Laser & Optoelectronics Progress, Volume. 62, Issue 6, 0611003(2025)

Design of Microscopy Imaging System Based on ZYNQ

Cunlong Su1,2、*, Jing Liu3, Runna Liu3, Shuwei Yang1, Keding Yan2, Siyi He2, and Yang Zhou2
Author Affiliations
  • 1School of Intelligent Science and Engineering, Xi'an Peihua University, Xi'an 710125, Shaanxi , China
  • 2School of Electronic and Information Engineering, Xi'an Technological University, Xi'an 710021, Shaanxi , China
  • 3School of Medicine, Xi'an International University, Xi'an 710077, Shaanxi , China
  • show less
    Figures & Tables(11)
    Component block diagram of small microimaging system
    Hardware composition diagram of small microscopic imaging system. (a) Imaging optical path diagram; (b) mechanical structure design diagram; (c) LED spotlight; (d) homogeniser; (e) physical diagram of achromatic objective lens; (f) manual slide; (g) physical diagram of OV5640 camera; (h) ZYNQ main control
    Block diagram of acquisition system architecture
    Top-level design of image acquisition system
    Flowchart of image acquisition system software
    Software design diagram for small microscopic imaging system. (a) Overall flow chart of microscopic image acquisition software; (b) graphical user interface of microscopic image acquisition software
    System hardware and software joint diagram
    System resolution test. (a) Physical image of USAF-1951 resolution board; (b) image of resolution board captured under 10× objective lens
    Bright-field microscopic images captured under 10× objective lens. (a) Human blood cell; (b) longitudinal section of myocardium; (c) transverse section of spinal cord; (d) transverse section of Cycas ferruginea leaf; (e) slice of fish skin; (f) corn starch
    • Table 1. Parameter of flat field achromatic objective

      View table

      Table 1. Parameter of flat field achromatic objective

      Magnifying powerNumerical apertureWorking distance /mmOptical resolution /nmIdeal pixel size /μm
      10×0.252.04976‒18544.88‒9.27
      20×0.400.65610‒11596.10‒11.59
      40×0.650.56375‒7137.50‒14.26
    • Table 2. USAF-1951 resolution test table

      View table

      Table 2. USAF-1951 resolution test table

      Element01234567
      1500.00250.00125.0062.5031.2515.637.813.91
      2445.45222.72111.3655.6827.8413.926.963.48
      3396.85198.4399.2149.6124.8012.406.203.10
      4353.55176.7888.3944.1922.1011.055.522.76
      5314.98157.4978.7539.3719.699.844.922.46
      6280.62140.3170.1535.0817.548.774.382.19
    Tools

    Get Citation

    Copy Citation Text

    Cunlong Su, Jing Liu, Runna Liu, Shuwei Yang, Keding Yan, Siyi He, Yang Zhou. Design of Microscopy Imaging System Based on ZYNQ[J]. Laser & Optoelectronics Progress, 2025, 62(6): 0611003

    Download Citation

    EndNote(RIS)BibTexPlain Text
    Save article for my favorites
    Paper Information

    Category: Imaging Systems

    Received: Jul. 8, 2024

    Accepted: Aug. 28, 2024

    Published Online: Mar. 6, 2025

    The Author Email: Cunlong Su (1921014513@qq.com)

    DOI:10.3788/LOP241644

    CSTR:32186.14.LOP241644

    Topics