Journal of Applied Optics, Volume. 46, Issue 2, 282(2025)

Design and application of fluorescence concentration measurement system in large area water based on LIF

Haoping WAN1,2、*
Author Affiliations
  • 1Institute of Disaster Prevention, Mitigation and Water Engineering Safety, Jiangxi Academy of Water Science and Engineering, Jiangxi 330029, China
  • 2Jiangxi Provincial Technology Innovation Center for Ecological Water Engineering in Poyang Lake Basin, Jiangxi 330029, China
  • show less
    Figures & Tables(14)
    Hardware topology diagram of concentration measurement system
    Physical images of collection devices and sensors
    Structure diagram of photoelectric sensor
    Structure diagram of isolation reflector
    Measurement principle diagram of photoelectric sensor
    Schematic diagram of AC and DC amplification circuits
    Structure diagram of collection box
    Calibration flume
    Concentration-voltage relationship curves of 1#~10# sensors
    Simple experimental setup
    Sensors layout and effect diagram of concentration field
    • Table 1. Characteristic parameters of sensors

      View table
      View in Article

      Table 1. Characteristic parameters of sensors

      传感器编号kaR2
      1#48.5150.17581.000 0
      2#37.7560.62011.000 0
      3#40.5250.69580.9997
      4#55.7590.533 00.9998
      5#58.0780.59140.9992
      6#47.4180.927 00.9998
      7#29.7060.512 00.9994
      8#33.3030.65640.9999
      9#24.8110.10060.9995
      10#33.3010.54290.9983
    • Table 2. Concentration determination results

      View table
      View in Article

      Table 2. Concentration determination results

      传感器 编号样本水体浓度/(mg/L)
      0.0100.0150.0200.0250.0300.0350.040
      1#0.01040.01520.02090.02470.03130.03580.0412
      2#0.01050.01570.02030.02670.03130.03590.0419
      3#0.01050.01540.02090.02690.03110.03670.0409
      4#0.01070.01530.02050.02570.03140.03610.0407
      5#0.01060.01520.02060.02550.03150.03660.0418
      6#0.01050.01570.02040.02580.03130.03690.0426
      7#0.01030.01550.02050.02580.02950.03540.0429
      8#0.00980.01550.02080.02580.03130.03530.0435
      9#0.01090.01520.02070.02480.02970.03640.0406
      10#0.00970.01580.02060.02590.03080.03570.0396
    • Table 3. Sensor characteristic parameters in variable water

      View table
      View in Article

      Table 3. Sensor characteristic parameters in variable water

      传感器编号kaR2
      1#48.3100.19391.000 0
      2#37.5980.62360.9999
      3#40.8210.67430.9997
      4#56.0130.54360.9997
      5#57.8640.58620.9993
      6#47.7020.765 00.9998
      7#29.7860.50140.9994
      8#33.4640.648 00.9998
      9#24.8860.10840.9996
      10#33.1860.54860.9986
    Tools

    Get Citation

    Copy Citation Text

    Haoping WAN. Design and application of fluorescence concentration measurement system in large area water based on LIF[J]. Journal of Applied Optics, 2025, 46(2): 282

    Download Citation

    EndNote(RIS)BibTexPlain Text
    Save article for my favorites
    Paper Information

    Category:

    Received: Oct. 8, 2024

    Accepted: --

    Published Online: May. 13, 2025

    The Author Email: Haoping WAN (whp0985@163.com)

    DOI:10.5768/JAO202546.0201007