Acta Optica Sinica, Volume. 40, Issue 22, 2212005(2020)

Quantum Vacuum Measurement Based on Fabry-Perot Laser Resonant Cavity

Wenjie Jia, Zhenhua Xi, Dong Fan, Meng Dong, Chengyao Wu, and Yongjun Cheng*
Author Affiliations
  • Research Center of Vacuum Measurement Technology, Lanzhou Institute of Physics, Lanzhou, Gansu 730000, China
  • show less
    Figures & Tables(6)
    Structural diagram of quantum vacuum measurement device based on Fabry-Perot resonant cavity
    Test results of laser frequency uncertainty. (a) Variation of beat frequency with time; (b) evaluation result of Allan deviation
    Relationship between measured and retrieved vacuum degrees
    • Table 1. Partial test result of laser frequency

      View table

      Table 1. Partial test result of laser frequency

      SerialnumberTime /sΔν /MHzν /MHz
      100384350383.4336
      240000.0007384350383.4329
      380000.0010384350383.4326
      4120000.0017384350383.4319
      5160000.0021384350383.4315
      6200000.0025384350383.4311
      7240000.0030384350383.4306
      8280000.0034384350383.4301
      9320000.0039384350383.4297
      10360000.0044384350383.4292
    • Table 2. Measurement results of refractive index

      View table

      Table 2. Measurement results of refractive index

      Serial numberVacuum degree /Pan-1Serial numberVacuum degree /Pan-1
      11112.73023×10-712100512.51911×10-5
      26201.54968×10-613202305.07167×10-5
      310342.58229×10-614300897.54434×10-5
      422075.52102×10-615402551.00935×10-4
      531697.93029×10-616503041.26143×10-4
      639981.00120×10-517604981.51719×10-4
      750401.26219×10-518700381.75653×10-4
      860671.51961×10-519801922.01135×10-4
      970381.76357×10-520901042.26002×10-4
      1080442.01531×10-5211001862.51302×10-4
      1190462.26681×10-5
    • Table 3. Measured and retrieved results of vacuum degree

      View table

      Table 3. Measured and retrieved results of vacuum degree

      Measured value /PaBeat frequency /MHzDeviation of modulusRetrieved value /PaRelative deviation
      10056-595.357100523.90×10-4
      20235227.8613202332.40×10-5
      30094744.0019300849.60×10-5
      402601554.0725402481.21×10-4
      50309757.3332502981.06×10-4
      605031599.6738604967.50×10-5
      700431811.0044700385.10×10-5
      801971119.605180198-7.00×10-6
      901091689.905790113-3.50×10-5
      100191928.3064100199-8.40×10-5
    Tools

    Get Citation

    Copy Citation Text

    Wenjie Jia, Zhenhua Xi, Dong Fan, Meng Dong, Chengyao Wu, Yongjun Cheng. Quantum Vacuum Measurement Based on Fabry-Perot Laser Resonant Cavity[J]. Acta Optica Sinica, 2020, 40(22): 2212005

    Download Citation

    EndNote(RIS)BibTexPlain Text
    Save article for my favorites
    Paper Information

    Category: Instrumentation, Measurement and Metrology

    Received: May. 29, 2020

    Accepted: Aug. 3, 2020

    Published Online: Oct. 25, 2020

    The Author Email: Cheng Yongjun (ouc2016opjwj@sina.com)

    DOI:10.3788/AOS202040.2212005

    Topics