High Power Laser and Particle Beams, Volume. 34, Issue 11, 114002(2022)

Calibration method of capacitance sensor for particle accelerator wire position measurement

Luping Yan1, Lan Dong1,2、*, Tong Wang1,2, Shang Lu1, Yuanying Han1, Xiaoyang Liu1, Luyan Zhang1, Haoyue Yan1, Na Ma1,2, Zhenqiang He1,2, Zhiyong Ke1,2, Lingling Men1,2, Bo Li1,2, Xiaolong Wang1,2, Jing Liang1,2, and Tao Luo1,2
Author Affiliations
  • 1Institute of High Energy Physics, Chinese Academy of Sciences, Beijing 100049, China
  • 2Dongguan Neutron Science Center, Dongguan 523803, China
  • show less
    Figures & Tables(14)
    Vibration wire pre-collimation unit measurement system
    Capacitive displacement sensor system
    Working principle diagram and working curve of capacitive displacement sensor
    Differential structure principle and working curve
    Circuit principle of the capacitance sensor
    Input and output of the frequency discriminator
    Wire position calibration system
    34901A multiplexer
    The relation between voltage and position of wire in horizontal and vertical directions
    WPS calibration route
    WPS calibration processing program
    • Table 1. Phase I long term stability test

      View table
      View in Article

      Table 1. Phase I long term stability test

      X/μm Y/μm X/μm Y/μm
      average methodpairwise cycle method
      maximum deviation1.66.23.87.3
      sensor 011minimum deviation−2.5−3.5−2.9−7.6
      standard deviation0.81.71.02.2
      maximum deviation3.65.74.410.7
      sensor 015minimum deviation−3.2−5.8−4.1−7.0
      standard deviation1.51.91.92.9
    • Table 2. Phase II long term stability test

      View table
      View in Article

      Table 2. Phase II long term stability test

      X/μm Y/μm X/μm Y/μm
      average methodpairwise cycle method
      maximum deviation5.66.910.67.4
      sensor 011minimum deviation−2.2−3.5−12.3−6.9
      standard deviation1.01.31.51.9
      maximum deviation1.75.72.27.5
      sensor 015minimum deviation−1.5−2.5−2.0−7.3
      standard deviation0.71.70.92.4
    • Table 3. Phase III long term stability test

      View table
      View in Article

      Table 3. Phase III long term stability test

      X/μm Y/μm X/μm Y/μm
      average methodpairwise cycle method
      maximum deviation5.96.14.56.5
      sensor 011minimum deviation−3.6−4.4−4.7−5.6
      standard deviation1.01.61.01.9
      maximum deviation−2.64.72.96.8
      sensor 015minimum deviation−2.7−3.5−3.5−6.5
      standard deviation0.71.40.91.9
    Tools

    Get Citation

    Copy Citation Text

    Luping Yan, Lan Dong, Tong Wang, Shang Lu, Yuanying Han, Xiaoyang Liu, Luyan Zhang, Haoyue Yan, Na Ma, Zhenqiang He, Zhiyong Ke, Lingling Men, Bo Li, Xiaolong Wang, Jing Liang, Tao Luo. Calibration method of capacitance sensor for particle accelerator wire position measurement[J]. High Power Laser and Particle Beams, 2022, 34(11): 114002

    Download Citation

    EndNote(RIS)BibTexPlain Text
    Save article for my favorites
    Paper Information

    Category: Particle Beams and Accelerator Technology

    Received: Oct. 20, 2021

    Accepted: Jun. 10, 2022

    Published Online: Oct. 18, 2022

    The Author Email: Lan Dong (dongl@ihep.ac.cn)

    DOI:10.11884/HPLPB202234.220447

    Topics