Acta Optica Sinica, Volume. 42, Issue 9, 0930002(2022)

Spectral Calibration and Matching Method for Ultraviolet-Visible Polarization Imaging Spectrometer

Jingjing Shi1,2, Mengfan Li2、*, Yadong Hu2, Gaojun Chi2, Xiangjing Wang2, and Jin Hong1,2
Author Affiliations
  • 1School of Environmental Science and Optoelectronic Technology, University of Science and Technology of China, Hefei 230026, Anhui, China
  • 2Anhui Institute of Optics and Fine Mechanics, Chinese Academy of Sciences, Hefei 230031, Anhui, China
  • show less
    Figures & Tables(17)
    Optical system of instrument
    Schematic diagram of spectral modulation module
    Field of view calibration site
    Image of parallel light acquired by detector
    Fitting results. (a) S beam; (b) P beam
    Pixel collection at different field of view. (a) S beam; (b) P beam
    Field of view calibration regression lines. (a) S beam; (b) P beam
    Schematic diagram of spectral calibration experiment
    Spectral image of mercury lamp acquired by detector
    Spectra of mercury lamp. (a) S beam; (b) P beam
    Spectral response curves of instrument at 404.66 nm of mercury lamp characteristic spectral line. (a) S beam; (b) P beam
    Spectral calibration regression lines. (a) S beam; (b) P beam
    Schematic diagram of outdoor experiment
    Solar spectrum acquired by instrument
    • Table 1. Wavelength-pixel matching relationship

      View table

      Table 1. Wavelength-pixel matching relationship

      Wavelength /nmPixel of center peak
      S beamP beam
      365.02820.79821.72
      404.66966.11966.93
      407.78977.57978.42
      435.831080.591081.42
      546.071485.681486.46
    • Table 2. Uncertainty of spectral calibrationnm

      View table

      Table 2. Uncertainty of spectral calibrationnm

      Channelσ21σ22σ23σ2
      S0.010.0270.03160.043
      P0.010.0270.03860.048
    • Table 3. Comparison of calibration value and standard value of Fraunhofer linesnm

      View table

      Table 3. Comparison of calibration value and standard value of Fraunhofer linesnm

      Absorption elementStandard wavelengthCalibration wavelengthDeviation
      S beamP beamS beamP beam
      Ca+393.37393.38393.330.010.04
      Ca+396.84396.82396.80.020.04
      410.18410.16410.10.020.08
      434.05434.09434.040.040.01
      Fe466.81466.83466.780.020.03
      486.13486.09486.120.040.01
    Tools

    Get Citation

    Copy Citation Text

    Jingjing Shi, Mengfan Li, Yadong Hu, Gaojun Chi, Xiangjing Wang, Jin Hong. Spectral Calibration and Matching Method for Ultraviolet-Visible Polarization Imaging Spectrometer[J]. Acta Optica Sinica, 2022, 42(9): 0930002

    Download Citation

    EndNote(RIS)BibTexPlain Text
    Save article for my favorites
    Paper Information

    Category: Spectroscopy

    Received: Aug. 25, 2021

    Accepted: Nov. 22, 2021

    Published Online: May. 6, 2022

    The Author Email: Li Mengfan (mfli@aiofm.ac.cn)

    DOI:10.3788/AOS202242.0930002

    Topics