Acta Optica Sinica, Volume. 40, Issue 16, 1630002(2020)

Detection and Correction of Fringe Count Error in Spaceborne Infrared Hyperspectral Fourier Transform Spectrometer

Chang Liu1,2,3, Hailiang Shi1,2,3, Zhiwei Li1,3, Chunqiang Wu4, Chengli Qi4, and Wei Xiong1,2,3、*
Author Affiliations
  • 1Anhui Institute of Optics and Fine Mechanics, Hefei Institute of Physical Science, Chinese Academy of Sciences, Hefei, Anhui 230031, China
  • 2University of Science and Technology of China, Hefei, Anhui 230026, China
  • 3Key Laboratory of Optical Calibration and Characterization of Chinese Academy of Sciences, Hefei, Anhui 230031, China
  • 4Key Laboratory of Radiometric Calibration and Validation for Environmental Satellite, National Satellite Meteorological Center, Beijing 100081, China
  • show less
    Figures & Tables(11)
    Interferogram diagram when FCE occurs
    Effect of FCE at different positions on phase
    Effect of FCE at different positions on spectrum. (a) Overall view; (b) partial enlargement
    Effects of FCE at different positions on phase difference and spectral deviation. (a) Effect on phase difference; (b) effect on spectral deviation
    Schematics of FCE detection. (a) Interference data with FCE; (b) phase extraction function; (c) phase of phase extraction function
    FCE calibration results. (a) Spectral phase of the interferogram with FCE; (b) linearly fits the phase extraction function; (c)phase contrast chart after FCE correction; (d) spectrum contrast chart after FCE correction
    FCE correction in earth scene. (a) Phase extraction function linear fitting in earth scene; (b) phase comparison before and after correction
    Test results of actual data
    • Table 1. FCE error simulation parameters

      View table

      Table 1. FCE error simulation parameters

      ParameterValue
      Spectral range /cm-1648 to 1136
      Spectral resolution /cm-10.625
      Sampling point19456
      Lost point1
      Lost positionleft edge,center,right edge
    • Table 2. Instrument related parameters

      View table

      Table 2. Instrument related parameters

      ParameterTechnical indicator
      Scan cycle /s10
      Residence steps40
      Effective residence steps33
      Scan times30
      Wave bandLongwave, midwave,shortwave
    • Table 3. Algorithm detection and correction under different FCE conditions

      View table

      Table 3. Algorithm detection and correction under different FCE conditions

      FCE appear positionCalculatedoffset pointsFitted variancePhase differencebefore correctionPhase differenceafter correction
      One point lost1.03190.2352339.228621.1
      LeftThree points lost3.03750.23051045.235.2764
      Five points lost5.17130.2348693.924428.8488
      One point lost0.93628.0109339.219521.0571
      CenterThree points lost3.03530.18531045.335.2145
      Five points lost5.16740.1694694.76728.7564
    Tools

    Get Citation

    Copy Citation Text

    Chang Liu, Hailiang Shi, Zhiwei Li, Chunqiang Wu, Chengli Qi, Wei Xiong. Detection and Correction of Fringe Count Error in Spaceborne Infrared Hyperspectral Fourier Transform Spectrometer[J]. Acta Optica Sinica, 2020, 40(16): 1630002

    Download Citation

    EndNote(RIS)BibTexPlain Text
    Save article for my favorites
    Paper Information

    Category: Spectroscopy

    Received: Apr. 20, 2020

    Accepted: May. 21, 2020

    Published Online: Aug. 7, 2020

    The Author Email: Xiong Wei (frank@aiofm.ac.cn)

    DOI:10.3788/AOS202040.1630002

    Topics