Journal of Atmospheric and Environmental Optics, Volume. 19, Issue 2, 175(2024)

Establishment of ozone- and nitrous oxide-based actinometry for quantifying vacuum ultra-violate radiation intensities and its application

YANG Dongsen1、#, LI Wanying2、#, CHEN Jiangyao2, LIU Xinran1, and ZHENG Jun1
Author Affiliations
  • 1School of Environmental Science and Engineering, Nanjing University of Information Science & Technology,Nanjing 210044, China
  • 2Institute of Environmental Health and Pollution Control, School of Environmental Science and Engineering, Guangdong University of Technology, Guangzhou 510006, China
  • show less
    Figures & Tables(6)
    Schematic of ultra-violet radiation intensities calibration instruments based on actinometry
    Nitrous oxide-based ultra-violet radiation intensities calibration results
    Ozone-based calibration of ultra-violet radiation intensities calibration results
    Calibration results of gaseous sulfuric acid during field measurement based on actinometry
    • Table 1. The value and uncertainty of reaction rate constant kn at 25 °C

      View table
      View in Article

      Table 1. The value and uncertainty of reaction rate constant kn at 25 °C

      knReaction rate/(cm3·molecule-1·s-1)Uncertainty/%
      k24.0 × 10-11±5
      k32.6 × 10-11±10
      k57.2 × 10-11±10
      k64.4 × 10-11±10
      k7< 1.0 × 10-12±10
    • Table 2. Uncertainty source of ultra-violet radiation intensities calibration based on actinometry

      View table
      View in Article

      Table 2. Uncertainty source of ultra-violet radiation intensities calibration based on actinometry

      Uncertainty sourceUncertainty/%Reference
      NOx concentration2NOx instrument (Thermo 17i)
      Cross section of N2O2Edwards G D, et al. 2003[20]
      Quantum yield of N2O< 1Edwards G D, et al. 2003[20]
      Flowrate< 1MKS MFC
      Reaction rate~10Atkinson R, et al. 2004[22]
      Overall N2O calibration method< 16
      O3 concentration< 2O3 instrument (Thermo 49i)
      Cross section of O2~60Creasey D J, et al. 2000[26]
      Quantum yield of O3Creasey D J, et al. 2000[26]
      Flowrate< 1MKS MFC
      Reaction rateBurkholder J B, et al. 2015[27]
      Over all O3 calibration method> 60
    Tools

    Get Citation

    Copy Citation Text

    Dongsen YANG, Wanying LI, Jiangyao CHEN, Xinran LIU, Jun ZHENG. Establishment of ozone- and nitrous oxide-based actinometry for quantifying vacuum ultra-violate radiation intensities and its application[J]. Journal of Atmospheric and Environmental Optics, 2024, 19(2): 175