Chinese Optics Letters, Volume. 20, Issue 11, 111201(2022)

Improving the sensitivity of DC magneto-optical Kerr effect measurement to 10−7rad/Hz

Junying Ma, Feng Gu, Ying Xu, Jiaming Le, Fanlong Zeng, Yizheng Wu, and Chuanshan Tian*
Author Affiliations
  • State Key Laboratory of Surface Physics and Department of Physics, Fudan University, Shanghai 200433, China
  • show less
    Figures & Tables(5)
    (a) Sketch of the DC-MOKE setup. HWP1 and HWP2 stand for half-wave plates, and P is the polarizer. The arrows illustrate the polarization state after the optics. (b) Fluctuation of MOKE signal in 1 h after control of the temperature within ±1 mK for the laser tube and polarizing optics in a sealed environment. (c) Statistic analysis of the MOKE noise in (b) using Gaussian distribution.
    (a) MOKE signal (red line) fluctuates along with the laser intensity (blue line) as the laser tube temperature is drifting. The fine spectral feature is the fingerprints of the gain medium. (b) Top, mode structure of a red (632.8 nm) He–Ne laser. The adjacent longitudinal modes, labeled as s-mode (blue line) and p-mode (red line), are orthogonally polarized. Bottom shows that the measured intensity variances of the s-mode (blue) and p-mode (red) are out of phase in a He–Ne laser with cavity length of 25 cm. (c) The fluctuation of laser intensity after temperature control of ±1 mK for the laser tube (inset). (d) Comparison of the polarization noise with (red) and without (blue) the Brewster window.
    (a) and (b) Variation of DC-MOKE signal (red line) when modulating the temperature (blue line) of (a) the polarizer and (b) Wollaston prism, respectively. (c) Comparison of MOKE noise in sealed and unsealed condition after subtracting the drifting background.
    (a) Hysteresis loops at five different positions of a wedge-shaped Ni thin film on SiO2 substrate. (b) Noise measured at the bare SiO2 substrate.
    Noise spectrum of our MOKE apparatus measured by an SR830 lock-in amplifier.
    Tools

    Get Citation

    Copy Citation Text

    Junying Ma, Feng Gu, Ying Xu, Jiaming Le, Fanlong Zeng, Yizheng Wu, Chuanshan Tian, "Improving the sensitivity of DC magneto-optical Kerr effect measurement to 10−7rad/Hz," Chin. Opt. Lett. 20, 111201 (2022)

    Download Citation

    EndNote(RIS)BibTexPlain Text
    Save article for my favorites
    Paper Information

    Category: Instrumentation, Measurement, and Optical Sensing

    Received: Mar. 26, 2022

    Accepted: Jun. 6, 2022

    Published Online: Jul. 19, 2022

    The Author Email: Chuanshan Tian (cstian@fudan.edu.cn)

    DOI:10.3788/COL202220.111201

    Topics