Spectroscopy and Spectral Analysis, Volume. 41, Issue 2, 467(2021)

Experimental Research on Monitoring of BTX Concentration Based on Differential Optical Absorption Spectroscopy

Hai-ming ZHENG*, Xiao-peng ZHU, Shuai-shuai FENG, and Gui-hong JIA
Author Affiliations
  • Department of Mechanical Engineering, North China Electric Power University, Baoding 071003, China
  • show less
    Figures & Tables(12)
    Experimental system1: Nitrogen source; 2: PTFE tube; 3: CS200 flow control valve; 4: BTX standard fluid; 5: Silicone hose; 6: BT100-2J precision peristaltic pump; 7: Atomizer; 8: Mixing gas chamber; 9: Heating device; 10: Filter; 11: Ultraviolet enhanced xenon light source; 12: Absorption cell; 13: Exhaust gas outflow end; 14: Data acquisition system; 15: PC data processing system; 16: Waste liquid recovery tank; 17:Shell
    Schematic of smooth denoising
    Schematic of Wavelet transform denoising
    Comparison of benzene absorption cross-section
    Comparison of filtering effect
    Absorbance of gaseous benzene
    Benzene fitting curve
    Differential absorbance curve of benzene
    BTX mixed concentration inversion process(a): BTX differential absorbance; (b): BTX differential absorbance and xylene differential absorbance;(c): Differential absorbance of benzene and toluene; (d) Differential absorbance of benzene
    • Table 1. Inversion results of gaseous benzene based on integrated area method and least square method

      View table
      View in Article

      Table 1. Inversion results of gaseous benzene based on integrated area method and least square method

      气态苯标准浓度/(mg·m-3)积分面积法/(mg·m-3)误差/%最小二乘法/(mg·m-3)误差/%
      6.256.42-2.76.240.2
      12.3313.23-7.312.201.1
      22.7123.51-3.521.863.7
    • Table 2. BTX mixed gas ratio (mg·m-3)

      View table
      View in Article

      Table 2. BTX mixed gas ratio (mg·m-3)

      组号二甲苯甲苯
      15.217.628.62
      27.518.2510.52
      310.8511.7513.24
      415.2416.2513.24
      520.1222.1024.25
      625.2426.7530.30
    • Table 3. Inversion results of different concentrations mixed gases (mg·m-3)

      View table
      View in Article

      Table 3. Inversion results of different concentrations mixed gases (mg·m-3)

      项目组号
      123456
      二甲苯标准值5.217.5110.8515.2420.1225.24
      测量值5.187.5210.6745.1119.8525.20
      误差/%0.570.131.650.851.340.15
      甲苯标准值7.628.2511.7516.2522.1026.75
      测量值7.237.9811.0315.7521.2415.92
      误差/%5.113.156.123.072.533.14
      标准值8.6210.5213.2413.2424.2530.30
      测量值7.919.9512.1018.5622.0527.82
      误差/%8.245.428.615.549.078.21
    Tools

    Get Citation

    Copy Citation Text

    Hai-ming ZHENG, Xiao-peng ZHU, Shuai-shuai FENG, Gui-hong JIA. Experimental Research on Monitoring of BTX Concentration Based on Differential Optical Absorption Spectroscopy[J]. Spectroscopy and Spectral Analysis, 2021, 41(2): 467

    Download Citation

    EndNote(RIS)BibTexPlain Text
    Save article for my favorites
    Paper Information

    Category: Research Articles

    Received: Jan. 13, 2020

    Accepted: Apr. 25, 2020

    Published Online: Apr. 8, 2021

    The Author Email: Hai-ming ZHENG (haimingzheng@163.com)

    DOI:10.3964/j.issn.1000-0593(2021)02-0467-06

    Topics