    Schematic of the experimental system
    Characteristics of model parameters under different azimuth angles. (a) Maximum value; (b) minimum value; (c) mean value; (d) variance
    Model parameters when the fast axis azimuth is 0°. (a) m00; (b) m01; (c) m02; (d) m03
    Model parameters when the fast axis azimuth is 15°. (a) m00; (b) m01; (c) m02; (d) m03
    Model parameters when the fast axis azimuth is 30°. (a) m00; (b) m01; (c) m02; (d) m03
    Model parameters when the fast axis azimuth is 45°. (a) m00; (b) m01; (c) m02; (d) m03
    Modulated images of horizontally polarized light under different fast axis azimuth angles. (a) 0°; (b) 15°; (c) 30°; (d) 45°
    • Table 1. Detection accuracy under different fast axis azimuth angles

      View table

      Table 1. Detection accuracy under different fast axis azimuth angles

      Azimuth-45°-30°-15°15°30°45°
      S1-AVME0.00070.00210.00100.00070.00110.00470.0006
      S2-AVME0.00070.00360.00080.00100.00090.00520.0007
      S3-AVME0.32360.00620.00050.00040.00080.01340.3313
      Average value0.10830.00400.00080.00070.00090.00780.1109
    • Table 2. Measurement accuracy under different fast axis azimuth angles when the LED output current is 64 mA

      View table

      Table 2. Measurement accuracy under different fast axis azimuth angles when the LED output current is 64 mA

      Azimuth-45°-30°-15°15°30°45°
      S1-AVME0.02650.00340.01760.00260.00260.01760.0134
      S2-AVME0.05290.00210.04930.00250.00270.04930.0247
      S3-AVME0.15470.00240.00890.00280.00200.00890.3500
      Average AVME0.07800.00260.02530.00260.00240.02530.1294
    • Table 3. Measurement accuracy under different fast axis azimuth angles when the LED output current is 65 mA

      View table

      Table 3. Measurement accuracy under different fast axis azimuth angles when the LED output current is 65 mA

      Azimuth-45°-30°-15°15°30°45°
      S1-AVME0.02650.00370.01840.00190.00240.01300.0145
      S2-AVME0.05290.00170.05380.00270.00200.02320.0259
      S3-AVME0.15470.00210.00740.00230.00240.00410.2761
      Average AVME0.07800.00250.02650.00230.00230.01340.1055
    • Table 4. Measurement accuracy under different fast axis azimuth angles when the LED output current is 66 mA

      View table

      Table 4. Measurement accuracy under different fast axis azimuth angles when the LED output current is 66 mA

      Azimuth-45°-30°-15°15°30°45°
      S1-AVME0.02450.00820.00280.00240.00900.14020.0389
      S2-AVME0.01020.01130.00200.00240.01430.24890.2194
      S3-AVME0.16840.00320.00190.00170.00440.07000.2728
      Average AVME0.06770.00760.00220.00210.00920.15300.1770
    Tools

    Get Citation

    Copy Citation Text

    Xinxin Zhang, Tianlei Ning, Lihui Liu, Yanqiu Li. Influence of Fast Axis Angle of Wave Plate on Accuracy of Spatially Modulated Polarization Measurements[J]. Acta Optica Sinica, 2024, 44(16): 1612003

    Download Citation

    EndNote(RIS)BibTexPlain Text
    Save article for my favorites
    Paper Information

    Category: Instrumentation, Measurement and Metrology

    Received: Mar. 20, 2024

    Accepted: Apr. 29, 2024

    Published Online: Aug. 5, 2024

    The Author Email: Liu Lihui (liulihui@bit.edu.cn), Li Yanqiu (liyanqiu@bit.edu.cn)

    DOI:10.3788/AOS240753

    CSTR:32393.14.AOS240753

    Topics