Laser & Optoelectronics Progress, Volume. 60, Issue 7, 0712005(2023)

Research on Imaging Quality Comprehensive Evaluation by Multi-Parameters of Line Structured Light Stripe

Taotao Li1、*, Pei Shen1, Xiaoqian Zhong1, Xianlei Xu2, and Jiangying Liu1
Author Affiliations
  • 1School of Electromechanical Engineering, Pingxiang University, Pingxiang 337055, Jiangxi, China
  • 2State Key Laboratory of Coal Resources and Safe Mining, China University of Mining and Technology (Beijing), Beijing 100083, China
  • show less
    Figures & Tables(11)
    One-dimensional image entropy changes of light stripe image set
    Local contrast changes of light stripe image set
    Definition coefficient changes of light stripe image set
    Synchronous changes of light stripe image set's H, C, β,and stripe centers extraction deviation Ra
    Light stripe imaging results of metal plate under different exposure time t. (a) t=200 μs; (b) t=1000 μs; (c) t=2000 μs; (d) t=2500 μs; (e) t=5000 μs
    Hign-precision rack
    3D measurement results of 7th tooth in high-precision rack. (a) 3D reconstruction of tooth profile; (b) Rssa varies with t
    • Table 1. Corresponding t and H ranges of light stripe imaging quality empirical evaluation results

      View table

      Table 1. Corresponding t and H ranges of light stripe imaging quality empirical evaluation results

      Characteristic parameterHigh quality light stripe imagePoor quality light stripe image
      Region #2Region #1Region #3
      Exposure time t /μs500-4000100-5004000-15000
      One-dimensional image entropy H2.73-5.461.68-2.735.46-7.06
      Image local contrast C0.5-0.70.7-1.00-0.5
    • Table 2. Value ranges of H, C, β,and Ra of high quality light stripe images in image set U

      View table

      Table 2. Value ranges of H, C, β,and Ra of high quality light stripe images in image set U

      Evaluation parameterValue range
      Exposure time t /μs1200 ≤ t ≤ 2700
      One-dimensional image entropy H3.81 ≤ H ≤ 4.90
      Image local contrast C0.50 ≤ C ≤ 0.56
      Light stripe definition coefficient β0.95 ≤ β ≤ 1.00
      Stripe centers extraction deviation Ra0.26 ≤ Ra ≤ 0.30
    • Table 3. Multi-parameters comprehensive evaluation results of light stripe images in Fig. 5

      View table

      Table 3. Multi-parameters comprehensive evaluation results of light stripe images in Fig. 5

      Exposure time t /μsOne-dimensional image entropy HImage local contrast CLight stripe definition coefficient βComprehensive evaluation result E
      2002.490.980.3135.2
      10003.980.560.9497.2
      20004.640.510.9598.4
      25004.910.500.9396.3
      50006.230.460.8570.8
    • Table 4. Comparison of evaluation results of different methods

      View table

      Table 4. Comparison of evaluation results of different methods

      MethodRelevant evaluation parameterExposure time t /μsVariation range of Rssa
      Method 1α≈4.52600~32000.057~0.090
      Method 2R≈160,D≈2100800~20000.022~0.120
      Method 3Llaser≈2102200~30000.034~0.060
      Proposed method3.81 ≤ H ≤ 4.10,0.52 ≤ C ≤ 0.56,0.95 ≤ β ≤ 1.001200~24000.022~0.042
    Tools

    Get Citation

    Copy Citation Text

    Taotao Li, Pei Shen, Xiaoqian Zhong, Xianlei Xu, Jiangying Liu. Research on Imaging Quality Comprehensive Evaluation by Multi-Parameters of Line Structured Light Stripe[J]. Laser & Optoelectronics Progress, 2023, 60(7): 0712005

    Download Citation

    EndNote(RIS)BibTexPlain Text
    Save article for my favorites
    Paper Information

    Category: Instrumentation, Measurement and Metrology

    Received: Feb. 28, 2022

    Accepted: Mar. 30, 2022

    Published Online: May. 24, 2023

    The Author Email: Taotao Li (ltaotao1988@126.com)

    DOI:10.3788/LOP220829

    Topics