Laser & Optoelectronics Progress, Volume. 60, Issue 17, 1722001(2023)

A Straight-Axis Multi-Channel Optical Detection System for Nucleic Acid On-Site Detection

Bangchao Xi1,2,3,4, Jiayu Yang1,2,3,4, Shaolei Huang1,2,3,4, Haozheng Dai1,2,3,4, Juntian Zeng1,2,3,4, and Dongxu Zhang1,2,3,4、*
Author Affiliations
  • 1School of Public Health, Xiamen University, Xiamen 361102, Fujian , China
  • 2State Key Laboratory of Molecular Vaccinology and Molecular Diagnostic(Xiamen University), Xiamen 361102, Fujian , China
  • 3National Institute of Diagnostics and Vaccine Development in Infection Diseases(Xiamen University), Xiamen 361102, Fujian , China
  • 4State Drug Administration Key Laboratory of Infectious Disease Detection Technology Research and Evaluation, Xiamen 361102, Fujian , China
  • show less
    Figures & Tables(13)
    Schematic diagram of the orthogonal optical path structure
    Schematic diagram of light passing through optical components
    Schematic diagram of multi-channel integration and switching device
    Rapid amplification supporting device. (a) Microfluidic chip; (b) amplification heating light path; (c) optical fiber with linear spot; (d) physical map of multi-channel integration and switching device
    Block diagram of the hardware structure of the detection system
    Light Tools simulation. (a) Optical path structure; (b) spot size
    Test environment and conditions. (a) Fully automatic nucleic acid detection device; (b) internal structure diagram
    Resolution comparison between the straight-axis multi-channel optical detection systems and commercial PCR instruments.(a) FAM; (b) HEX; (c) ROX
    Channel crosstalk data histogram
    Detection curves of straight-axis multi-channel optical detection systems. (a) Random result 1; (b) random result 2; (c) random result 3; (d) random result 4
    • Table 1. Parameters of LED

      View table

      Table 1. Parameters of LED

      LEDCenter wavelength /nmHalf bandwidth /nmHalf-width value /(°)CompanyFluorescence channel
      NSPB510BS46520±15NICHIAFAM
      NSPG310B52035±15NICHIAHEX
      HSE575H-L51257535±12Huashang OptoelectronicsROX
      NSPR310S63515±15NICHIACY5
    • Table 2. Parameters of optical path distance

      View table

      Table 2. Parameters of optical path distance

      Fluorescence channelDistance between LED and excitation lens /mmDistance between excitation lens and filter/mmDistance between fiber and emission lens /mmDistance between emission lens and filter /mm
      FAM7.77.37.64.3
      HEX7.77.37.64.3
      ROX7.17.37.64.3
      CY57.77.37.64.3
    • Table 3. Fluorescent dye detection repeatability test results

      View table

      Table 3. Fluorescent dye detection repeatability test results

      DyeFAMHEXROX
      ConcentrationHighModerateLowHighModerateLowHighModerateLow
      CV /%0.020.100.720.100.180.830.170.540.80
    Tools

    Get Citation

    Copy Citation Text

    Bangchao Xi, Jiayu Yang, Shaolei Huang, Haozheng Dai, Juntian Zeng, Dongxu Zhang. A Straight-Axis Multi-Channel Optical Detection System for Nucleic Acid On-Site Detection[J]. Laser & Optoelectronics Progress, 2023, 60(17): 1722001

    Download Citation

    EndNote(RIS)BibTexPlain Text
    Save article for my favorites
    Paper Information

    Category: Optical Design and Fabrication

    Received: Aug. 3, 2022

    Accepted: Sep. 15, 2022

    Published Online: Sep. 13, 2023

    The Author Email: Dongxu Zhang (zhangdongxu@xmu.edu.cn)

    DOI:10.3788/LOP222231

    Topics