Opto-Electronic Engineering, Volume. 52, Issue 6, 250049(2025)

Optical design and applications of endoscopic optical coherence tomography probe with large field of view

Haoyang Yu1,2, Jinyuan Hu1,2, Jingye Gu1,2, Sujian Wu1,2, Jinyu Fan1,2, Yi He1,2, and Guohua Shi1,2、*
Author Affiliations
  • 1School of Biomedical Engineering, University of Science and Technology of China, Suzhou, Jiangsu 215163, China
  • 2Jiangsu Key Laboratory of Medical Optics, Suzhou Institute of Biomedical Engineering and Technology, Chinese Academy of Sciences, Suzhou, Jiangsu 215163, China
  • show less
    Figures & Tables(12)
    Structure diagram of optical system
    Optimize performance results. (a) Curves of optical transfer function; (b) Curves of optical transfer function after system inversion; (c) Distortion
    Schematic diagram of the endoscopic swept source optical coherence tomography system
    Structure of endoscopic probe
    Physical diagram of endoscopic probe
    Infrared detection card image captured by endoscope probe
    Tape images in various forms taken by endoscope probe. (a) Tape image taken by large field of view endoscopic probe; (b) Tape image taken by gradient refractive index lens endoscopic probe; (c) Deep fracture; (d) Shallow fracture
    Endoscopic probe image of finger
    Endoscopic probe images of isolated porcine eyes. (a) Iris; (b) Sclera; (c) Retina
    • Table 1. Endoscopic objective lens design specifications

      View table
      View in Article

      Table 1. Endoscopic objective lens design specifications

      Design parameterData
      Working wavelength/nm1000~1100
      Field of view/mm6
      Working length/mm<10
      F3.6
      Diameter/mm1.6
      MTF>0.3@120 lp/mm
    • Table 2. Tolerance data

      View table
      View in Article

      Table 2. Tolerance data

      ItemValue
      Radius/fringe±1.0
      Thickness/mm±0.02
      Surface tilt (X and Y)/(°)±0.016
      Surface decenter (X and Y)/mm±0.01
      Element tilt (X and Y)/(°)±0.050
      Element decenter (X and Y)/mm±0.01
      S+A irregularity/fringe±0.2
      Refractive index±0.0005
      Abbe number/%±0.5
    • Table 3. Results of tolerance analysis

      View table
      View in Article

      Table 3. Results of tolerance analysis

      ItemMTF value0.00 mm0.13 mm0.22 mm0.30 mm0.44 mm
      >90%0.226059680.262297630.235847850.230009390.187731940.17901241
      >80%0.285987010.322112250.290404740.296267690.243668940.25094786
      >50%0.384426670.414925600.390797300.387745710.355241710.35156355
      >20%0.457186260.483680730.459003750.459884230.428656500.42745950
      >10%0.483695660.509587850.487491360.485480720.461760840.45997851
    Tools

    Get Citation

    Copy Citation Text

    Haoyang Yu, Jinyuan Hu, Jingye Gu, Sujian Wu, Jinyu Fan, Yi He, Guohua Shi. Optical design and applications of endoscopic optical coherence tomography probe with large field of view[J]. Opto-Electronic Engineering, 2025, 52(6): 250049

    Download Citation

    EndNote(RIS)BibTexPlain Text
    Save article for my favorites
    Paper Information

    Category: Article

    Received: Feb. 23, 2025

    Accepted: Apr. 24, 2025

    Published Online: Sep. 3, 2025

    The Author Email: Guohua Shi (史国华)

    DOI:10.12086/oee.2025.250049

    Topics