Acta Optica Sinica, Volume. 42, Issue 6, 0611001(2022)

Design of Medium/Short Wave Infrared Radiation Calibration System Based on Imaging/Light Intensity Change Response

Sen Yang and Houqing Zhang*
Author Affiliations
  • School of Mechanical and Electrical Engineering, Northeast Forestry University, Harbin, Heilongjiang 150040, China
  • show less
    Figures & Tables(12)
    Radiation measurement system based on light intensity change response
    Radiation measurement system based on imaging/light intensity change response
    Principle block diagram of system
    Calibration diagram of system
    Process of multi-parameter calibration method
    Calibration curve. (a) 2.05--2.55 μm; (b) 3.70--4.80 μm
    • Table 1. Technical index of system

      View table

      Table 1. Technical index of system

      IndexValue
      Measurement uncertainty /%≤3
      Waveband /μm2.05--2.55, 3.70--4.80
      Irradiance /(W·cm-2)6×10-10-7×10-4
      Field angle /(°)±0.6
      Signal to noise ratio≥100
    • Table 2. Uncertainty of each component of system

      View table

      Table 2. Uncertainty of each component of system

      ComponentRelative standard uncertainty /%
      Temperature of blackbody1.00
      Emissivity of blackbody0.30
      Aperture of blackbody1.00
      Transmittance of collimator0.42
      Voltage measurement accuracy of system0.74
      Calibration reproducibility of system0.85
      Combined standard uncertainty1.88
    • Table 3. Parameter setting of simulator

      View table

      Table 3. Parameter setting of simulator

      Radiation sourceParameterValue
      Resistor arrayGray scale256
      Waveband /μm3.70--4.80
      Temperature range /℃1000--1200
      BlackbodyTemperature interval /℃50
      Field stop area /mm20.146
      Waveband /μm2.05--2.55
    • Table 4. Measurement result of irradiance

      View table

      Table 4. Measurement result of irradiance

      Field stop area /mm2Irradiance /(W·cm-2)
      0.0316.105×10-9
      0.1963.934×10-8
      0.7861.589×10-7
      1.5403.169×10-7
      3.1426.454×10-7
    • Table 5. Temperature measurement result of equivalent blackbody

      View table

      Table 5. Temperature measurement result of equivalent blackbody

      Temperature of blackbody /Temperature of equivalent blackbody /℃
      Measured by existing radiometerMeasured by new radiometer
      1000410.0407.8
      1050424.6422.2
      1100438.4436.3
      1150452.1449.6
      1200465.0462.6
    • Table 6. Measurement result of energy ratio

      View table

      Table 6. Measurement result of energy ratio

      Measured objectWavebandEnergy ratio
      Interference AMedium/short wave1∶3.91
      Interference BMedium/short wave1∶4.02
      Interference CMedium/short wave1∶3.91
      Object/interference AMedium wave1∶3.83
      Object/interference BMedium wave1∶4.35
      Object/interference CMedium wave1∶5.41
    Tools

    Get Citation

    Copy Citation Text

    Sen Yang, Houqing Zhang. Design of Medium/Short Wave Infrared Radiation Calibration System Based on Imaging/Light Intensity Change Response[J]. Acta Optica Sinica, 2022, 42(6): 0611001

    Download Citation

    EndNote(RIS)BibTexPlain Text
    Save article for my favorites
    Paper Information

    Category: Imaging Systems

    Received: Aug. 26, 2021

    Accepted: Sep. 27, 2021

    Published Online: Mar. 8, 2022

    The Author Email: Zhang Houqing (1695641968@qq.com)

    DOI:10.3788/AOS202242.0611001

    Topics