Optically pumped magnetometers (OPMs) operating in the spin-exchange relaxation-free (SERF) regime have been advanced significantly, owing to the demand for miniaturization, multichannel integration, low-cost manufacturing, and high sensitivity[
Chinese Optics Letters, Volume. 20, Issue 8, 081202(2022)
Zero field optically pumped magnetometer with independent dual-mode operation
We propose a dual-mode optically pumped magnetometer (OPM) that can flexibly switch between single-beam modulation mode and double-beam DC mode. Based on a
1. Introduction
Optically pumped magnetometers (OPMs) operating in the spin-exchange relaxation-free (SERF) regime have been advanced significantly, owing to the demand for miniaturization, multichannel integration, low-cost manufacturing, and high sensitivity[
SERF magnetometers have been investigated widely, which can be categorized into the double-beam mode (with pump and probe lights) and the single-beam mode (with only one pump light), based on the structure and detection principle used[
Recently, single-beam mode SERF magnetometers have been applied widely in the field of biomagnetic measurement, owing to their suitable characteristics for miniaturization, array integration, and cost effectiveness. Osborne et al. have reported a zero field OPM for biomedical applications[
Sign up for Chinese Optics Letters TOC Get the latest issue of Advanced Photonics delivered right to you!Sign up now
In this study, we propose a dual-mode SERF magnetometer based on a miniaturized vapor cell that can flexibly switch between the single-beam modulation mode and the double-beam DC mode. The double-beam DC mode is designed with suitable sensitivity to precisely measure the noise floor of a mu-metal magnetic shield, and the single-beam modulation mode exhibits bandwidth characteristics suitable for biomagnetic measurements. Our design is suitable for miniaturized OPMs with multiple functions, including magnetic-shield background noise measurement and medical imaging, which will aid in realizing portable magnetic sensing devices.
2. Principle of SERF Magnetometer
To calculate the spin polarization of alkali metal atoms, the Bloch equation is applied in the SERF regime[
Assuming that the direction of the pump light is along the axis, the direction of the magnetic field to be measured is along the axis. When and are equal to zero, the steady-state components in Eq. (1) are given by[
Equations (4) and (5) describe the combined effects of the optical pumping and magnetic field on the polarization, represented by the spectrometric absorption () and dispersion () curves[
For the single-beam mode, the circularly polarized pump light along the axis is the only beam. When varies, the absorption exhibits a zero field Lorentz resonance, the half-width at half-maximum is , and the maximum value is proportional to . Then, a modulating magnetic field of amplitude and angular frequency is applied along the axis, . The response of in the zero field modulation mode can be described as a Fourier series. The dominant term in the solution is the first harmonic, as given by[
Here, is the Bessel function of the first kind, and is the constant of proportionality.
For the double-beam mode, a beam of linearly polarized probe light is added along the axis on the foundation of the single-beam mode. The rotation angle of the probe light is proportional to . Equation (4) is approximately linear along the axis near , and this component of the magnetic field may be extracted directly from the dispersive lineshape. The response of to the DC mode is given by Eq. (4).
During the operation of the dual-mode SERF magnetometer in the linear region, the term can be ignored when . The polarizability of the axis and axis can be uniformly described by
The polarizability () cannot be obtained directly, but instead the intensity of the laser is measured by photodiodes (PDs) as the response () of the magnetometer. From Eq. (7), we can further simplify the response as . Although the response has a similar form, the corresponding method of operation and scale factor are significantly different. In the balanced polarimetry technique, measuring the optical rotation angle can aid in achieving a higher magnetic field sensitivity, and, for lock-in amplifier demodulation polarization, it can effectively suppress low-frequency noise.
3. Experimental Setup
The experimental setup of the SERF magnetometer is illustrated in Fig. 1. The central part is a small-size alkali-metal vapor cell with external dimensions of and interior dimensions , filled with a droplet of and approximately 650 torr of as the buffer and quenching gas. The vapor cell was heated to approximately 150°C using a 200 kHz AC electronic current. The temperature was monitored and stabilized using a non-magnetic Pt1000 for real-time closed-loop control.
