Acta Optica Sinica, Volume. 40, Issue 7, 0712001(2020)

Synchronous Measurement Method of Trans-Micron Scale Mixed Particle Size

Rong Zhao1, Kewei Pan2, Bin Yang1、*, Li Ping1, and Xiaoshu Cai1
Author Affiliations
  • 1Shanghai Key Laboratory of Multiphase Flow and Heat Transfer in Power Engineering, University of Shanghai for Science and Technology, Shanghai 200093, China
  • 2Solid Rocket Motor Technology Development Research Center, Shanghai Space Propulsion Technology Research Institute, Shanghai 201109, China
  • show less
    Figures & Tables(8)
    Structural diagram of measurement system
    Schematic of extinction measurement
    Schematic of particle two-phase flow measurement by image method
    Typical signal extinction spectra
    Inverse results of particle size distribution by typical signal extinction method. (a) Single 900 nm standard particle sample solution; (b) mixed 900 nm + 51 μm standard particle sample solution
    Typical image preprocessing. (a) Filtered and de-noised image; (b) grayscale normalized image; (c) background homogenized image; (d) binarized image
    Measurement results of particle size distribution by typical signal image method
    • Table 1. Particle size measurement results and relative errors

      View table

      Table 1. Particle size measurement results and relative errors

      SamplenumberStandard particle sizeMulti-wavelength lightextinction methodImage method
      Submicron-tenmicronsOver tenmicronsMeasurementresultRelativeerror /%MeasurementresultRelativeerror /%
      1900 nm76.9 μm(864.8±4.7) nm3.91(76.71±0.16) μm0.25
      2900 nm51.0 μm(880.1±2.7) nm2.21(52.33±0.10) μm2.61
      3900 nm15.1 μm(874.1±9.8) nm2.88(15.60±0.08) μm3.31
      4500 nm51.0 μm(490.9±2.8) nm1.81(52.00±0.17) μm1.96
      5700 nm51.0 μm(685.1±2.4) nm2.13(52.05±0.16) μm2.05
      62.1 μm51.0 μm(2.18±0.01) μm3.59(52.14±0.16) μm2.24
      73.1 μm51.0 μm(3.20±0.01) μm3.38(52.10±0.17) μm2.16
      85.1 μm51.0 μm(5.29±0.01) μm3.65(52.31±0.13) μm2.57
      99.7 μm51.0 μm(9.29±0.04) μm4.19(52.07+0.18) μm2.10
    Tools

    Get Citation

    Copy Citation Text

    Rong Zhao, Kewei Pan, Bin Yang, Li Ping, Xiaoshu Cai. Synchronous Measurement Method of Trans-Micron Scale Mixed Particle Size[J]. Acta Optica Sinica, 2020, 40(7): 0712001

    Download Citation

    EndNote(RIS)BibTexPlain Text
    Save article for my favorites
    Paper Information

    Category: Instrumentation, Measurement and Metrology

    Received: Oct. 17, 2019

    Accepted: Dec. 9, 2019

    Published Online: Apr. 15, 2020

    The Author Email: Yang Bin (yangbin@usst.edu.cn)

    DOI:10.3788/AOS202040.0712001

    Topics