Acta Optica Sinica, Volume. 37, Issue 8, 0811003(2017)

An Iterative All-Geometric-Parameter Calibration Method for Cone-Beam Computed Laminography System

Jingyu Wang, Yu Han, Lei Li, Xiaoqi Xi, Jianbang Liu, and Bin Yan*
Author Affiliations
  • College of Information System Engineering, Information Engineering University, Zhengzhou, Henan 450002, China
  • show less
    Figures & Tables(10)
    Scanning configuration of the cone beam CL system
    Projections of phantom
    Convergence curve of F
    Final relative error
    Reconstruction results of Shepp-Logan phantom. (a) Images of XY slice; (b) images of XZ slice; (c) images of YZ slice
    Reconstruction results of the PCB phantom. (a) Ideal image; (b) image corrected by Zhang's method; (c) image corrected by the proposed method
    • Table 1. Parameter calibration results with noise level of 0.05

      View table

      Table 1. Parameter calibration results with noise level of 0.05

      ParameterZhang's method in CT systemZhang's method in CL system
      Real valueMeasured valueReal valueMeasured value
      η /(°)0.0350.0340.0290.028
      pX /mm4.5874.59539.86139.657
      pZ /mm0.6640.8671212.9321205.948
      sX /mm55.03954.836
      sY /mm14501450.02514501445.835
      sZ /mm-2-0.456-23.771
      t7.2507.25212.24212.209
    • Table 2. Parameter calibration results with noise level of 0.1

      View table

      Table 2. Parameter calibration results with noise level of 0.1

      ParameterZhang's method in CT systemZhang's method in CL system
      Real valueMeasured valueReal valueMeasured value
      η /(°)0.0350.0340.0290.028
      pX /mm4.5874.57339.86139.649
      pZ /mm0.6640.3601212.9321206.385
      sX /mm55.04955.529
      sY /mm14501449.78114501447.381
      sZ /mm-2-1.362-25.069
      t7.2507.25212.24212.201
    • Table 3. Parameter calibration results with noise level of 0.05

      View table

      Table 3. Parameter calibration results with noise level of 0.05

      ParameterReal valueZhang's methodProposed method
      η /(°)0.0290.0280.028
      pX /mm39.86139.65739.742
      pZ /mm1212.9321205.9481212.919
      sX /mm54.8365.055
      sY /mm14501445.8351451.483
      sZ /mm-23.771-0.261
      t12.24212.20912.231
    • Table 4. Parameter calibration results with noise level of 0.1

      View table

      Table 4. Parameter calibration results with noise level of 0.1

      ParameterReal valueZhang's methodProposed method
      η /(°)0.0290.0280.028
      pX /mm39.86139.64939.673
      pZ /mm1212.9321206.3851212.971
      sX /mm55.5295.089
      sY /mm14501447.3811450.921
      sZ /mm-25.069-0.929
      t12.24212.20112.225
    Tools

    Get Citation

    Copy Citation Text

    Jingyu Wang, Yu Han, Lei Li, Xiaoqi Xi, Jianbang Liu, Bin Yan. An Iterative All-Geometric-Parameter Calibration Method for Cone-Beam Computed Laminography System[J]. Acta Optica Sinica, 2017, 37(8): 0811003

    Download Citation

    EndNote(RIS)BibTexPlain Text
    Save article for my favorites
    Paper Information

    Category: Imaging Systems

    Received: Mar. 10, 2017

    Accepted: --

    Published Online: Sep. 7, 2018

    The Author Email: Yan Bin (ybspace@hotmail.com)

    DOI:10.3788/AOS201737.0811003

    Topics