Chinese Journal of Lasers, Volume. 50, Issue 9, 0907208(2023)

Integrated Design and Implementation of Ocular Diopter and Biological Parameters Measurement

Cheng Wang1, Jun Zhu1, Lineng Xiong1, Jianrong Dong1, Lei Dai1, Huazhong Xiang1, Minghui Chen1, Gang Zheng1, Jie Zhao2、*, and Dawei Zhang3、**
Author Affiliations
  • 1Institute of Biomedical Optics and Optometry, Key Laboratory of Medical Optical Technology and Instruments, Ministry of Education, School of Health Science and Engineering, University of Shanghai for Science and Technology, Shanghai 200093, China
  • 2Department of Ophthalmology, Yangpu District Shidong Hospital, Shanghai 200438, China
  • 3Engineering Research Center of Optical Instruments and Systems, Ministry of Education, Key Laboratory of Modern Optical Systems, University of Shanghai for Science and Technology, Shanghai 200093, China
  • show less
    Figures & Tables(9)
    Optical path diagram with integrated design of eye refraction and biological parameter measurement
    Image formed by reflection of cornea and lens and distance x measurement
    Eye axis length measurement
    Corneal curvature measurement
    Anterior chamber depth measurement
    Diopter measurement image
    • Table 1. Analysis of diopter measurement of standard model eye

      View table

      Table 1. Analysis of diopter measurement of standard model eye

      Diopter of standard model eye /DMean±standard deviationt valueP value
      -20-19.93±0.170.8750.431
      -15-14.97±0.160.3990.710
      -10-10.07±0.09-1.7230.160
      -5-5.05±0.07-1.4140.230
      00.09±0.082.4990.070
      54.95±0.07-1.4140.230
      1010.02±0.130.3430.749
      1514.93±0.12-1.3000.263
      2019.95±0.45-0.2470.817
    • Table 2. Analysis of measured human eye biological parameters and diopter

      View table

      Table 2. Analysis of measured human eye biological parameters and diopter

      ParameterSwTRTCoV /%ICC(95%CI)
      AL /mm0.0150.0420.0600.999
      K0.0200.0550.2560.994
      ACD /mm0.0990.2762.9710.968
      Spherical degree /D0.0430.1203.6560.894
      Cylindrica degree/D0.0670.1877.1440.680
    • Table 3. Analysis of parameter measurement results of IOL Master and experimental prototype

      View table

      Table 3. Analysis of parameter measurement results of IOL Master and experimental prototype

      ParameterMean±standard deviation

      t

      value

      P

      value

      PrototypeIOL Master 500
      AL /mm25.29±1.2325.36±1.09-0.5320.617
      K7.74±0.327.75±0.31-0.2610.807
      ACD /mm3.35±0.283.58±0.31-1.8340.141
    Tools

    Get Citation

    Copy Citation Text

    Cheng Wang, Jun Zhu, Lineng Xiong, Jianrong Dong, Lei Dai, Huazhong Xiang, Minghui Chen, Gang Zheng, Jie Zhao, Dawei Zhang. Integrated Design and Implementation of Ocular Diopter and Biological Parameters Measurement[J]. Chinese Journal of Lasers, 2023, 50(9): 0907208

    Download Citation

    EndNote(RIS)BibTexPlain Text
    Save article for my favorites
    Paper Information

    Category: Optical Diagnostics and Therapy

    Received: Nov. 28, 2022

    Accepted: Feb. 22, 2023

    Published Online: Apr. 24, 2023

    The Author Email: Zhao Jie (13761231212@163.com), Zhang Dawei (usstoe@163.com)

    DOI:10.3788/CJL221460

    Topics