Figure 1.Experimental setup of the SERF magnetometer for independent single-beam mode and double-beam mode operation. PMF, polarization maintaining fiber; OFC, optical fiber collimator; LP, linear polarizer; QWP, quarter-wave plate; PD, photodiode; HWP, half-wave plate; PBS, polarizing-beam splitter; TIA, transimpedance amplifier; LIA, lock-in amplifier; Ref, reference signal; DAQ, data acquisition.
As shown in Fig. 1, when the SERF magnetometer operates in the single-beam mode, the pump light tuned to the Rb D1 resonance line is generated by a distributed feedback (DFB) laser. The pump beam is delivered from the fiber collimator and circularly polarized by a set of linear polarizers and quarter-wave plates. After it is transmitted to the vapor cell, the pump light is detected by PD1 and converted into a voltage signal by an ultra-low-noise PD amplifier (Thorlabs, PDA200C). The lock-in amplifier (Zurich Instruments, MFLI) generates a modulated magnetic field with a frequency of along the sensitive axis and synchronously demodulates the magnetic field information detected by the PD.
When the SERF magnetometer operates in the double-beam mode, the pump light maintains the configuration in the single-beam mode. An additional probe beam is detuned to roughly 100 GHz from the Rb D1 resonance line. The probe beam passes through a fiber collimator and linear polarizer, yielding a linearly polarized beam. After passing through the vapor cell along the axis, the probe beam is geometrically split by a half-wave plate and a polarization beam splitter cube and detected using two PDs (PD2 and PD3). Through the balanced polarimetry technique, the difference between PD2 and PD3 outputs can be used as the response signal of the double-beam mode.
The modulation magnetic field (single-beam mode) and calibration magnetic field were added to the tri-axial coil through a summing amplifier. The output response signal was recorded using a data acquisition system (National Instruments, PXIe-4499). A four-layer cylindrical mu-metal magnetic shield was used to counteract the ambient magnetic field.
4. Results and Discussion
In this section, the SERF magnetometer performance in both modes was optimized separately. The laser frequency detuning and light intensity of the pump and probe lights were optimized first. Under the optimal light parameters, we discuss the optimal operating temperature to identify the optimal sensitivity and bandwidth.
4.1. Performance optimization
In the double-beam mode, the vapor cell is heated to 150°C, and the background magnetic field is compensated to zero by the tri-axial coil. The actual response signal () is determined by the scale factor , which is given by
We first optimized the laser parameters of the probe light. The pump light frequency was maintained at the resonance absorption, and the optical intensity was . Equations (8) and (9) describe the combined effects of the probe light intensity () and frequency () on the scale factor. When the probe light intensity is fixed at , changing the probe frequency detuning from to 220 GHz, the response signal () is detected with calibration magnetic field () of 100 pT rms, and the dependence is shown in Fig. 2(a). The response signal reaches the maximum value when the probe light is blue detuned by about 100 GHz. The experimental data is well fitted with Eq. (9), as shown in Fig. 2(a) with the red solid line. The optimal frequency detuning for maximum signal depends on optical depth and pressure-broadening width. By setting , the maximum scale factor will be obtained when the detuning () reaches the optimal value, which is given by
Figure 2.Response signal versus (a) frequency detuning of probe light, (b) light intensity of probe light, (c) frequency detuning of pump light, and (d) light intensity of pump light.
In Fig. 2(b), the probe light intensity varies from to , and the probe light frequency is maintained at 100 GHz blue detuning. From Eq. (8), the scale factor is positively correlated with probe light intensity. The response signal increased with probe light intensity, and the experimental data is well fitted in Fig. 2(b) as a red solid line. However, when the light intensity increases above , the corresponding probe light intensity noise increases substantially, resulting in the signal-to-noise ratio not being significantly improved. In the experiment, the probe light intensity of was adopted.
