Chinese Journal of Lasers, Volume. 47, Issue 12, 1204001(2020)

Method for Volume Error Calibration of CMM Based on Inverse-Distance Weighting Algorithm

Chen Hongfang*, Zhang Shuang, Liang Chaowei, Tang Liang, and Shi Zhaoyao
Author Affiliations
  • Beijing Engineering Research Center of Precision Measurement Technology and Instruments, Beijing University of Technology, Beijing 100124, China
  • show less
    Figures & Tables(13)
    Schematic of laser tracing multi-station measurement system
    Schematic of smallest hexahedron which contains point J
    Laser tracing multi-station measurement system
    Spatial distribution of sampling points (circle is interpolation sample data, and asterisk is accuracy evaluation data)
    IDW interpolation results. (a) Volume error in x direction; (b) volume error in y direction; (c) volume error in z direction
    Difference between volume error obtained by IDW algorithm and volume error measured by laser tracing multi-station measurement system
    Schematic of trilinear interpolation model
    Volume error results obtained by trilinear interpolation algorithm. (a) Volume error in x direction; (b) volume error in y direction; (c) volume error in z direction
    • Table 1. Theoretical station coordinates of laser tracer

      View table

      Table 1. Theoretical station coordinates of laser tracer

      Station No.X /mmY /mmZ /mm
      1-50-70-620
      2520-50-470
      35201300-620
      4-401250-470
      570-70-620
      6-6040-620
    • Table 2. Position coordinates of laser tracer when CMM has errors

      View table

      Table 2. Position coordinates of laser tracer when CMM has errors

      Station No.X /mmY /mmZ /mm
      1-50.0000-69.9999-620.0000
      2520.0001-50.0003-470.0001
      3520.00021300.0004-620.0002
      4-39.99981249.9997-470.0000
      570.0002-69.9996-619.9999
      6-60.000140.0000-620.0000
    • Table 3. Calibration errors of station coordinates

      View table

      Table 3. Calibration errors of station coordinates

      Station No.ΔX /mmΔY /mmΔZ /mm
      1-0.06×10-40.00010.20×10-4
      21.00×10-4-0.0003-0.80×10-4
      32.00×10-40.0004-2.00×10-4
      42.00×10-4-0.00030.30×10-4
      52.00×10-40.00041.00×10-4
      6-0.90×10-4-0.0900×10-40.09×10-4
    • Table 4. Station coordinates of laser tracer and initial measurement distances

      View table

      Table 4. Station coordinates of laser tracer and initial measurement distances

      jXj /mmYj /mmZj /mmdj /mm
      1-55.765970.0572-613.4493-212.7153
      2522.1023-51.4716-465.7633-253.9001
      3526.38571310.3801-613.4811-185.4420
      4-41.32141269.6249-465.7788-206.6375
      572.6585-77.6446-613.4448-212.7214
      6-65.641841.8233-613.4556-212.9387
    • Table 5. Algorithm accuracy

      View table

      Table 5. Algorithm accuracy

      DirectionIDW algorithmTrilinear interpolationalgorithm
      RMSER2RMSER2
      x0.0002550.9626290.0003070.943569
      y0.0002250.9732570.0002900.961579
      z0.0003450.9847130.0003600.984293
    Tools

    Get Citation

    Copy Citation Text

    Chen Hongfang, Zhang Shuang, Liang Chaowei, Tang Liang, Shi Zhaoyao. Method for Volume Error Calibration of CMM Based on Inverse-Distance Weighting Algorithm[J]. Chinese Journal of Lasers, 2020, 47(12): 1204001

    Download Citation

    EndNote(RIS)BibTexPlain Text
    Save article for my favorites
    Paper Information

    Category: Measurement and metrology

    Received: Jun. 12, 2020

    Accepted: --

    Published Online: Nov. 17, 2020

    The Author Email: Hongfang Chen (chf0302@126.com)

    DOI:10.3788/CJL202047.1204001