Journal of Applied Optics, Volume. 43, Issue 6, 1107(2022)

Graphical offline analysis software of electron multiplier pulse signals

Huaxing PENG1,2, Baojun YAN1、*, Shulin LIU1,3、**, Binting ZHANG1,2, and Wenlu WEI1,2
Author Affiliations
  • 1State Key Laboratory of Particle Detection and Electronics, Institute of High Energy Physics, Chinese Academy of Sciences, Beijing 100049, China
  • 2School of Physical Sciences, University of Chinese Academy of Sciences, Beijing 100049, China
  • 3School of Nuclear Science and Technology, University of Chinese Academy of Sciences, Beijing 100049, China
  • show less
    Figures & Tables(17)
    Screenshot of program interface
    Flow chart of program
    Diagram of pulse signal waveform and integration interval selection
    Charge integral spectrum
    Charge integral spectrum and fitting analysis
    Post-pulse test and analysis
    Leading-edge time analysis and goodness of fitting
    Physical drawing and schematic diagram of testing device
    Comparison of measurement results for PMT by oscilloscope and QDC
    Post-pulse time and charge distribution
    Dark noise and post-pulse waveform
    Relationship between post-pulse rate and threshold
    Leading-edge time distribution
    Baseline deduction discussion
    • Table 1. Parameters for signal generator channel 1 and channel 2

      View table
      View in Article

      Table 1. Parameters for signal generator channel 1 and channel 2

      通道频率/ kHz幅值/ V延时/ ns脉冲宽度/ ns
      113.174050
      21−0.8255300
    • Table 2. Measurement results of oscilloscope

      View table
      View in Article

      Table 2. Measurement results of oscilloscope

      电压/Vq1σ1μqreal /C增益分辨率
      12501.992730.7558920.2369463.98546E-132490912.50.379324846
      13002.53470.9241340.2292745.0694E-1331683750.364593048
      13503.003321.071090.2443166.00664E-1337541500.356635324
      14003.51851.165210.2228927.037E-1343981250.33116669
      14504.359951.347280.2091328.7199E-135449937.50.309012718
      15005.657531.754770.2583431.13151E-127071912.50.31016539
      15506.119771.736340.187441.22395E-127649712.50.283726349
    • Table 3. Measurement results of QDC

      View table
      View in Article

      Table 3. Measurement results of QDC

      电压/Vq1σ1μqreal /C增益分辨率
      125012.57954.111520.1989413.9399E-132462437.1250.326842879
      130015.52504.770350.1881264.86243E-133039018.7500.307268921
      135019.23115.691740.1866186.02318E-133764487.8250.295965389
      140022.79966.716700.1815187.14083E-134463021.7000.294597274
      145028.43698.406800.2021618.90644E-135566523.1750.295629974
      150033.81029.410060.1870011.05894E-126618346.6500.278320152
      155040.676811.364500.1951361.27400E-127962483.6000.279385301
    Tools

    Get Citation

    Copy Citation Text

    Huaxing PENG, Baojun YAN, Shulin LIU, Binting ZHANG, Wenlu WEI. Graphical offline analysis software of electron multiplier pulse signals[J]. Journal of Applied Optics, 2022, 43(6): 1107

    Download Citation

    EndNote(RIS)BibTexPlain Text
    Save article for my favorites
    Paper Information

    Category: Research Articles

    Received: Jul. 18, 2022

    Accepted: --

    Published Online: Nov. 18, 2022

    The Author Email:

    DOI:10.5768/JAO202243.0604009

    Topics