Acta Optica Sinica, Volume. 37, Issue 9, 0922001(2017)

Optical Design of Portable Non-Mydriatic Fundus Camera with Large Field of View

Xiaoheng Wang1,2 and Qingsheng Xue1、*
Author Affiliations
  • 1 Changchun Institute of Optics, Fine Mechanics and Physics, Chinese Academy of Sciences, Changchun, Jilin 130033, China
  • 2 University of Chinese Academy of Sciences, Beijing 100049, China
  • show less
    Figures & Tables(16)
    Optical system structure of the fundus camera
    Color image of the retina
    Eye model
    Structure of the ophthalmic lens. (a) Prototype; (b) optimization result
    Initial structure of the imaging system. (a) Without spectroscope; (b) with spectroscope
    Structure of the imaging system combined with eye model
    MTF curves of the imaging system
    Aberration analysis diagram. (a) Ray fan plot; (b) spot diagram; (c) field curvature and distortion
    Imaging simulation of the system. (a) Source image; (b) simulated image plane
    Schematic of the annular illumination
    Lighting system construction
    Schematic of the ring light source
    (a) MTF, (b) spot diagram and (c) relative illumination distribution of the illumination system
    • Table 1. Main designed specifications of the fundus camera

      View table

      Table 1. Main designed specifications of the fundus camera

      ParameterValue
      Field of view /(°)40
      Smallest pupil diameter that can be detected /mm3.5
      Diopter compensation /D-10~10
      Size of the receiver /mm10
      Wavelength of focus /nmNear-infrared light: 870
      Wavelength of illuminate /nmVisible light: 486~656
    • Table 2. Tolerance range of the optical system

      View table

      Table 2. Tolerance range of the optical system

      ParameterValue
      RadiusN=2
      Thickness /mm0.02
      Decenter X/Y /mm0.02
      Tilt X/Y /(″)30
      S+A irregularityΔN=0.2
      Index0.0001
      Abbe0.0083
    • Table 3. MTF of the imaging system at 70 lp/mm

      View table

      Table 3. MTF of the imaging system at 70 lp/mm

      ParameterValue
      Nominal0.489
      Best0.474
      Worst0.359
      Mean0.436
      Standard deviation0.032
    Tools

    Get Citation

    Copy Citation Text

    Xiaoheng Wang, Qingsheng Xue. Optical Design of Portable Non-Mydriatic Fundus Camera with Large Field of View[J]. Acta Optica Sinica, 2017, 37(9): 0922001

    Download Citation

    EndNote(RIS)BibTexPlain Text
    Save article for my favorites
    Paper Information

    Category: Optical Design and Fabrication

    Received: Apr. 18, 2017

    Accepted: --

    Published Online: Sep. 7, 2018

    The Author Email: Xue Qingsheng (qshxue2006@163.com)

    DOI:10.3788/AOS201737.0922001

    Topics