Chinese Journal of Lasers, Volume. 49, Issue 13, 1310002(2022)

Multichannel Pure Rotational Raman Spectroscopic System for Absolute-Probing Temperature Lidar Throughout a Day

Shichun Li1,2、*, Xu Wang1, Penghui Zhang1, Wenhui Xin1,2, Dengxin Hua1,2、**, Xianglong Hu1, and Yuehui Song1,2
Author Affiliations
  • 1School of Mechanical and Precision Instrument Engineering, Xi’an 710048, Shaanxi, China
  • 2Collaborative Innovation Centre for Modern-Equipment Environmentally-Conscious Manufacturing, Xi’an 710048, Shaanxi, China
  • show less
    Figures & Tables(10)
    Pure rotational Raman spectra of nitrogen molecular. (a) Relationship between even-rotational quantum number and adjacent Raman lines; (b) backscattering cross-section
    Pure rotational Raman laser lidar for atmospheric temperature absolute detection throughout a day
    Relationship between FSR of FPI and wavelength difference between adjacent pure rotational quantum number Raman spectra
    Matching of FPI transmittance with pure rotational Raman spectra of N2. (a) Deviation analysis; (b) schematic
    Maximum and minimum transmittance of FPI under different effective fineness values. (a) Maximum transmittance; (b) minimum transmittance
    Performance analysis of pure rotational Raman lidar. (a) Lidar echo power in typical channel; (b) signal-noise-ratio (SNR) in typical channel
    Atmospheric temperature profile with absolute probing retrieval
    • Table 1. Primary specifications of pure rotational Raman lidar

      View table

      Table 1. Primary specifications of pure rotational Raman lidar

      ParameterValue
      Exciting wavelength /nm532.0
      Pulse energy /mJ300
      Pulse duration /ns10
      Pulse repetition frequency /Hz20
      Diameter of telescope /mm300
      Field of view /mrad0.5
      Spatial resolution /m3.75(@25 ns)
      PMT quantum efficiency0.17
    • Table 2. Primary parameters of FPI

      View table

      Table 2. Primary parameters of FPI

      ParameterValue
      a0.001
      R0.9519
      n1.0003
      h /μm629.300
      FD45.63
      FWMH /pm6
      Maximum transmittance0.556
      Suppression ratio to elastic scattering /dB30
    • Table 3. Performance parameters of multi-channel Raman spectroscope

      View table

      Table 3. Performance parameters of multi-channel Raman spectroscope

      ParameterValue
      J=6J=8J=10J=12J=14J=16J=18J=20
      Transmittance0.3150.2670.2420.2400.2660.3310.4290.395
      Suppression rate /dB5760.56560.567.969.36969.5
    Tools

    Get Citation

    Copy Citation Text

    Shichun Li, Xu Wang, Penghui Zhang, Wenhui Xin, Dengxin Hua, Xianglong Hu, Yuehui Song. Multichannel Pure Rotational Raman Spectroscopic System for Absolute-Probing Temperature Lidar Throughout a Day[J]. Chinese Journal of Lasers, 2022, 49(13): 1310002

    Download Citation

    EndNote(RIS)BibTexPlain Text
    Save article for my favorites
    Paper Information

    Category: remote sensing and sensor

    Received: Oct. 26, 2021

    Accepted: Dec. 22, 2021

    Published Online: Jun. 13, 2022

    The Author Email: Li Shichun (lsczqz@xaut.edu.cn), Hua Dengxin (dengxinhua@xaut.edu.cn)

    DOI:10.3788/CJL202249.1310002

    Topics