Polarization is an essential property of light, and measuring the polarization state fast and precisely is an important issue in many scientific research fields, such as ellipsometry[
Chinese Optics Letters, Volume. 19, Issue 2, 021201(2021)
Spatially modulated polarimetry based on a vortex retarder and Fourier analysis
We report a spatially modulated polarimetry scheme by using a zero-order vortex half-wave retarder (ZVHR) and a spatial Fourier analysis method. A ZVHR is employed to analyze the input polarized light and convert it into a vectorial optical field, and an analyzer is set after the ZVHR to form an hourglass intensity pattern due to the spatial polarization modulation. Then, the input light’s Stokes parameters can be calculated by spatial Fourier analysis of the hourglass pattern with a single shot. The working principle of the polarimeter has been analyzed by the Stokes–Mueller formalism, and some quantitative measuring experiments of different polarization states have been demonstrated. The experimental results indicate that the proposed polarimeter is accurate, robust, and simple to use.
1. Introduction
Polarization is an essential property of light, and measuring the polarization state fast and precisely is an important issue in many scientific research fields, such as ellipsometry[
To overcome the adverse effects of conventional methods and to satisfy the increasing demands for fast and precise measurement, researchers have proposed many new polarimetric schemes, such as a liquid crystal retarder (LCR) polarimeter[
The vortex retarder is a special wave plate that has a constant retardance across the clear aperture, but its fast axis rotates continuously along the azimuth[
Sign up for Chinese Optics Letters TOC Get the latest issue of Advanced Photonics delivered right to you!Sign up now
In our earlier research, a spatially modulated ellipsometric scheme by using a vortex retarder has been demonstrated theoretically[
2. Theoretical and Experimental Investigation
The schematic of the experimental setup is shown in Fig. 1, which includes two parts, the polarization state generation (PSG) system and the polarization state detection (PSD) system. In the PSG system, an ordinary He–Ne laser with the wavelength of 632.8 nm was employed as an optical source, and an integrating sphere and a positive lens (focal length 100 mm) were utilized to produce a uniform and nearly collimated light field for spatial polarization modulation. The incident beams with different states of polarization (SOPs) were generated by a linear polarizer (Thorlabs, LPVISE100-A) and a quarter-wave plate (Thorlabs, WPQ10E-633), which were mounted on the motorized rotation stages (Thorlabs, KPRM1E). In the PSD system, the zero-order vortex half-wave retarder (ZVHR) (Thorlabs, WPV10L-633) was utilized to convert the unknown input polarized light into a vectorial optical field, and an hourglass-shaped intensity distribution can be obtained when this vectorial optical field is analyzed by a linear polarizer (Thorlabs, LPVISE100-A)[
Figure 1.Schematic of our experimental configuration. IS, integrating sphere; L, positive lens; P, polarizer; Q, quarter-wave plate; ZVHR, zero-order vortex half-wave retarder; A, analyzer.
The operating principle of our proposed polarimeter (PSD system) can be analyzed with the Mueller matrix formula as
Equation (3) represents a truncated spatial Fourier series whose coefficients are determined by the Stokes parameters of the incident polarized beam[
If the first Stokes parameter is normalized (i.e., ), and and are also normalized with respect to , then the fourth Stokes parameter can be calculated as
However, the measured accuracy of Stokes parameters will be impaired by many factors in real experiments, such as ambient stray light, imperfections of optical components, and image noise. The comprehensive effect of these adverse factors will cause a decrease of the image contrast, even though the incident light of the polarimeter is a fully polarized light, and the obtained spatially modulated image will be similar to the situation where a partially polarized light is input into the polarimeter. Therefore, the measurement errors caused by the above adverse factors can be calibrated by attributing them to the decrease of DOP, and the actual DOP can be estimated by employing a typical linearly polarized light with known polarization state as a probe light. Since the parameters , , and can be calculated by the proposed scheme directly, the DOP of a linearly polarized probe light can be obtained as
Before the formal measurements, an arbitrary linearly polarized light is employed as a probe light to estimate the corresponding DOP of the current environment, and the DOP can be obtained according to Eq. (6) by using the measured results of and . Based on the calibrating result of DOP in the current measuring environment, the full Stokes parameters of any incident light can be obtained according to the Eqs. (4) and (5).
The basic working principle and measuring procedure of the proposed scheme will be introduced as follows. Firstly, in the PSG system, the transmission axis of the polarizer and the fast axis of the quarter-wave plate were rotated to 0° for generating a horizontal linearly polarized input light, which was also employed as the probe light to estimate the DOP. The input light was modulated and detected by the PSD system, the corresponding intensity image was recorded by the camera and shown in Fig. 2(a), and the radius of the hourglass-shaped intensity image is about 800 pixels. As shown in Fig. 2(a), the marginal and central parts were not well modulated, owing to the vignetting effect of the optical system and the imperfection of optical components, so a ring area was cut from Fig. 2(a) and chosen as a region of interest (ROI), which is beneficial for improving the measuring accuracy and image processing speed. To make full use of the modulated information in the image, a thicker ring area is expected, but more defective pixels may be included at the same time, which also limits the measurement accuracy. In our measurement, the ring thickness was chosen as 200 pixels, and its inner and outer radii were 300 pixels and 500 pixels, respectively, which is shown in Fig. 2(b). To obtain the intensity modulation curve , the integrals were performed from the center of the ring image along the radius at each azimuth angle , the integral values at each azimuth angle are glued to form the integral curve, and then the intensity modulation curve was obtained by normalizing the integral curve. In the real measurement, these procedures were achieved by performing 180 times Radon transformations[
Figure 2.Typical measured results of linearly polarized light. (a) The recorded hourglass intensity image of horizontal linearly polarized light. (b) A ring ROI cut from (a). (c) The normalized intensity modulation curves as a function of azimuth angle
3. Results and Discussion
To validate the accuracy and feasibility of our polarimeter, the quantitative polarization measuring experiments have been carried out in two steps. At first, the incident beams with nine particular polarization states have been measured, including the 0°, 45°, 90°, and 135° linearly polarized light. The left and right circularly polarized light and three elliptically polarized lights were generated by the PSG system, and then they were examined by the PSD system, respectively. The measured results have been shown in Table 1.
