Chinese Journal of Lasers, Volume. 50, Issue 9, 0907104(2023)

Polarization Spatial Frequency Domain Imaging System and Phantom Verification

Zhiguo Shi1,2, Yang Zhang2, Quanfu Wang2, Zhongsheng Li2, Xia Wang2, Meili Dong2, Jingshu Ni2, Yao Huang2, Shengzhao Zhang1, Yikun Wang2, and Yuanzhi Zhang2、*
Author Affiliations
  • 1Department of Biomedical Engineering, Anhui Medical University, Hefei 230009, Anhui, China
  • 2Anhui Provincial Engineering Technology Research Center for Biomedical Optical Instrument, Anhui Provincial Engineering Laboratory for Medical Optical Diagnosis&Treatment Technology and Instrument, Anhui Institute of Optics and Fine Mechanics, Hefei Institutes of Physical Science, Chinese Academy of Sciences, Hefei 230031, Anhui, China
  • show less
    Figures & Tables(9)
    Polarization spatial frequency domain imaging system. (a) Schematic; (b) physical drawing
    System operation flow chart
    Relationship between m11 and reflectance of gradient diffuse reflector
    Schematic of decomposition experiment verification
    Polarization decomposition validation test results. (a) Experimental results of gradient fat emulsion solution; (b) experimental results of gradient glass sheet angle; (c) comparative experimental results of full wave plate and quarter wave plate
    Experimental results of polarization spatial frequency domain imaging and traditional polarized light imaging. (a) Experimental results for placing diffuse reflection plate with 0% reflectance at the bottom of fat emulsion solution; (b) experimental results for placing diffuse reflection plate with 100% reflectance at the bottom of fat emulsion solution
    Burn test results. (a) Burn photos of pigskin, sorted by heating time; (b) curves of depolarization coefficient and diattenuation coefficient; (c) phase delay curve
    • Table 1. Parameter setting in gradient fat emulsion solution experiment

      View table

      Table 1. Parameter setting in gradient fat emulsion solution experiment

      No.Volume fraction of fatemulsion solution /%Glass sheet angle /(°)Wave plate
      10.04120π/2
      20.08120π/2
      30.12120π/2
      40.16120π/2
      50.20120π/2
      60.24120π/2
      70.40120π/2
    • Table 2. Parameter setting in gradient glass sheet angle test

      View table

      Table 2. Parameter setting in gradient glass sheet angle test

      No.Volume fraction of fatemulsion solution /%Glass sheetangle /(°)Wave plate
      10.260π/2
      20.275π/2
      30.290π/2
      40.2105π/2
      50.2120π/2
    Tools

    Get Citation

    Copy Citation Text

    Zhiguo Shi, Yang Zhang, Quanfu Wang, Zhongsheng Li, Xia Wang, Meili Dong, Jingshu Ni, Yao Huang, Shengzhao Zhang, Yikun Wang, Yuanzhi Zhang. Polarization Spatial Frequency Domain Imaging System and Phantom Verification[J]. Chinese Journal of Lasers, 2023, 50(9): 0907104

    Download Citation

    EndNote(RIS)BibTexPlain Text
    Save article for my favorites
    Paper Information

    Category: Biomedical Optical Imaging

    Received: Dec. 5, 2022

    Accepted: Feb. 7, 2023

    Published Online: Apr. 24, 2023

    The Author Email: Yuanzhi Zhang (yzzhang@aiofm.ac.cn)

    DOI:10.3788/CJL221483

    Topics