Laser & Optoelectronics Progress, Volume. 55, Issue 2, 021401(2018)

Effect of Sample Temperature on Characteristic Parameters of Double-Pulse Laser Induced Al Plasma

Xiaomei Lin, Lei Zhong, and Jingjun Lin*
Author Affiliations
  • College of Electrical & Electronic Engineering, Changchun University of Technology, Changchun, Jilin 130000, China
  • show less
    Figures & Tables(14)
    Experimental setup diagram of LIBS
    Contrast of Al spectral line intensity induced by SP-LIBS and DP-LIBS
    Effect of Ar flow on spectral line intensity
    Al spectra collected by DP-LIBS technology at different sample temperatures
    Effect of sample temperature on intensity of Al spectral lines induced by DP-LIBS technology
    Boltzmann oblique line of Al Ⅰ atom induced by double pulse laser at room temperature (13 ℃)
    Effect of sample temperature on electron temperature of Al plasma induced by double pulse laser
    Fitting spectral lines of different linetypes of Al atom at 308.22 nm
    Effect of sample temperature on electron density of Al plasma induced by double pulse laser
    Effect of sample temperature on FWHM of different Al spectral lines induced by double pulse laser
    Effect of sample temperature on morphology and size of plasma plume
    • Table 1. Al atom spectral lines used in Boltzmann oblique line

      View table

      Table 1. Al atom spectral lines used in Boltzmann oblique line

      AtomWavelength /nmEk /cm-1gkAki /(107 s-1)
      Al I221.0145344.16540.87
      226.9144168.84767.58
      237.2142144.41120.58
      256.8038929.41341.92
      257.5138933.96863.60
      265.2537689.40721.42
      266.0437689.40722.84
      308.2232435.45345.87
      309.2732436.79667.29
    • Table 2. Comparison of different linear fitting for Al atom spectral lines at 308.22 nm

      View table

      Table 2. Comparison of different linear fitting for Al atom spectral lines at 308.22 nm

      FittinglinetypeFWHMR-SquareAICAkaikeweight
      Gauss0.160990.98948220.704580.39956
      Lorentz0.144460.99040218.869761
      Voigt0.129890.91065266.368074.85146×10-11
    • Table 3. Original morphology area of plasma plume at different sample temperatures

      View table

      Table 3. Original morphology area of plasma plume at different sample temperatures

      Sample temperature /℃13160260360460560660700750
      S /(pixel·inch-1)477453935721593961916512696369966981
    Tools

    Get Citation

    Copy Citation Text

    Xiaomei Lin, Lei Zhong, Jingjun Lin. Effect of Sample Temperature on Characteristic Parameters of Double-Pulse Laser Induced Al Plasma[J]. Laser & Optoelectronics Progress, 2018, 55(2): 021401

    Download Citation

    EndNote(RIS)BibTexPlain Text
    Save article for my favorites
    Paper Information

    Category: Lasers and Laser Optics

    Received: Jul. 10, 2017

    Accepted: --

    Published Online: Sep. 10, 2018

    The Author Email: Jingjun Lin ( 1570498510@qq.com)

    DOI:10.3788/LOP55.021401

    Topics