When analyzing the pump light, the two-mode transimpedance amplifier uses different gains, and therefore, the response signal is divided by the maximum value to be normalized, as shown in Figs. 2(c) and 2(d). For the pump light frequency, the resonance absorption frequency is 377.084 THz ( D1 line). Sweeping the pump light frequency near the absorption point with detuning from to 220 GHz, the corresponding response signal curve is detected and shown in Fig. 2(c) on the red line. The response signal is an approximately Lorentzian line shape and reaches the maximum value when the detuning is within 0–10 GHz. When the frequency is maintained near the resonance absorption, the fictitious magnetic field generated by light shifts is effectively suppressed.
In Fig. 2 (d), the pump light intensity varies from to , and the experimental data is well fitted with Eq. (10), as shown in Fig. 2(d) with the red solid line. The response reaches the maximum when the pump intensity is . As predicted by Eq. (10), the scale factor reaches a maximum when the pumping rate is equal to the transverse relaxation rate (). Since pumping rate is proportional to the intensity of the pump light, reached the maximum under an ideal pump light intensity, which is given by
In the single-beam mode, the vapor cell temperature and compensation magnetic field settings were consistent with the double-beam mode. The response signal of the single-beam mode measured by PD1 is demodulated with a lock-in amplifier, and the scale factor () has the following form:
The response signal varied with the frequency detuning of the pump light, as shown by the blue line in Fig. 2(c). The response signal reaches a maximum when the pump light frequency remains at the central frequency. The frequency detuning increases, and the response signal gradually becomes weaker.
From Eq. (13), the scale factor is positively correlated with pump light intensity, which satisfies . In Fig. 2(d) on the blue line, as the intensity increases, the value of the corresponding response signal also increases. The experimental data is well fitted with Eqs. (13) and (15), as shown in Fig. 2(d) with the blue solid line. Although the response signal of the single-beam mode becomes stronger as the pump light intensity increases, the corresponding noise also increases. Optimal performance of the single-beam mode can be achieved when the light intensity reaches .
Figure 3 depicts the response signal of the SERF magnetometer as a function of the temperature from 128°C to 180°C. The two modes have the maximum response value corresponding to the optimal operation temperature. At this temperature, the polarization rate of the alkali-metal atoms is approximately 1/2, and the SERF magnetometer achieves optimal performance. It is worth noting that the optimal temperature of the two modes is different: for the single-beam mode, it is 146°C, and, for the double-beam mode, it is 170°C. The main reason for this difference is that the introduction of the modulation magnetic field increases the transverse relaxation time, making it easier to achieve in the single-beam mode. In actual use, we can first use the double-beam mode (at 170°C) to calibrate the ambient magnetic field noise, and then use the single-beam mode (at 146°C) to perform the biomagnetic measurement. It pre-evaluates the background magnetic field and effectively reduces power consumption and surface temperature in the MEG system.
Figure 3.Response signal as a function of temperature at the optimal optical parameters.
4.2. Sensitivity and bandwidth
Based on the above analysis and experiments, it can be concluded that the optimal response signal can be obtained by controlling the light parameters and the vapor cell temperature to quickly identify the best sensitivity of the SERF magnetometer system. We apply a 100 pT rms sinusoidal calibration signal at 30 Hz along the axis and record the frequency spectrum of the magnetometer response for 150 s. The power spectral density is calculated and averaged for 1 Hz bin. The sensitivity noise floors for single-beam and dual-beam modes are shown in blue and red dashed lines in Fig. 4, respectively. The magnetic field sensitivity is higher than in the single-beam mode and in the double-beam mode under the previous optimal configuration. Figure 5 shows the normalized frequency response of the single-beam mode and double-beam mode with the 100 pT rms calibration magnetic field. The bandwidths at were 141 Hz and 65 Hz for the single-beam mode and double-beam mode, respectively.
Figure 4.Sensitivity of the SERF magnetometer. The probe noise is recorded by blocking the pump light and turning on the probe light. The electronic noise is recorded by blocking the pump light and probe light.
Figure 5.Frequency response of the 100 pT rms calibrated magnetic field at frequency of 1–200 Hz and fitting curve in the single-beam mode and double-beam mode, respectively.
