Acta Optica Sinica, Volume. 40, Issue 7, 0712003(2020)

Polarization Detection Performance of Directional Polarimetric Camera

Bihai Tu1,2,3、*, Jin Hong1,3、**, Pingping Yao1,3, Binhuan Meng1,3, Yinlin Yuan1,3, Miaomiao Zhang1,2,3, and Jianwen Weng1,2,3
Author Affiliations
  • 1Anhui Institute of Optics and Fine Mechanics, Chinese Academy of Sciences, Hefei, Anhui 230031, China
  • 2University of Science and Technology of China, Hefei, Anhui 230026, China
  • 3Key Laboratory of Optical Calibration and Characterization, Chinese Academy of Sciences, Hefei, Anhui 230031, China
  • show less
    Figures & Tables(24)
    Diagram of instrument system composition
    Diagram of on-orbit work
    Schematic diagram of DPC imaging system
    Schematic diagram of optical system
    865 nm polarized effect of lens
    910 nm polarized effect of lens
    Schematic diagram of performance measurement experiment setup
    Measurement range of experiment
    Original image of DPC
    DOLP image of DPC
    Flow chart of data processing
    Diffusion function curve
    2D diffusion function diagram
    Global coefficient matrix data. (a) Region (6,6); (b) region (7,8)
    Stray light correction. (a) RGB image; (b) corrected data at wavelength of 865 nm; (c) ratio of stray light signal to effective data
    DPC data preview
    Polarization change curve of sunglint
    Angle sensitivity curve
    Data at wavelength of 865 nm. (a) Radiation intensity; (b) degree of polarization; (c) sunglint angle; (d) scattering angle
    Spectral response curve of 865 nm channel
    Polarization analysis of sunglint. (a) Simulation data; (b) DPC data; (c) deviation distribution; (d) polarization comparison scatter plot
    • Table 1. Preset parameters of light source

      View table

      Table 1. Preset parameters of light source

      Waveband /nmDOLP of different positions
      P1P2P3P4P5P6P7P8P9
      49000.10290.15280.20370.25450.30530.40630.50710.6073
      67000.10000.15000.20000.25000.30000.40000.50000.6000
      86500.09990.14850.19810.24760.29730.39660.49620.5961
    • Table 2. Measured value at wavelength of 865 nm

      View table

      Table 2. Measured value at wavelength of 865 nm

      Preset DOLPMeasured value under different field-of-view angles
      45°30°15°-15°-30°-45°
      00.00360.00270.00450.00390.00340.00480.0063
      0.09990.09750.09630.09780.09540.09540.09540.0984
      0.14850.14600.14730.14560.14520.14880.14830.1477
      0.19810.19520.19730.19630.19300.19810.19750.1948
      0.24760.24780.24670.24600.24400.24750.24680.2480
      0.29730.29470.29730.29600.29310.29690.29550.2951
      0.39660.39500.39600.39560.39340.39610.39660.3954
      0.49620.49370.49520.49460.49350.49570.49580.4950
      0.59610.59500.59720.59530.59470.59630.59550.5938
    • Table 3. Measured value at wavelength of 865 nm

      View table

      Table 3. Measured value at wavelength of 865 nm

      DOLPMeasured value under different zenith angles
      -55°-50°-45°-40°-35°-30°-25°
      By DPC0.19960.23750.28560.31040.32660.35260.3401
      By CE3180.20880.24890.29040.32110.34010.35710.3548
      P|0.00920.01140.00480.01070.01350.00450.0147
      DOLPMeasured value under different zenith angles
      -20°-15°-10°-5°10°
      By DPC0.34170.32150.30560.27460.24810.2110.1791
      By CE3180.35140.33820.31040.28220.24830.21140.1796
      P|0.00970.01670.00480.00760.00020.00040.0005
    Tools

    Get Citation

    Copy Citation Text

    Bihai Tu, Jin Hong, Pingping Yao, Binhuan Meng, Yinlin Yuan, Miaomiao Zhang, Jianwen Weng. Polarization Detection Performance of Directional Polarimetric Camera[J]. Acta Optica Sinica, 2020, 40(7): 0712003

    Download Citation

    EndNote(RIS)BibTexPlain Text
    Save article for my favorites
    Paper Information

    Category: Instrumentation, Measurement and Metrology

    Received: May. 27, 2019

    Accepted: Dec. 12, 2019

    Published Online: Apr. 15, 2020

    The Author Email: Tu Bihai (tubihai@aiofm.ac.cn), Hong Jin (hongjin@aiofm.ac.cn)

    DOI:10.3788/AOS202040.0712003