Acta Optica Sinica, Volume. 40, Issue 15, 1506003(2020)

Calibration for Acoustic Pressure Sensitivity of Optical Fiber Hydrophone Based on Fizeau Interference

Yandong Pang1,2, Ciming Zhou1, Junbin Huang4, Hongcan Gu4, Xi Chen1,2, Chenguang Zhao1,3, Qing Zhou1,2, and Dian Fan1、*
Author Affiliations
  • 1National Engineering Laboratory of Fiber Optic Sensor Technology, Wuhan University of Technology, Wuhan, Hubei 430070, China
  • 2School of Information Engineering, Wuhan University of Technology, Wuhan, Hubei 430070, China
  • 3School of Mechanical and Electronic Engineering, Wuhan University of Technology, Wuhan, Hubei 430070, China
  • 4Department of Weaponry Engineering, Naval University of Engineering, Wuhan, Hubei 430032, China
  • show less
    Figures & Tables(12)
    Structure of Fizeau interferometric optical fiber hydrophone system
    Acoustic pressure field calibration model of hydrophone with different postures. (a) Vertical posture; (b) horizontal posture
    Flow chart of optical fiber hydrophone phase change demodulation algorithm
    Acoustic pressure sensitivity response under different radiuses
    Acoustic pressure sensitivity errors under different postures of hydrophone. (a) Horizontal posture; (b) vertical posture
    Experimental schematic of Fizeau interferometric optical fiber hydrophone
    Fixed design of optical fiber hydrophone
    Output time-domain waveforms under different postures. (a) Vertical posture; (b) horizontal posture
    Acoustic pressure sensitivity of hydrophone under different postures
    Sensitivity difference at different postures
    Sensitivity calibration model with depth reading difference
    • Table 1. [in Chinese]

      View table

      Table 1. [in Chinese]

      Frequency /Hzh=2 cmh=6 cmh=15 cm
      Actual sensitivityMeasured sensitivityErrorActual sensitivityMeasured sensitivityErrorActual sensitivityMeasured sensitivityError
      100-144.138-147.9023.764-160.215-161.6241.409-164.029-164.490.461
      1000-138.433-142.2533.820-143.361-144.5191.158-164.724-165.2170.493
      2000-160.541-163.8663.325-174.309-175.4371.128-174.716-174.7570.041
    Tools

    Get Citation

    Copy Citation Text

    Yandong Pang, Ciming Zhou, Junbin Huang, Hongcan Gu, Xi Chen, Chenguang Zhao, Qing Zhou, Dian Fan. Calibration for Acoustic Pressure Sensitivity of Optical Fiber Hydrophone Based on Fizeau Interference[J]. Acta Optica Sinica, 2020, 40(15): 1506003

    Download Citation

    EndNote(RIS)BibTexPlain Text
    Save article for my favorites
    Paper Information

    Category: Fiber Optics and Optical Communications

    Received: Dec. 19, 2019

    Accepted: Apr. 30, 2020

    Published Online: Aug. 14, 2020

    The Author Email: Fan Dian (fandian@whut.edu.cn)

    DOI:10.3788/AOS202040.1506003

    Topics