For the mm-level vapor cell, is the ideal sensitivity. However, when the pump light is blocked, the probe noise of the magnetometer is only , as shown in Fig. 4. Therefore, it can be concluded that approaches the noise floor of the four-layer mu-metal magnetic-shielding cylinder. According to the physical parameters of the cylinder, the noise floor is consistent with the theoretical value of the magnetic noise by the Johnson current[
5. Conclusion
In summary, we have developed a dual-mode SERF magnetometer with mm-level vapor cells using two Rb D1 line lasers, which can switch between the single-beam modulation mode and the double-beam DC mode for different applications. When operating under appropriate conditions, this SERF magnetometer shows a magnetic field sensitivity of in the single-beam mode and probe noise of in the double-beam mode. The noise floor of the multilayer mu-metal magnetic-shielding cylinder was measured to be in the double-beam mode. Our design provides a method for realizing miniaturized OPMs with multiple functions, including magnetic-shield background noise measurement and medical imaging.
[1] V. Shah, M. V. Romalis. Spin-exchange relaxation-free magnetometry using elliptically polarized light. Phys. Rev. A, 80, 013416(2009).
[2] I. K. Kominis, T. W. Kornack, J. C. Allred, M. V. Romalis. A subfemtotesla multichannel atomic magnetometer. Nature, 422, 596(2003).
[3] E. Boto, N. Holmes, J. Leggett, G. Roberts, V. Shah, S. S. Meyer, L. D. Muñoz, K. J. Mullinger, T. M. Tierney, S. Bestmann, G. R. Barnes, R. Bowtell, M. J. Brookes. Moving magnetoencephalography towards real-world applications with a wearable system. Nature, 555, 657(2018).
[4] H. B. Dang, A. C. Maloof, M. V. Romalis. Ultrahigh sensitivity magnetic field and magnetization measurements with an atomic magnetometer. Appl. Phys. Lett., 97, 151110(2010).
[5] H. H. Nelson, J. R. McDonald. Multisensor towed array detection system for UXO detection. IEEE Trans. Geosci. Remote Sens., 39, 1139(2001).
[6] D. Sheng, S. Li, N. Dural, M. V. Romalis. Subfemtotesla scalar atomic magnetometry using multipass cells. Phys. Rev. Lett., 110, 160802(2013).
[7] H. Xia, A. Ben-Amar Baranga, D. Hoffman, M. V. Romalis. Magnetoencephalography with an atomic magnetometer. Appl. Phys. Lett., 89, 211104(2006).
[8] A. P. Colombo, T. R. Carter, A. Borna, Y. Jau, C. N. Johnson, A. L. Dagel, P. D. Schwindt. Four-channel optically pumped atomic magnetometer for magnetoencephalography. Opt. Express, 24, 15403(2016).
[9] Y. Yan, G. Liu, H. Lin, K. Yin, J. Lu. VCSEL frequency stabilization for optically pumped magnetometers. Chin. Opt. Lett., 19, 121407(2021).
[10] O. Alem, T. H. Sander, R. Mhaskar, J. LeBlanc, H. Eswaran, U. Steinhoff, Y. Okada, J. Kitching, L. Trahms, S. Knappe. Fetal magnetocardiography measurements with an array of microfabricated optically pumped magnetometers. Phys. Med. Biol., 60, 4797(2015).
[11] S. P. Krzyzewski, A. R. Perry, V. Gerginov, S. Knappe. Characterization of noise sources in a microfabricated single-beam zero-field optically-pumped magnetometer. J. Appl. Phys., 126, 044504(2019).
[12] W. Happer, H. Tang. Spin-exchange shift and narrowing of magnetic resonance lines in optically pumped alkali vapors. Phys. Rev. Lett., 31, 273(1973).
[13] I. M. Savukov, M. V. Romalis. Effects of spin-exchange collisions in a high-density alkali-metal vapor in low magnetic fields. Phys. Rev. A, 71, 023405(2005).
[14] J. Tang, Y. Zhai, L. Cao, Y. Zhang, L. Li, B. Zhao, B. Zhou, B. Han, G. Liu. High-sensitivity operation of a single-beam atomic magnetometer for three-axis magnetic field measurement. Opt. Express, 29, 15641(2021).
