Optics and Precision Engineering, Volume. 33, Issue 11, 1691(2025)

Enhancing sensitivity of multi-pass cavity enhanced Raman gas detection

Shangjin LI1,2, Tenglong DING1,2, Yanqun ZHANG1,2, Dianqiang SU1,2, Zhonghua JI1,2, Weiguang MA1,2, and Yanting ZHAO1,2、*
Author Affiliations
  • 1State Key Laboratory of Quantum Optics and Quantum Optics Technologies and Devices, Institute of Laser Spectroscopy, Shanxi University, Taiyuan030006, China
  • 2Collaborative Innovation Center of Extreme Optics, Shanxi University, Taiyuan030006, China
  • show less
    Figures & Tables(12)
    Schematic diagram of the Raman apparatus with multiple reflection cavity enhancement
    Optical simulation and actual spot patterns of different reflection modes
    Comparison of Raman scattering signal intensities for three reflection modes
    Effect of collection system angle on intensity of Raman scattering signals
    Influence of P-polarization and S-polarization on intensity of Raman scattering signals
    Signal intensity of single-path collection and dual-path collection in collection system
    Raman spectra of carbon dioxide in air
    Raman spectra of liquefied petroleum gas
    Relationship between Raman scattering signal intensity and gas concentration
    • Table 1. Signal intensity and signal-to-noise ratio for three different reflection modes

      View table
      View in Article

      Table 1. Signal intensity and signal-to-noise ratio for three different reflection modes

      Reflection modeSignal strength/countsNoise intensity/countsSNR
      Coelom9564023.9
      Near-common chamber elliptical2 4094356.0
      Near-symcentric chamber linear3 4894675.8
    • Table 2. Signal intensity and signal-to-noise ratio for five collection system angles

      View table
      View in Article

      Table 2. Signal intensity and signal-to-noise ratio for five collection system angles

      Collect system angle/(°)Signal strength/countNoise intensity/countSNR
      3022 00187252.9
      4515 01380187.7
      607 5657897.0
      753 3978341.0
      901 4698218.2
    • Table 3. Signal intensity and signal-to-noise ratio for dual-path collection and single-path collection

      View table
      View in Article

      Table 3. Signal intensity and signal-to-noise ratio for dual-path collection and single-path collection

      Branch collection and dual collectionSignal strength/countNoise intensity/countSNR
      A road25 21776331.8
      B road21 14872293.7
      B road+A road40 34481498.1
    Tools

    Get Citation

    Copy Citation Text

    Shangjin LI, Tenglong DING, Yanqun ZHANG, Dianqiang SU, Zhonghua JI, Weiguang MA, Yanting ZHAO. Enhancing sensitivity of multi-pass cavity enhanced Raman gas detection[J]. Optics and Precision Engineering, 2025, 33(11): 1691

    Download Citation

    EndNote(RIS)BibTexPlain Text
    Save article for my favorites
    Paper Information

    Category:

    Received: Mar. 5, 2025

    Accepted: --

    Published Online: Aug. 14, 2025

    The Author Email:

    DOI:10.37188/OPE.20253311.1691

    Topics