Acta Optica Sinica, Volume. 39, Issue 1, 0112002(2019)

Alignment Error Analysis and Optimization of Wave Plate Fast Axis in Four-Path Full Stokes Simultaneous Polarization Imaging System

Yulong Yin1,2,3、*, Xiaobing Sun1,3、*, Maoxin Song1,3, and Wei Chen1,2,3
Author Affiliations
  • 1 Anhui Institute of Optics and Fine Mechanics, Chinese Academy of Sciences, Hefei, Anhui 230031, China
  • 2 University of Science and Technology of China, Hefei, Anhui 230026, China
  • 3 Key Laboratory of Optical Calibration and Characterization, Chinese Academy of Sciences, Hefei, Anhui 230031, China
  • show less
    Figures & Tables(14)
    Schematic of four-path full Stokes simultaneous polarization imaging system
    Measurement errors of Stokes parameters for different incident lights. (a) Natural light; (b) 0° linearly polarized light; (c) 90° linearly polarized light; (d) 45° linearly polarized light; (e) 135° linearly polarized light; (f) right circularly polarized light; (g) left circularly polarized light
    Stokes parameter distributions of 1000 incident light sampling points selected on the Poincaré sphere. (a) 3D distribution; (b) Stokes parameter value
    Stokes parameter measurement errors. (a) Only 1° alignment error of HWP fast axis; (b) only 1° alignment error of QWP fast axis
    Influences of alignment errors of fast axes on polarization measurement accuracy when sampling points with different degrees of polarization used as incident lights. (a) Only HWP fast axis; (b) only QWP fast axis; (c) alignment coupling error of HWP and QWP fast axes
    2-norm conditional number κ of measurement matrix M versus Tp and Ts
    Polarization measurement accuracy of system versus Tp and Ts under concurrence of 1° alignment errors of HWP and QWP fast axes
    Optimization of alignment errors of HWP and QWP fast axes
    ΔA versus alignment error α of HWP fast axis and alignment error β of QWP fast axis
    Polarization measurement accuracy ΔP(α,β,P) versus alignment error α of HWP fast axis and β of QWP fast axis for different degrees of polarization of 1, 0.8, 0.5, 0.2 and 0.1
    • Table 1. Variation in polarization measurement accuracy ΔP(α,β,P=1) of system with alignment error α of HWP fast axis and alignment error β of QWP fast axis

      View table

      Table 1. Variation in polarization measurement accuracy ΔP(α,β,P=1) of system with alignment error α of HWP fast axis and alignment error β of QWP fast axis

      β /(°)ΔP(α,β,P=1)/%
      α=-0.3°α=-0.2°α=-0.15°α=-0.1°α=0°α=0.1°α=0.15°α=0.2°α=0.3°
      -1.23.482.962.752.542.182.482.702.933.44
      -0.62.652.011.731.461.071.461.722.012.65
      -0.522.561.911.611.340.921.331.601.902.56
      -0.22.331.591.230.880.350.881.231.592.33
      02.281.521.140.7600.761.141.522.28
      0.22.331.591.230.880.350.881.231.592.33
      0.522.571.911.601.330.921.341.611.912.57
      0.62.642.011.721.451.071.471.722.022.64
      1.23.432.922.692.492.202.522.732.973.48
    • Table 2. Variation in polarization measurement accuracy A(α,β,P=1) of system with alignment error α of HWP fast axis and alignment error β of QWP fast axis

      View table

      Table 2. Variation in polarization measurement accuracy A(α,β,P=1) of system with alignment error α of HWP fast axis and alignment error β of QWP fast axis

      β /(°)A(α,β,P=1) /%
      α=-0.3°α=-0.2°α=-0.15°α=-0.1°α=0°α=0.1°α=0.15°α=0.2°α=0.3°
      -1.24.304.304.304.304.304.304.304.304.30
      -0.63.852.572.122.122.122.122.122.573.85
      -0.523.852.571.931.831.831.831.932.573.85
      -0.23.852.571.931.290.701.291.932.573.85
      03.852.571.931.2901.291.932.573.85
      0.23.852.571.931.290.701.291.932.573.85
      0.523.852.571.931.831.831.831.932.573.85
      0.63.852.572.122.122.122.122.122.573.85
      1.24.304.304.304.304.304.304.304.304.30
    • Table 3. Variation in polarization measurement accuracy ΔP(α,β,P=5%) of system with alignment error α of HWP fast axis and alignment error β of QWP fast axis

      View table

      Table 3. Variation in polarization measurement accuracy ΔP(α,β,P=5%) of system with alignment error α of HWP fast axis and alignment error β of QWP fast axis

      β /(°)ΔP(α,β,P=5%)/%
      α=-0.5°α=-0.4°α=-0.1°α=0°α=0.1°α=0.4°α=0.5°
      -2.52.432.010.890.620.882.012.44
      -2.42.411.990.850.580.841.992.42
      -1.02.281.830.480.150.481.832.28
      -0.52.271.820.470.060.471.822.27
      02.271.820.4500.451.822.27
      0.52.271.820.470.060.471.822.27
      1.02.281.830.480.150.481.832.27
      2.42.421.990.840.580.851.992.42
      2.52.432.010.880.620.892.012.44
    • Table 4. Variation in polarization measurement accuracy A(α,β,P=5%) with alignment error α of HWP fast axis and alignment error β of QWP fast axis

      View table

      Table 4. Variation in polarization measurement accuracy A(α,β,P=5%) with alignment error α of HWP fast axis and alignment error β of QWP fast axis

      β /(°)A(α,β,P=5%) /%
      α=-0.5°α=-0.4°α=-0.1°α=0°α=0.1°α=0.4°α=0.5°
      -2.52.472.020.930.930.932.022.47
      -2.42.471.990.870.870.871.992.47
      -1.02.471.980.490.250.491.982.47
      -0.52.471.980.490.110.491.982.47
      02.471.980.4900.491.982.47
      0.52.471.980.490.110.491.982.47
      1.02.471.980.490.250.491.982.47
      2.42.471.990.870.870.871.992.47
      2.52.472.020.930.930.932.022.47
    Tools

    Get Citation

    Copy Citation Text

    Yulong Yin, Xiaobing Sun, Maoxin Song, Wei Chen. Alignment Error Analysis and Optimization of Wave Plate Fast Axis in Four-Path Full Stokes Simultaneous Polarization Imaging System[J]. Acta Optica Sinica, 2019, 39(1): 0112002

    Download Citation

    EndNote(RIS)BibTexPlain Text
    Save article for my favorites
    Paper Information

    Category: Instrumentation, Measurement and Metrology

    Received: Jul. 4, 2018

    Accepted: Sep. 5, 2018

    Published Online: May. 10, 2019

    The Author Email:

    DOI:10.3788/AOS201939.0112002