Optics and Precision Engineering, Volume. 33, Issue 7, 1051(2025)

Determining performance of radiographic examination system with duplex-wire type image quality indicator

Qingguo YANG*... Bozhong TAN, Xiaozhong HE, Ruo TANG, Liu YANG, Shuqing LIAO, Yunfei SUN, Chaofan MA and Jianjun DENG |Show fewer author(s)
Author Affiliations
  • Institute of Fluid Physics, China Academy of Engineering Physics, Mianyang621900, China
  • show less
    Figures & Tables(33)
    Schematic diagram of radiographic system
    Schematic diagram of duplex-wire section radiography
    Ideal and actual projection value curves of duplex-wire profile
    Contrast transfer function curve
    CTF curves for different source-object distances
    CTF curves for different offset distances of duplex-wire IQI with source-object distance of 100 mm
    CTF curves for different offset distance of duplex-wire IQI with source-object distance of 200 mm
    CTF curves for different ray energies
    CTF curves for LSF with normal distribution of different characteristic sizes
    Relationship between image unsharpness and characteristic size of line spread function with normal distribution
    Comparison between interpolated and fitted CTF curves
    Petal-type electron accelerator x-ray source
    DR image and average section curve of duplex-wire type IQI with magnification of 1.00
    CTF curves with magnification of 1.00
    DR image and average section curve of duplex-wire type IQI with magnification of 9.40
    CTF curves with magnification of 9.40
    LSF curves with magnification of 9.40
    MTF curves with magnification of 9.40
    DR image and average section curve of duplex-wire type IQI with magnification of 2.23
    CTF curves with magnification of 2.23
    LSF curves with magnification of 2.23
    MTF curves with magnification of 2.23
    DR image and average section curve of duplex-wire type IQI with magnification of 1.00
    CTF curves with magnification of 1.00
    DR image and average section curve of duplex-wire type IQI with magnification of 8.90
    CTF curves with magnification of 8.90
    LSF curves with magnification of 8.90
    MTF curves with magnification of 8.90
    DR image and average section curve of duplex-wire type IQI with magnification of 3.43
    CTF curves with magnification of 3.43
    LSF curves with magnification of 3.43
    MTF curves with magnification of 3.43
    • Table 1. Parameters of duplex-wiretype IQI

      View table
      View in Article

      Table 1. Parameters of duplex-wiretype IQI

      No.Line pair value/(lp·mm-1Wire diameter&spacing/mm
      D41.2500.40
      D51.5630.32
      D62.0000.25
      D72.5000.20
      D83.1250.16
      D93.8500.130
      D105.0000.100
      D116.2500.080
      D127.9370.063
      D1310.0000.050
      D1412.5000.040
      D1515.6250.032
      D1620.0000.025
      D1725.0000.020
    Tools

    Get Citation

    Copy Citation Text

    Qingguo YANG, Bozhong TAN, Xiaozhong HE, Ruo TANG, Liu YANG, Shuqing LIAO, Yunfei SUN, Chaofan MA, Jianjun DENG. Determining performance of radiographic examination system with duplex-wire type image quality indicator[J]. Optics and Precision Engineering, 2025, 33(7): 1051

    Download Citation

    EndNote(RIS)BibTexPlain Text
    Save article for my favorites
    Paper Information

    Category:

    Received: Dec. 23, 2024

    Accepted: --

    Published Online: Jun. 23, 2025

    The Author Email: Qingguo YANG (yungore@163.com)

    DOI:10.37188/OPE.20253307.1051

    Topics