Chinese Optics Letters, Volume. 6, Issue 7, 502(2008)
Superiority of zoom lens coupling in designing a novel X-ray image detector
We design a novel X-ray image detector by lens coupling a Gd2O2S:Tb intensifying screen with a high performance low-light-level (L3, which often means luminescence less than 10^(-3) Lux) image intensifier. Different coupling effects on imaging performance between zoom lens and fix-focus lens are analyzed theoretically. In experiment, for designing a detector of 15-inch visual field, the system coupled by zoom lens is of 12.25-lp/cm resolution, while the one with fix-focus lens is 10 lp/cm. The superiority of zoom lens is validated. It is concluded that zoom lens preserves the image information better than fix-focus lens and improves the imaging system’s performance in this design, which is referential to the design of other optical imaging systems.
Get Citation
Copy Citation Text
Chunyu Yu, Benkang Chang, Dianxiu Wei, "Superiority of zoom lens coupling in designing a novel X-ray image detector," Chin. Opt. Lett. 6, 502 (2008)
Received: Jan. 22, 2008
Accepted: --
Published Online: Jul. 11, 2008
The Author Email: Chunyu Yu (yukate@163.com)