Chinese Journal of Lasers, Volume. 47, Issue 9, 911001(2020)

Parameter Optimization of Laser-Induced Breakdown Spectroscopy Experimental Device Based on Response Surface Methodology

Pan Lijian, Chen Weifang*, Zhou Yanfeng, and Cui Rongfang
Author Affiliations
  • College of Mechanical and Electrical Engineering, Nanjing University of Aeronautics and Astronautics, Nanjing, Jiangsu 210001, China
  • show less
    Figures & Tables(12)
    Schematic of LIBS experimental system
    LIBS spectra of aluminum alloy
    Results of single factor experiment. (a) Laser energy; (b) delay time; (c) depth of focus
    Normal probability distribution of residuals
    Distribution of predicted and actual values
    Effect of interaction between laser energy and delay time on SBR of spectral line. (a) Response surface diagram; (b) contour map
    Effect of interaction between laser energy and depth of focus on SBR of spectral line. (a) Response surface diagram; (b) contour map
    Effect of interaction between delay time and depth of focus on SBR of spectral line. (a) Response surface diagram; (b) contour map
    • Table 1. Factors and levels for single factor experiment

      View table

      Table 1. Factors and levels for single factor experiment

      FactorValue
      Laser energy /mJ80,90,100,110,120
      Delay time /μs1,2,3,4,5
      Depth of focus /mm-1,0,1,2,3
    • Table 2. Factors and levels for response surface methodology

      View table

      Table 2. Factors and levels for response surface methodology

      FactorVariablesymbolLevel
      -101
      Laser energy /mJX1100110120
      Delay time /μsX2123
      Depth of focus /mmX3123
    • Table 3. Design parameters and results of experiments

      View table

      Table 3. Design parameters and results of experiments

      Experiment No.X1 / mJX2 /μsX3 / mmRSBR
      1120236.42
      2110227.12
      3110227.31
      4120127.02
      5110316.61
      6110227.42
      7100216.48
      8110116.85
      9100236.29
      10110336.45
      11110227.39
      12110227.27
      13100326.54
      14120326.87
      15120217.04
      16100126.35
      17110136.56
    • Table 4. Analysis of variance

      View table

      Table 4. Analysis of variance

      SourceSum of squareDegree of freedomMean square errorFPSignificant
      Model2.3190.2617.430.0005Y
      X10.3610.3624.240.0017Y
      X20.01210.0120.820.3965
      X30.2010.2013.470.0080Y
      X1X20.02910.0291.960.2040
      X1X30.04610.0463.140.1198
      X2X30.004210.00420.290.6088
      X120.4710.4731.790.0008Y
      X220.3110.3121.380.0024Y
      X320.7110.7148.290.0002Y
      Residual0.1070.015
      Lack of fit0.04730.0161.130.4380N
      Pure error0.05640.014
    Tools

    Get Citation

    Copy Citation Text

    Pan Lijian, Chen Weifang, Zhou Yanfeng, Cui Rongfang. Parameter Optimization of Laser-Induced Breakdown Spectroscopy Experimental Device Based on Response Surface Methodology[J]. Chinese Journal of Lasers, 2020, 47(9): 911001

    Download Citation

    EndNote(RIS)BibTexPlain Text
    Save article for my favorites
    Paper Information

    Category: spectroscopy

    Received: Mar. 30, 2020

    Accepted: --

    Published Online: Sep. 16, 2020

    The Author Email: Weifang Chen (meewfchen@nuaa.edu.cn)

    DOI:10.3788/CJL202047.0911001

    Topics