Acta Optica Sinica, Volume. 39, Issue 10, 1011002(2019)

Three-Dimensional Reconstruction of Micro-Scale Flow Field Based on Light Field Microscopic Imaging

Xianglei Song1, Shu Li2, Mengtao Gu1, Biao Zhang1, and Chuanlong Xu1、*
Author Affiliations
  • 1Key Laboratory of Energy Thermal Conversion and Control of Ministry of Education, School of Energy and Environment, Southeast University, Nanjing, Jiangsu 210096, China
  • 2Department of Science and Technology Development, Nanjing Institute of Measurement and Testing Technology, Nanjing, Jiangsu 210037, China
  • show less
    Figures & Tables(17)
    Schematic of the light field microscopic imaging (fo, ft and fμ are the focal length of the objective lens, the tube lens and the microlens, respectively)
    Schematic of the light field Micro-PIV system (Ll is the flange distance of C lens mount)
    Schematic of the light field microscopic imaging and the deconvolution reconstruction for micro-scale flow
    Imaging process of a point light source in light field microscopic imaging system
    PSFs of the light field microscopic imaging system obtained by calculation at different depths
    Photography of the experimental system
    Assembled cage light field camera system. (a) Schematic; (b) photography
    PSFs of the light field microscopic imaging system obtained by experiments at different depths
    VSSIM evaluation of PSFs calculated based on wave optics theory and obtained by experiment
    Reconstructed particle at z=40 μm. (a) Reconstructed tracer particle light intensity distribution; (b) front view; (c) side view
    Variation of the length of elongation with the defocusing distance (Rx and Rz are the results of theoretical calculation, and Rxe and Rze are the results of experiments)
    Variation of the reconstructed position error with the defocusing distance
    Light field image of the tracer particles and reconstruction by simulation at particle concentration of 2.75×107 cm-3. (a) Light field image of tracer particles; (b) reconstructed tracer particles
    Reconstructed position errors of tracer particles at x and y directions
    Reconstructed position errors of tracer particles at z direction
    Light field images and reconstructed tracer particles by experiments. (a)(c) Light field images obtained at particle concentration of 2.75×106 cm-3 and 2.75×107 cm-3, respectively; (b)(d) reconstructed tracer particles at particle concentration of 2.75×106 cm-3 and 2.75×107 cm-3, respectively
    • Table 1. Specific parameters of the light field microscopic imaging system

      View table

      Table 1. Specific parameters of the light field microscopic imaging system

      EquipmentSymbolParameterValue
      MicroscopeMObjective lens magnification10
      NANumerical aperture of objective lens0.3
      foFocal length of objective lens /mm20
      ftFocal length of tube lens /mm200
      Cameranpx×npyResolution2352×1768
      PpPixel pitch /μm5.5
      Microlens arrayDMicrolens pitch /μm136
      fμFocal length of microlens /μm2260
    Tools

    Get Citation

    Copy Citation Text

    Xianglei Song, Shu Li, Mengtao Gu, Biao Zhang, Chuanlong Xu. Three-Dimensional Reconstruction of Micro-Scale Flow Field Based on Light Field Microscopic Imaging[J]. Acta Optica Sinica, 2019, 39(10): 1011002

    Download Citation

    EndNote(RIS)BibTexPlain Text
    Save article for my favorites
    Paper Information

    Category: Imaging Systems

    Received: Apr. 4, 2019

    Accepted: Jun. 21, 2019

    Published Online: Oct. 9, 2019

    The Author Email: Xu Chuanlong (chuanlongxu@seu.edu.cn)

    DOI:10.3788/AOS201939.1011002

    Topics