Acta Optica Sinica, Volume. 44, Issue 4, 0412006(2024)

Anti-Vibration White Light Interferometry Based on Non-Uniform Fast Fourier Transform

Liang Xue, Renhui Guo*, Yang Liu, Yu Qian, Jinwei Jiang, and Jianxin Li
Author Affiliations
  • School of Electronic and Optical Engineering, Nanjing University of Science and Technology, Nanjing 210094, Jiangsu , China
  • show less
    Figures & Tables(12)
    Dual-channel imaging micro-interference measurement optical path
    Flow diagram of interference signal correction algorithm
    Comparison of interference data under ideal and vibration conditions. (a) Phase shift interval; (b) interference light intensity of white light interference signal
    White light interference light intensity curve before and after correction. (a) Before correction; (b) after correction
    Simulate results of step surface restoration. (a) Restoration result of step surface with vibration; (b) result of step surface restoration after correction by NUFFT algorithm; (c) one-dimensional contour of step faces; (d) deviation from one-dimensional contour of step face
    Simulate results of step surface restoration. (a) Restoration result of step surface with vibration; (b) result of step surface restoration after correction by NUFFT algorithm
    White light interferogram. (a) Complete white light interferogram; (b) white light interferogram of area near step face
    Quasi-monochromatic light interferogram processing results. (a) Phase shift interval; (b) phase shift interval deviation
    White light interference light intensity curve before and after correction. (a) Before correction; (b) after correction
    Reference results and NUFFT algorithm correction results. (a) Measurement result of Veeco interferometer; (b) step morphology calculated before correction; (c) step morphology calculated by NUFFT algorithm after correction ; (d) one-dimensional contour of step surface obtained by Veeco interferometer; (e) one-dimensional contour of step surface obtained after NUFFT algorithm correction
    • Table 1. Measurement results of front and rear step surfaces before and after correction

      View table

      Table 1. Measurement results of front and rear step surfaces before and after correction

      Whether to correctAverage height /μmRRt /μmSSq /μm

      Height deviation /

      %

      Veeco reference0.13930.25440.03580.21
      After correction0.14190.24200.03812.08
      Before correction0.15460.38030.104911.22
    • Table 2. Measurement results of repeatability experiment on post-correction step surface

      View table

      Table 2. Measurement results of repeatability experiment on post-correction step surface

      No.Average height /μmRRt /μmSSq /μm
      10.14190.24200.0381
      20.14080.24000.0378
      30.14030.23910.0377
      40.14280.24340.0385
      50.14220.24250.0382
      60.13990.23850.0376
      70.14160.24140.0380
      80.14110.24070.0379
      90.14010.23890.0377
      100.13960.23800.0375
    Tools

    Get Citation

    Copy Citation Text

    Liang Xue, Renhui Guo, Yang Liu, Yu Qian, Jinwei Jiang, Jianxin Li. Anti-Vibration White Light Interferometry Based on Non-Uniform Fast Fourier Transform[J]. Acta Optica Sinica, 2024, 44(4): 0412006

    Download Citation

    EndNote(RIS)BibTexPlain Text
    Save article for my favorites
    Paper Information

    Category: Instrumentation, Measurement and Metrology

    Received: Oct. 9, 2023

    Accepted: Dec. 13, 2023

    Published Online: Feb. 27, 2024

    The Author Email: Guo Renhui (grh@njust.edu.cn)

    DOI:10.3788/AOS231628

    CSTR:32393.14.AOS231628

    Topics