Laser & Optoelectronics Progress, Volume. 55, Issue 5, 053003(2018)

Measurements of CO2 Temperature and Concentration in High Temperature Environment Based on Tunable Diode Laser Absorption Spectroscopy

Haibin Cui*, Fei Wang, and Meiyi Li
Author Affiliations
  • College of Energy Engineering, State Key Laboratory of Clean Energy Utilization, Zhejiang University, Hangzhou, Zhejiang 310027, China
  • show less
    Figures & Tables(11)
    Absorption line strength of (a) CO, (b) H2O and (c) CO2 in the wavelength range of 1-3 μm
    Simulated absorption spectra
    Simulated results of line pairs. (a) Line pair 1; (b) line pair 2; (c) line pair 3
    (a) Variations of line strength with temperature; (b) line strength ratio and their sensitivity with temperature
    Diagram of the high temperature measurement system
    (a) Spectral absorption signal and multi-peak fitting results using Voigt function and (b) fitting residual
    • Table 1. CO2 spectral absorption-line parameters

      View table

      Table 1. CO2 spectral absorption-line parameters

      LinepairCenterwavenumberν0 /cm-1Centerwavelengthλ /nmLine-intensityS0 (T= 900 K) /(10-7 cm-2·Pa-1)Low transitionenergy E″/cm-1
      14990.4182003.84022.812462.6758
      4990.6652003.74106.551276.4475
      24995.97022001.61324.331702.4031
      4995.97982001.60940.1514059.1287
      4995.98442001.60754.91586.8582
      4996.10852001.557824.2273.868
      35001.49871999.400722.7519.535
      5001.91591999.23392.382193.5452
      5001.94681999.22161.482551.011
      5002.48691999.005721.8578.0115
    • Table 2. Measured temperature under different operating conditions

      View table

      Table 2. Measured temperature under different operating conditions

      Volume fraction ofCO2/%Measured temperature /K
      873 K973 K1073 K1173 K1273 K
      10881.6994.41087.81160.01260.1
      8868.9946.81055.61160.91284.6
      6890.7996.71100.31205.71241.6
      4898.3950.11042.11135.31228.6
    • Table 3. Deviation of measured temperatures with given temperature

      View table

      Table 3. Deviation of measured temperatures with given temperature

      Volume fraction ofCO2 /%Deviation /%
      873 K973 K1073 K1173 K1273 K
      100.982.201.38-1.11-1.01
      8-0.47-2.70-1.62-1.030.91
      62.032.442.552.79-2.47
      42.90-2.35-2.88-3.21-3.49
    • Table 4. Measured CO2 concentration under different operating conditions

      View table

      Table 4. Measured CO2 concentration under different operating conditions

      Volume fraction ofCO2/%Measured volume fraction of CO2
      873 K973 K1073 K1173 K1273 K
      109.909.809.8610.1410.20
      88.128.158.098.178.14
      66.126.186.086.115.89
      44.123.873.893.853.87
    • Table 5. Deviation of measured CO2 concentration with given value under different operating conditions

      View table

      Table 5. Deviation of measured CO2 concentration with given value under different operating conditions

      Volume fraction ofCO2 /%Deviation /%
      873 K973 K1073 K1173 K1273 K
      10-1.00-2.00-1.401.402.00
      81.501.881.132.131.75
      62.003.001.331.83-1.83
      43.75-3.25-4.75-3.75-3.25
    Tools

    Get Citation

    Copy Citation Text

    Haibin Cui, Fei Wang, Meiyi Li. Measurements of CO2 Temperature and Concentration in High Temperature Environment Based on Tunable Diode Laser Absorption Spectroscopy[J]. Laser & Optoelectronics Progress, 2018, 55(5): 053003

    Download Citation

    EndNote(RIS)BibTexPlain Text
    Save article for my favorites
    Paper Information

    Category: Spectroscopy

    Received: Oct. 13, 2017

    Accepted: --

    Published Online: Sep. 11, 2018

    The Author Email: Haibin Cui (cuihaibin1991@zju.edu.cn)

    DOI:10.3788/LOP55.053003

    Topics