Laser & Optoelectronics Progress, Volume. 59, Issue 6, 0617023(2022)

Modified Beer-Lambert Law for NdYAG Laser Transmission in Blood

Xiaoxi Dong1, Yaqun Kong2, Huijuan Yin1, Weinan Dong1, Jizhi Zhao3, and Jichun Yang1、*
Author Affiliations
  • 1Institute of Biomedical Engineering, Chinese Academy of Medical Sciences & Peking Union Medical College, Tianjin 300192, China
  • 2Beijing Friendship Hospital, Capital Medical University, Beijing 100050, China
  • 3Peking Union Medical College Hospital, Chinese Academy of Medical Sciences & Peking Union Medical College, Beijing 100730, China
  • show less
    Figures & Tables(7)
    Transmitted laser energy detecting system of Nd∶YAG. (a) Schematic diagram of system; (b) principle of detection
    Comparison between theoretical value and real detected value. (a) Relationship between detected depth and target depth; (b) relationship between geometry-dependent factor and detected depth
    Residual analysis of geometry-dependent factor and detected depth by four fitting methods. (a) Linear regression residual; (b) log regression residual; (c) index regression residual; (d) power regression residual
    Relationship between laser decay and depth in blood
    • Table 1. Optical properties of blood

      View table

      Table 1. Optical properties of blood

      Blood sortProportion in venous blood /%22μa /cm-1[192223μs' /cm-1[212425
      Oxygenated blood58.783.023.41
      Deoxygenated blood41.220.286.61
      Venous blood1.894.73
    • Table 2. Selected scales of transferpettor, theoretical thickness and transmitted laser intensity

      View table

      Table 2. Selected scales of transferpettor, theoretical thickness and transmitted laser intensity

      Target thickness /mmScale of transferpettor /μLTheoretical thickness /mmDetected thickness /mmTransmitted laser intensity /VRreal /%
      00001.40±0.02
      1.001950.990.95±0.020.97±0.0930.86
      1.252451.241.15±0.020.87±0.0438.35
      1.502951.491.37±0.030.58±0.0358.73
      1.753451.751.65±0.050.43±0.0269.21
      2.003952.001.87±0.080.32±0.0677.41
    • Table 3. Parameters and regression decision coefficient for the fitting of geometry-dependent factor

      View table

      Table 3. Parameters and regression decision coefficient for the fitting of geometry-dependent factor

      NameFormulaabR2
      Linear fittingGd=ad+b1.0699-0.76510.9295
      Log fittingGd=alnd+b1.62000.25340.9139
      Index fittingGd=aexp(bd)0.1902-0.10240.8873
      Power fittingGd=adb0.38081.82190.6289
    Tools

    Get Citation

    Copy Citation Text

    Xiaoxi Dong, Yaqun Kong, Huijuan Yin, Weinan Dong, Jizhi Zhao, Jichun Yang. Modified Beer-Lambert Law for NdYAG Laser Transmission in Blood[J]. Laser & Optoelectronics Progress, 2022, 59(6): 0617023

    Download Citation

    EndNote(RIS)BibTexPlain Text
    Save article for my favorites
    Paper Information

    Category: Medical Optics and Biotechnology

    Received: Nov. 12, 2021

    Accepted: Jan. 7, 2022

    Published Online: Mar. 8, 2022

    The Author Email: Jichun Yang (yjc@tju.edu.cn)

    DOI:10.3788/LOP202259.0617023

    Topics