Laser & Optoelectronics Progress, Volume. 61, Issue 23, 2312004(2024)

Improved Quadratic Normalized for Eliminating Effect of Light Source Fluctuation

Hongbo Gao1, Xiefei Shi1、*, Peng Wang2, Xiaoman Shen1, Jihua Yang1, Limei Zeng1, and Weiguo Zhao3
Author Affiliations
  • 1Institute of Optics, National Institute of Measurement and Testing Technology, Chengdu 610021, Sichuan , China
  • 2Institute of Flow, National Institute of Measurement and Testing Technology, Chengdu 610021, Sichuan , China
  • 3College of Metrology Measurement and Instrument, China Jiliang University, Hangzhou 310000, Zhejiang , China
  • show less
    Figures & Tables(9)
    Structure of proposed new test system
    Structure and optical path of QNMD
    Experimental facility
    Reflex response curve of light source undulation experiment
    Quadratic normalized coefficient
    • Table 1. Response of standard white plate when wavelength is 550 nm

      View table

      Table 1. Response of standard white plate when wavelength is 550 nm

      Detection timesReflex responseIncident responseReflection coefficient
      189.25992.2750.9676
      289.28292.314
      389.28792.222
      489.29692.268
      589.31692.295
      689.30592.330
      789.30092.244
      889.27192.308
      989.30292.292
      1089.31692.285
    • Table 2. Response of the sample when wavelength is 550 nm

      View table

      Table 2. Response of the sample when wavelength is 550 nm

      Detection timesReflex responseIncident responseReflection coefficient
      177.88277.8870.9243
      284.30384.293
      377.88277.887
      484.30384.293
      577.88277.887
      684.30384.293
      777.88277.887
      884.30384.293
      977.88277.887
      1084.30384.293
    • Table 3. Comparison of measurement results

      View table

      Table 3. Comparison of measurement results

      ParameterStandard valueMeasured value
      X84.4184.54
      Y89.4289.66
      Z95.5195.84
      x0.31340.3131
      y0.33200.3320
    • Table 4. Comparison between the new and old methods

      View table

      Table 4. Comparison between the new and old methods

      ParameterStandard valueOld methodNew method
      Measured valueRelative error /%Measured valueRelative error /%
      X84.4185.541.3484.840.51
      Y89.4291.081.8689.880.51
      Z95.5196.601.1496.000.51
      x0.31340.31310.100.31340
      y0.33200.33360.480.33200
    Tools

    Get Citation

    Copy Citation Text

    Hongbo Gao, Xiefei Shi, Peng Wang, Xiaoman Shen, Jihua Yang, Limei Zeng, Weiguo Zhao. Improved Quadratic Normalized for Eliminating Effect of Light Source Fluctuation[J]. Laser & Optoelectronics Progress, 2024, 61(23): 2312004

    Download Citation

    EndNote(RIS)BibTexPlain Text
    Save article for my favorites
    Paper Information

    Category: Instrumentation, Measurement and Metrology

    Received: Jan. 12, 2024

    Accepted: Apr. 3, 2024

    Published Online: Nov. 14, 2024

    The Author Email: Xiefei Shi (21968664@qq.com)

    DOI:10.3788/LOP240507

    CSTR:32186.14.LOP240507

    Topics