Acta Optica Sinica, Volume. 40, Issue 20, 2029001(2020)

On-Orbit Performance Evaluation of On-Board Calibration Component of GF-5 Visible and Infrared Multispectral Imager

Wenxin Huang1、*, Liming Zhang1, Xiaolong Si1, Heyu Xu1, Hongyao Chen1, Xin Li1, and Yanhua Zhao2
Author Affiliations
  • 1Key Laboratory of Optical Calibration and Characterization, Chinese Academy of Sciences, Hefei, Anhui 230031, China
  • 2Beijing Institute of Space Mechanics & Electricity, Beijing 100094, China;
  • show less
    Figures & Tables(8)
    On-board calibration process on VIMI
    Working schematic diagram of SDRDM
    Change curve of SDRDM solar observation port signal with azimuth
    H factor change curve with time and azimuth
    H factor curve after correction for B3 and B5 bands
    • Table 1. H factor under the angle of similar azimuth

      View table

      Table 1. H factor under the angle of similar azimuth

      Date|Azimuth /(°)B1B3BrB5
      2018/8/28|157.7391111
      2019/4/16|158.9670.99610.98990.99170.9905
      2019/5/16|161.2061.00760.99260.99320.9926
      2019/8/28|159.8461.00730.99841.00201.0025
      2020/3/17|159.4151.00461.00161.00501.0076
      Average0.00490.00500.00580.0071
    • Table 2. Uncertainty of SDRDM monitoring measurement

      View table

      Table 2. Uncertainty of SDRDM monitoring measurement

      SourceUncertainty /%
      Relative SD BRDF measurement (μ1)0.50
      Relative SD cosine error of incidence (μ2)0.54
      Relationship between transmittance distribution and incident angle (μ3)0.35
      Stability of SDRDM (μ4)0.10
      Stray light (μ5)0.27
      Total uncertainty (U)0.86
    • Table 3. Uncertainty of the SD on-board emitted spectrum radiance

      View table

      Table 3. Uncertainty of the SD on-board emitted spectrum radiance

      SourceUncertainty /%
      SD BRDF measurement (μ1)1
      SD cosine error of incidence (μ2)0.54
      SD radiance non-uniformity (μ3)0.90
      Earth-sun distance correction factor (μ4)0.20
      SD stability monitor (μ5)0.86
      Measurement of SD H factor before launch (μ6)0.50
      Solar spectral irradiance (μ7)1
      Stray light (μ8)2.72
      Total uncertainty (U)3.40
    Tools

    Get Citation

    Copy Citation Text

    Wenxin Huang, Liming Zhang, Xiaolong Si, Heyu Xu, Hongyao Chen, Xin Li, Yanhua Zhao. On-Orbit Performance Evaluation of On-Board Calibration Component of GF-5 Visible and Infrared Multispectral Imager[J]. Acta Optica Sinica, 2020, 40(20): 2029001

    Download Citation

    EndNote(RIS)BibTexPlain Text
    Save article for my favorites
    Paper Information

    Category: Scattering

    Received: May. 25, 2020

    Accepted: Jul. 3, 2020

    Published Online: Sep. 30, 2020

    The Author Email: Huang Wenxin (huangwx@aiofm.ac.cn)

    DOI:10.3788/AOS202040.2029001

    Topics