Journal of Applied Optics, Volume. 45, Issue 3, 637(2024)

Application of optical angle gauge in lens center deviation measuring instrument calibration

Ying HAN*, Xinwei LI, Yongchun FAN, Lei YI, and Li LIU
Author Affiliations
  • China Airborne Missile Academy, Luoyang 471000, China
  • show less
    Figures & Tables(12)
    Schematic diagram of optical angle gauge[11]
    Schematic diagram of angle of deviation
    Upper and lower surface reflections
    Diagram of optical path design
    Adjustment of cross image
    Test method of lower optical path
    Placement diagram of optical angle gauge
    Collected reflections
    • Table 1. Requirements of deflection angle deviation and measurement uncertainty

      View table
      View in Article

      Table 1. Requirements of deflection angle deviation and measurement uncertainty

      等级偏向角偏向角偏差/(″)测量不确定度(k=2)/(″)
      一等2″~10″±10.1
      20″~1′±2
      2′~10′±3
      二等2″~10″±1.50.2
      20″~1′±3
      2′~10′±4
      三等2″~10″±20.5
      20″~1′±4
      2′~10′±5
    • Table 2. Reference data for nominal values of N-BK7 optical angle gauge

      View table
      View in Article

      Table 2. Reference data for nominal values of N-BK7 optical angle gauge

      波长/nm
      850660632.8546.1530518
      折射率1.509 841.514 211.515 091.518 721.519 581.520 28
      2′117.920118.93119.134119.974120.174120.334
      5′297.244299.79300.304302.422302.925303.330
    • Table 3. Measurement results of single optical path device

      View table
      View in Article

      Table 3. Measurement results of single optical path device

      参数
      光学角规标称值/(′)25
      光学角规名义值/(″)120.032302.925
      实测值/(″)231.338 6582.806 8
      231.261 6582.872 6
      231.257 6582.840 1
      231.301 6582.861 4
      231.310 2582.856 3
      231.332 8582.830 1
      231.258 3582.851 8
      231.292 6582.843 9
      231.260 6582.888 9
      231.280 4582.811 4
      平均值/(″)231.289 4582.846 3
      偏向角/(″)120.17302.84
      示值误差/(″)0.14−0.09
      重复性/(″)0.0310.026
    • Table 4. Measurement results of double optical path device

      View table
      View in Article

      Table 4. Measurement results of double optical path device

      参数光路方向
      上光路下光路
      光学角规标称值/(′)2525
      光学角规名义值/(″)120.032302.925120.032302.925
      实测值/(″)231.41582.63231.93582.41
      231.38582.56231.70582.44
      231.34582.60231.83582.48
      231.37582.58231.68582.41
      231.32582.51231.79582.38
      231.29582.57231.83582.37
      231.33582.63231.92582.43
      231.39582.56231.75582.48
      231.36582.64231.8582.42
      231.29582.53231.66582.37
      平均值/(″)231.35582.58231.79582.42
      偏向角/(″)120.00302.20120.24302.11
      示值误差/(″)−0.03−0.730.20−0.81
      重复性/(″)0.0410.0440.0930.040
    Tools

    Get Citation

    Copy Citation Text

    Ying HAN, Xinwei LI, Yongchun FAN, Lei YI, Li LIU. Application of optical angle gauge in lens center deviation measuring instrument calibration[J]. Journal of Applied Optics, 2024, 45(3): 637

    Download Citation

    EndNote(RIS)BibTexPlain Text
    Save article for my favorites
    Paper Information

    Category: Research Articles

    Received: Feb. 3, 2023

    Accepted: --

    Published Online: Jun. 2, 2024

    The Author Email: Ying HAN (hyloveconan@163.com)

    DOI:10.5768/JAO202445.0303001

    Topics