NUCLEAR TECHNIQUES, Volume. 48, Issue 1, 010401(2025)

Development of a 3D dose distribution validation water tank for rotating Gantry treatment terminals

Xincai KANG1,2, Faming LUO1,3、*, Youtang LI2, Zhiguo XU1,3, Ruishi MAO1,3, Zulong ZHAO1, and Yucong CHEN1
Author Affiliations
  • 1Institute of Modern Physics, Chinese Academy of Sciences, Lanzhou 730000, China
  • 2Lanzhou University of Technology, Lanzhou 730050, China
  • 3University of Chinese Academy of Sciences, Beijing 100049, China
  • show less
    Figures & Tables(12)
    Schematic of 3D water phantom measurement program
    Overall structural diagram of the 3D water tank
    Solidworks simulation results of box tank structure by Solidworks (a) Deformation of open box tank, (b) 8 N·m preload on tank top cover, (c) Deformation of the sealed tank, (d) Deformation of tank placed at 45° angle position
    Diagram of sealed water phantom model
    Diagram of multi-strip ionization chamber and flexible circuit boards
    Schematic of three-dimensional movement mechanism
    Collimation calibration of the 3D moving mechanisms (a) Collimation calibration platform of 3D moving mechanisms, (b) Position deviation during X-direction movement, (c) Position deviation during Y-direction movement, (d) Position deviation during Z-direction movement
    Illustration of sealing performance test of water phantom
    Snapshot of water phantom beam test at treatment room
    3D dose distribution of the 400 MeV·u-1 carbon ion pencil beam in water (a) Measured results with the water tank, (b) Simulation results of Geant4
    Comparison of the depth dose curves measured by the 3D water phantom with the results measured by the PTW MP3 water phantom (a) and relative deviation (b)
    • Table 1. Comparison of scanning mechanism of the 3D water tank and PTW MP3-P45

      View table
      View in Article

      Table 1. Comparison of scanning mechanism of the 3D water tank and PTW MP3-P45

      技术规格

      Specifications

      扫描范围

      Scanning range / mm

      重复精度

      Reproducibility / mm

      位置偏差

      Position deviation / mm

      PTW MP3-P45

      X: 500

      Y: 525

      Z: 400

      0.10.1

      三维水箱

      3D water tank

      X: 320

      Y: 160

      Z: 160

      0.1≤0.24
    Tools

    Get Citation

    Copy Citation Text

    Xincai KANG, Faming LUO, Youtang LI, Zhiguo XU, Ruishi MAO, Zulong ZHAO, Yucong CHEN. Development of a 3D dose distribution validation water tank for rotating Gantry treatment terminals[J]. NUCLEAR TECHNIQUES, 2025, 48(1): 010401

    Download Citation

    EndNote(RIS)BibTexPlain Text
    Save article for my favorites
    Paper Information

    Category: NUCLEAR ELECTRONICS AND INSTRUMENTATION

    Received: Jun. 5, 2024

    Accepted: --

    Published Online: Feb. 26, 2025

    The Author Email: LUO Faming (罗发明)

    DOI:10.11889/j.0253-3219.2025.hjs.48.240231

    Topics