Chinese Journal of Lasers, Volume. 45, Issue 4, 407005(2018)

Design of Imaging Detection System for Fluorescent Immune-Chromatographic Test Strip

Xu Xiaohan, Xia Guo*, Jing Shiqun, Wu Su, and Wang Guodong
Author Affiliations
  • [in Chinese]
  • show less
    Figures & Tables(14)
    Acquisition system of fluorescent immune-chromatographic detector
    Structural diagram of fluorescent immune-chromatographic test strip
    Working principle of fluorescent immune-chromatographic detector
    Segmentation result of fluorescent immune-chromatographic test strip. (a) Original image of fluorescent test strip; (b) binary image after segmentation
    Image of fluorescent immune-chromatographic test strip with noise
    Effect images of fluorescent area positioning.(a) Before positioning by algorithm; (b) after positioning by algorithm
    Picture and testing result of ESEQuant fluorescence analyzer. (a) Picture; (b) testing result
    Number of per column target points in binary image after segmentation
    Differences between calculated results using different extended parameters and ESEQuant results
    Standard fitting curves of concentrations
    • Table 1. Evaluation results of segmentation performance

      View table

      Table 1. Evaluation results of segmentation performance

      Fluorescenceconcentration /(mg·L-1)UMCR
      3.1250.99250.4244
      12.50.99720.5342
      250.99750.6153
      500.99700.6498
      1000.99560.6575
    • Table 2. Eigenvalues of test strip with different concentrations tested by ESEQuant and our instrument

      View table

      Table 2. Eigenvalues of test strip with different concentrations tested by ESEQuant and our instrument

      Fluorescenceconcentration /(mg·L-1)Our instrumentESEQuant
      00.22660.1338
      3.1250.36140.2516
      12.50.64610.5287
      251.01221.0018
      501.74501.6942
      1003.27923.4854
    • Table 3. Fitting results of concentration standard curves

      View table

      Table 3. Fitting results of concentration standard curves

      ParameterFitting result
      ESEQuantOur instrument
      A10.15620.2487
      A24303597929060
      x07707080280011000
      p1.04640.9956
      R20.996890.99944
    • Table 4. Repetitive test results of eigenvalues of test strip with different concentrations

      View table

      Table 4. Repetitive test results of eigenvalues of test strip with different concentrations

      Testing numberFluorescence concentration /(mg·L-1)
      3.12512.52550100
      10.30500.60581.15491.65684.6381
      20.30630.63631.19551.60744.5679
      30.31040.58881.10781.60004.5489
      40.32040.63861.08731.61744.5489
      50.32010.60001.11031.62194.5706
      60.32350.59481.10301.61004.5706
      70.32050.57941.13721.61844.5945
      80.32050.58711.08281.62844.5165
      90.32090.63761.15681.60074.5717
      100.31960.62471.14211.64784.6115
      Mean value0.31800.60931.12781.62094.5739
      SD0.00570.0230.03560.01900.0343
      Cv /%1.793.773.161.170.75
    Tools

    Get Citation

    Copy Citation Text

    Xu Xiaohan, Xia Guo, Jing Shiqun, Wu Su, Wang Guodong. Design of Imaging Detection System for Fluorescent Immune-Chromatographic Test Strip[J]. Chinese Journal of Lasers, 2018, 45(4): 407005