Chinese Journal of Lasers, Volume. 45, Issue 7, 0707001(2018)

Tissue Intrinsic Fluorescence Spectrum Recovery Algorithm and Its Application in Diabetes Screening

Yang Zhang1,2, Jingshu Ni1, Yuanzhi Zhang、* *, Zhaohui Fang3, Yikun Wang1,2, and Yong Liu1
Author Affiliations
  • 1 Institute of Applied Technology, Hefei Institutes of Physical Science, Chinese Academy of Sciences, Anhui Provincial Engineering Technology Research Center for Biomedical Optical Instrument, Institute of Applied Technology, Hefei, Anhui 230031, China
  • 2 Wan Jiang New Industy Technology Development Center, Tongling, Anhui 244000, China
  • 3 The First Affiliated Hospital of Anhui University of Chinese Medicine, Hefei, Anhui 230031, China
  • show less
    Figures & Tables(7)
    Schematics for measuring fluorescence and diffuse reflection spectra of tissue. (a) Fluorescence spectrum; (b) diffuse reflection spectrum
    Schematic of spectrum detection system
    Spectra of phantom sets with 0.96 mg/mL NADH and varying concentrations of Hb and polystyrene. (a)-(b) Fluorescence spectra; (c)-(d) diffuse reflection spectra of fluorescence excitation band; (e)-(f) diffuse reflection spectra of fluorescence emission band
    Relationship between the (a) phantom raw fluorescence or (b) fluorescence spectrum after recovery and mass concentration of NADH
    ROC curves of correction spectrum (kx=0.9, km=-0.8) and original spectrum (kx=0, km=0) for diabetes screening
    • Table 1. Composition of single-layer phantom

      View table

      Table 1. Composition of single-layer phantom

      PhantomnumberMass concentration /(mg·mL-1)
      NADHHbPolystyrene
      1X0.50.02
      2X1.00.02
      3X1.50.02
      4X0.50.04
      5X1.00.04
      6X1.50.04
      7X0.50.06
      8X1.00.06
      9X1.50.06
      10X00
    • Table 2. Absorption coefficient μa and scattering coefficientμs at typical excitation and emission wavelengths of NADH

      View table

      Table 2. Absorption coefficient μa and scattering coefficientμs at typical excitation and emission wavelengths of NADH

      Excitation wavelengthof 370 nmEmission wavelengthof 480 nm
      μa /cm-1μs /cm-1μa /cm-1μs /cm-1
      251082.692
      502175.2185
      753267.8277
    Tools

    Get Citation

    Copy Citation Text

    Yang Zhang, Jingshu Ni, Yuanzhi Zhang, Zhaohui Fang, Yikun Wang, Yong Liu. Tissue Intrinsic Fluorescence Spectrum Recovery Algorithm and Its Application in Diabetes Screening[J]. Chinese Journal of Lasers, 2018, 45(7): 0707001

    Download Citation

    EndNote(RIS)BibTexPlain Text
    Save article for my favorites
    Paper Information

    Category: biomedical photonics and laser medicine

    Received: Jan. 31, 2018

    Accepted: --

    Published Online: Sep. 11, 2018

    The Author Email: Zhang Yuanzhi (yzzhang@aiofm.ac.cn)

    DOI:10.3788/CJL201845.0707001

    Topics