[15] H. B. Dang, A. C. Maloof, M. V. Romalis. Ultrahigh sensitivity magnetic field and magnetization measurements with an atomic magnetometer. Appl. Phys. Lett., 97, 151110(2010).
[16] A. P. Colombo, T. R. Carter, A. Borna, Y. Y. Jau, C. N. Johnson, A. L. Dagel, D. D. Schwindt. Four-channel optically pumped atomic magnetometer for magnetoencephalography. Opt. Express, 24, 15403(2016).
[17] R. Wyllie, M. Kauer, R. T. Wakai, T. G. Walker. Optical magnetometer array for fetal magnetocardiography. Opt. Lett., 37, 2247(2012).
[18] D. Y. Ma, J. X. Lu, X. J. Fang, K. Yang, K. Wang, N. Zhang, B. C. Han, M. Ding. Parameter modeling analysis of a cylindrical ferrite magnetic shield to reduce magnetic noise. IEEE Trans. Ind. Electron., 69, 991(2021).
[19] J. Osborne, J. Orton, O. Alem. Fully integrated standalone zero field optically pumped magnetometer for biomagnetism. Proc. SPIE, 10548, 105481G(2018).
[20] V. P. Gerginov, L. Li, M. Gerginov, S. Krzyzewski, S. Knappe. Microfabricated magnetometers for imaging and communication. Proc. SPIE, 11296, 112963C(2020).
[21] O. Alem, T. Sander, R. Mhaskar, J. Leblanc, H. Eswaran, U. Steinhoff. Fetal magnetocardiography measurements with an array of microfabricated optically pumped magnetometers. Phys. Med. Biol., 60, 4797(2015).
[22] O. Alem, R. Mhaskar, J. Ricardo, D. Sheng, S. Knappe. Magnetic field imaging with microfabricated optically-pumped magnetometers. Opt. Express, 25, 7849(2017).
[23] S. J. Seltzer. Developments in alkali-metal atomic magnetometry(2008).
[24] J. Lu, Z. Qian, J. Fang. A fast determination method for transverse relaxation of spin-exchange-relaxation-free magnetometer. Rev. Sci. Instrum., 86, 043104(2015).
[25] M. P. Ledbetter, I. M. Savukov, V. M. Acosta, D. Budker, M. V. Romalis. Spin-exchange-relaxation-free magnetometry with Cs vapor. Phys. Rev. A, 77, 033408(2008).
[26] T. M. Tierney, N. Holmes, S. Mellor, J. D. López, G. R. Barnes. Optically pumped magnetometers: from quantum origins to multi-channel magnetoencephalography. NeuroImage, 199, 598(2019).
[27] R. E. Slocum, B. I. Marton. Measurement of weak magnetic fields using zero-field parametric resonance in optically pumped He4. IEEE Trans. Magn., 9, 221(1973).
[28] C. Cohen-Tannoudji, J. Dupont-Roc. Experimental study of Zeeman light shifts in weak magnetic fields. Phys. Rev. A, 5, 968(1972).
[29] S. K. Lee, M. V. Romalis. Calculation of magnetic field noise from high-permeability magnetic shields and conducting objects with simple geometry. J. Appl. Phys., 103, 084904(2008).
[30] J. Tang, Y. Zhai, B. Zhou, B. Han, G. Liu. Dual-axis closed loop of a single-beam atomic magnetometer: toward high bandwidth and high sensitivity. IEEE Trans. Instrum. Meas., 70, 1504808(2021).
Get Citation
Copy Citation Text
Shaowen Zhang, Jixi Lu, Ying Zhou, Fei Lu, Kaifeng Yin, Di Zhan, Yueyang Zhai, Mao Ye, "Zero field optically pumped magnetometer with independent dual-mode operation," Chin. Opt. Lett. 20, 081202 (2022)
Category: Instrumentation, Measurement, and Optical Sensing
Received: Dec. 6, 2021
Accepted: May. 7, 2022
Published Online: May. 27, 2022
The Author Email: Jixi Lu (lujixi@buaa.edu.cn), Ying Zhou (zhouyingphd@163.com)