Acta Optica Sinica, Volume. 44, Issue 24, 2428002(2024)

Center Wavelength Calibration of Fabry‒Perot Filter Based on Dual-Wavelength Intensity Modulation

Jin Zhu1,2, Lin Ma1,2、*, Zhengyu Jin1, and Hengyu Mu1,2
Author Affiliations
  • 1Yunnan Observatories, Chinese Academy of Sciences, Kunming 650216, Yunnan , China
  • 2University of Chinese Academy of Sciences, Beijing 100049, China
  • show less
    Figures & Tables(13)
    Intensity modulation curves of F-P filter
    Dual-wavelength intensity modulation curves of F-P filter
    Schematic of F-P filter calibration system
    Photograph of F-P filter calibration system
    Laser spot images at different controller values. (a) The -823rd step; (b) the -813th step; (c) the -803rd step
    Measured dual-wavelength intensity modulation curve
    Lorentz function fitting curve
    Relationship between cavity length of F-P filter and controller value
    Total error distribution of intensity modulation curve
    • Table 1. Specifications of ET100-FS-100 F-P filter

      View table

      Table 1. Specifications of ET100-FS-100 F-P filter

      ParameterValue
      Wavelength range /Å5000‒7000
      Fineness43.28@6330 Å
      Coating reflectivity /%93
      Coating absorptivity /%0.3
      Cavity length /μm1000
      Scan range /μm2 (electronic control)
      ControllerCS100
      Controller resolution /bit12
    • Table 2. Mean hstep, hstep error, and Δn error

      View table

      Table 2. Mean hstep, hstep error, and Δn error

      ParameterValue
      Mean hstep of 6329.92 Å /pm429.88
      Mean hstep of 6578.22 Å /pm429.73
      Mean hstep of dual-wavelength /pm429.80
      hstep error /pm0.05
      Δn error /step0.1
    • Table 3. Relationship between cavity length of F-P filter and controller value

      View table

      Table 3. Relationship between cavity length of F-P filter and controller value

      Peaknh /μm
      P1-812.9081001.50220
      P2-475.5461001.64712
      P1'-76.6331001.82168
      P2'289.8711001.97914
    • Table 4. Electronic component errors of F-P filter and resulting intensity errors

      View table

      Table 4. Electronic component errors of F-P filter and resulting intensity errors

      ParameterElectronic component errorhsteperror /pmStandard deviation /%
      Electrical noise±90 pm±90±0.37
      Quantization error±0.5κlsb±244.2±1.01
      Non-linearity error±0.05%±0.2442±0.001
    Tools

    Get Citation

    Copy Citation Text

    Jin Zhu, Lin Ma, Zhengyu Jin, Hengyu Mu. Center Wavelength Calibration of Fabry‒Perot Filter Based on Dual-Wavelength Intensity Modulation[J]. Acta Optica Sinica, 2024, 44(24): 2428002

    Download Citation

    EndNote(RIS)BibTexPlain Text
    Save article for my favorites
    Paper Information

    Category: Remote Sensing and Sensors

    Received: Dec. 25, 2023

    Accepted: Apr. 17, 2024

    Published Online: Dec. 17, 2024

    The Author Email: Ma Lin (malin@ynao.ac.cn)

    DOI:10.3788/AOS231981

    CSTR:32393.14.AOS231981

    Topics