Journal of Applied Optics, Volume. 46, Issue 3, 670(2025)

Calibration system for solar spectrograph radiance response

Jianwen ZHANG1, Weijie DENG1, Bo LI2、*, Heng WU2, and Guanyu LIN2
Author Affiliations
  • 1Key Laboratory of Optical System Advanced Manufacturing, Changchun Institute of Optics, Fine Mechanics and Physics, Chinese Academy of Sciences, Changchun 130033, China
  • 2The 1st Department of Space Optics, Changchun Institute of Optics, Fine Mechanics and Physics, Chinese Academy of Sciences, Changchun 130033, China
  • show less
    Figures & Tables(7)
    Partial band spectral response function obtained by direct measurement
    Schematic diagram of composition of solar radiance hyperspectral responsivity calibration system
    Transmittance curve of partial bands
    Relation between standard radiance and wavelength
    Partial band spectral response function
    • Table 1. Deviation comparison of irradiance calibration

      View table
      View in Article

      Table 1. Deviation comparison of irradiance calibration

      TypeDifference value
      Maximum irradiance0.175 8 W/mm2
      Average irradiance0.135 2 W/mm2
      Standard deviation−0.093 0
      Average deviation−0.009 5
      Distribution position and extension widthX: −0.036 0 μm σ: 0.032 4 mm Y: −0.012 1 μm σ: 0.032 3 mm
    • Table 2. Uncertainty calculation

      View table
      View in Article

      Table 2. Uncertainty calculation

      Component of uncertainty(k=2)Contributions/%
      Uncertainty of radiance of integrating sphereDiffuse reflection plate1.50
      Transmit standard radiometer error1.12
      Geometric error0.50
      Integral sphere error2.58
      Transfer light error1.60
      The instrument uncertaintyInstrumental repeatability1.00
      Wavelength repeatability and accuracy0.80
      Stray light1.00
      Non-linearity1.50
      Filter measurement uncertaintyAccuracy of photometer measurement1.50
      Accuracy of filter wavelength1.00
      Expanded uncertainty(k=2)4.60
    Tools

    Get Citation

    Copy Citation Text

    Jianwen ZHANG, Weijie DENG, Bo LI, Heng WU, Guanyu LIN. Calibration system for solar spectrograph radiance response[J]. Journal of Applied Optics, 2025, 46(3): 670

    Download Citation

    EndNote(RIS)BibTexPlain Text
    Save article for my favorites
    Paper Information

    Category:

    Received: Dec. 23, 2024

    Accepted: --

    Published Online: May. 28, 2025

    The Author Email: Bo LI (李博)

    DOI:10.5768/JAO202546.0303002

    Topics