanopositioning every 0.5 μm; the image is shown in Fig.
Figure 7.Best imaging figure and the intensity profile.
In Fig.
Because the object distance has a great influence on the imaging quality, we move the object distance every 0.5 μm from the best point on the nanopositioning to get some images; the compared results between the simulation and experiment are shown in Fig.
Figure 8.Comparison of the results between the simulation and experiment.
The experimental results are all integer multiples of the CCD pixel, so it is slightly worse than the simulative results with the abovementioned three errors of mirrors. At the abscissa 0 point, the simulative resolution is obviously better than the experimental resolution that is limited by the CCD pixel size.
In order to increase the precision of the results, the best object points were imaged ten times in each experiment.
From Fig.
Figure 9.Best object points are imaged ten times.
In conclusion, a Schwarzschild microscope is designed to image a mesh edge at a 13.5 nm wavelength. The NA and magnification of the microscope are 0.2 and 130, respectively. A CCD is used as a detector with an imaging area of
[1] S. Chen, S. Ma, Z. Wang. Chin. Opt. Lett., 14, 123401(2016).
[23] Z. Wang, J. Cao, B. Chen, Y. Ma, B. Chen, J. Zhang, Z. Wang, H. Gao, J. Lv, X. Chen. Acta Opt. Sin., 16, 531(1996).
[26] X. Wang, B. Mu, Y. Huang, J. Zhu, Z. Wang, P. He. Opt. Precis. Eng., 8, 19(2011).
[28]
Get Citation
Copy Citation Text
Shenghao Chen, Xin Wang, Qiushi Huang, Shuang Ma, Zhanshan Wang, "13.5 nm Schwarzschild microscope with high magnification and high resolution," Chin. Opt. Lett. 15, 043401 (2017)
Category: X-ray Optics
Received: Nov. 2, 2016
Accepted: Jan. 24, 2017
Published Online: Jul. 25, 2018
The Author Email: Zhanshan Wang (wangzs@tongji.edu.cn)