Laser & Optoelectronics Progress, Volume. 56, Issue 12, 121201(2019)

Measurement of Lens-Center Thickness Based on Low-Coherence Interference with Transmitted Illumination

Jingyou Liu1,2、* and Feng Lei1,2
Author Affiliations
  • 1 School of Physics and Electronic Electrical Engineering, Huaiyin Normal University, Huai'an, Jiangsu 223300, China
  • 2 Key Laboratory of Huai'an's Micro-Nano Optical Imaging, Huai'an, Jiangsu 223300, China
  • show less
    Figures & Tables(9)
    Principle diagram for measuring lens-center thickness based on low-coherence interference with transmitted illumination
    Principle diagram for adjusting optical path difference using pair of wedge prisms
    Equal-path interference fringes before and after lens. (a) Before lens; (b) after lens
    [in Chinese]
    Method for extending measuring range of lens-center thickness
    • Table 1. Measuring precision under different conditions

      View table

      Table 1. Measuring precision under different conditions

      nPnΔx /μmθsinθΔtC /nm
      5°44'0.1±200
      nPn=nP±22°52'0.05±100
      1°26'0.025±50
      6°12'0.10799±200
      1.475901.51390±23°5'0.05399±100
      1°33'0.02700±50
      5°20'0.09260±200
      1.513901.47590±22°39'0.04630±100
      1°20'0.02315±50
    • Table 2. Measuring results of lens-center thickness based on pair of edge prisms with wedge angle of 10°0'

      View table

      Table 2. Measuring results of lens-center thickness based on pair of edge prisms with wedge angle of 10°0'

      No.Nominal value of tested lensMeasured value of center thickness
      Dimeter ϕ /mmFocal length f' /mmCenter thickness TC /mm(tC±δ) /mmRelative error /%
      525.4-150.03.0±0.12.9286±0.00090.03
      712.7-25.42.0±0.12.0151±0.00080.04
      912.7150.04.0±0.14.0581±0.00090.02
      1225.4100.04.0±0.14.0375±0.00090.02
      1325.4-150.03.0±0.12.7990±0.00110.04
      1425.4400.03.5±0.13.5294±0.00080.02
      1810.0-20.01.5±0.11.4743±0.00100.07
      196.0-9.801.0±0.11.1094±0.00060.05
    • Table 3. Measuring results of lens-center thickness based on pair of edge prisms with wedge angle of 5°30'

      View table

      Table 3. Measuring results of lens-center thickness based on pair of edge prisms with wedge angle of 5°30'

      No.Nominal value of tested lensMeasured value of center thickness
      Dimeter ϕ /mmFocal length f' /mmCenter thickness TC /mm(tC±δ) /mmRelative error /%
      525.4-150.03.0±0.12.9227±0.00020.01
      712.7-25.42.0±0.12.0139±0.00090.05
      1325.4-150.03.0±0.12.7980±0.00070.02
      1425.4400.03.5±0.13.5226±0.00090.02
      1810.0-20.01.5±0.11.4778±0.00060.04
      196.0-9.801.0±0.11.1044±0.00050.05
    • Table 4. Examples for expanding measuring range

      View table

      Table 4. Examples for expanding measuring range

      No.Nominal value of tested plate or lensMeasurementCorrected measurement
      Dimeterϕ /mmFocal lengthf' /mmCenter thicknessTC /mm(tC±δ) /mmRelativeerror /%(tC±δ) /mmRelativeerror /%
      1--2.2±0.12.2331±0.00080.04--
      2--3.8±0.13.8447±0.00110.03--
      825.41505.0±0.12.8073±0.00090.035.0404±0.00120.02
      1725.4308.1±0.14.2573±0.00110.028.1020±0.00160.02
    Tools

    Get Citation

    Copy Citation Text

    Jingyou Liu, Feng Lei. Measurement of Lens-Center Thickness Based on Low-Coherence Interference with Transmitted Illumination[J]. Laser & Optoelectronics Progress, 2019, 56(12): 121201