|
It can be known from Table 1 that the measured Stokes parameters of the incident light were consistent with their theoretical values, and the maximum measured error (MME) and the average value of the absolute measured error (AVME) for the nine measurements were 0.0385, 0.0333, and 0.0460 (MME of , , and ) and 0.0154, 0.0233, and 0.0224 (AVME of , , and ), respectively, which indicated that the presented scheme is feasible, and its accuracy can meet the needs of general polarization measurement.
To confirm the feasibility and accuracy of our polarimeter further, more experiments have been carried out by changing the polarization state of incident light regularly. In the PSG system, the transmission axis of the polarizer was fixed at 20°, and the fast axis of the quarter-wave plate was rotated from 0° to 180° with a step of 5° to generate 37 different polarization states. The Stokes parameters of the 37 incident beams were measured and plotted in Fig. 3.
Figure 3.Experimental results measured from multiple regularly varied SOPs. (a) Experimental (black star, S1; red circle, S2; blue triangle, S3) and theoretical (black line, S1; red line, S2; blue line, S3) Stokes parameters corresponding to different SOPs. (b) Experimental (red star) and theoretical (blue line) data shown on a Poincaré sphere.
According to the experimental results shown in Fig. 3, we can easily find that the measured results were well consistent with the predicted values in both Stokes curves and the Poincaré sphere, and the fitted curves based on these discrete experimental results were also well matched with the theoretically predicted results. Moreover, the MME and the AVME for 37 continuous measurements have also been calculated, and the results were 0.0391, 0.0319, and 0.0555 (MME of , , and ), 0.0252, 0.0166, and 0.0203 (AVME of , , and ), respectively, which further indicated that the proposed scheme is available to measure the polarization state of the arbitrary polarized beam accurately and conveniently.
In summary, 9 discrete measurements with typical SOPs and 37 continuous measurements have been performed successfully to validate the feasibility of our polarimeter, and the results indicated that the measured error range of three Stokes parameters was mainly between 0.01 and 0.05, which is acceptable in most polarization measurement applications. However, some unavoidable measurement errors still exist, and these errors can probably be attributed to the misalignment of optical components, the imperfections of optical components, the image noise, and ambient stray light. The misalignment of the ZVHR and the analyzer will cause the actual bright areas’ orientation to deviate from the ideal position, and this would impair the accuracy of the polarimeter. To improve the accuracy further, the rotating stage with higher accuracy should be utilized in calibration to decrease the influence of misalignment, and better optical components and low noise camera should be employed to improve the modulation quality of the intensity image. Moreover, the whole polarimeter system can be encapsulated in a closed tube to avoid the adverse effect of ambient stray light.
Based on the theoretical analysis and operation procedure of the polarimeter, it is not difficult to conclude that the vortex retarder can be an efficient device for achieving spatial modulation in a polarimeter, and the proposed polarimetry is a novel scheme and easy to implement. This scheme has many distinctive advantages. Firstly, the polarimeter can measure all of the Stokes parameters of polarized light accurately at a single shot, which has potential for fast and precise measurement. Secondly, the configuration of our polarimeter is very simple; it only consists of a vortex retarder, a polarizer, and a camera, which is promising for miniaturization. Thirdly, no mechanical moving part is needed in the polarimeter, and it is very stable and robust. Moreover, the phase retardance change of the ZVHR caused by wavelength drift has little effect on the modulated intensity image, and the SOP can be measured from a single image. Thus, the measurement results are insensitive to the wavelength and power fluctuations of the light source.
However, this scheme still has some unpleasant defects, which are that the current method is only applicable to completely polarized beams, and the handedness of the input light cannot be recognized. Since the fourth Stokes parameter is determined by Eq. (5), the estimations of DOP and of partially polarized and unpolarized beams will be inaccurate, and the left-handed and right-handed polarized light cannot be distinguished. Fortunately, if a zero-order vortex quarter-wave retarder (ZVQR) is employed to substitute the ZVHR in the PSD system, in a similar way as it is done in a typical Stokes polarimeter using a quarter-wave plate[
4. Conclusion
In conclusion, a spatially modulated polarimetry scheme has been demonstrated based on the vortex retarder and spatial Fourier analysis. In a single acquisition of the spatially modulated image, the four Stokes parameters of input light can be measured, and the experimental results indicate that this approach is accurate, feasible, and convenient for examining the polarization states of light.
[1] H. Fujiwara. Spectroscopic Ellipsometry: Principles and Applications(2007).
[32] R. Deans. The Radon Transform and Some of Its Applications(2007).
[34] D. H. Goldstein. Polarized Light(2011).
Get Citation
Copy Citation Text
Chao Gao, Bing Lei, "Spatially modulated polarimetry based on a vortex retarder and Fourier analysis," Chin. Opt. Lett. 19, 021201 (2021)
Category: Instrumentation, Measurement, and Optical Sensing
Received: Jul. 20, 2020
Accepted: Sep. 4, 2020
Posted: Sep. 7, 2020
Published Online: Dec. 18, 2020
The Author Email: Bing Lei (cslb2017@163.com)