Acta Optica Sinica, Volume. 38, Issue 11, 1130002(2018)

Wavelength Calibration of Infrared Spectrometer Based on Hollow Planar Waveguide

Yue Wu1,2,3、*, Jiaxiang Liu1,3, Yonghua Fang1,3、*, and Min Yang1,2,3
Author Affiliations
  • 1 Anhui Institute of Optics and Fine Mechanics, Chinese Academy of Sciences, Hefei, Anhui 230031, China
  • 2 University of Science and Technology of China, Heifei, Anhui 230026, China
  • 3 Key Laboratory of Optical Calibration and Characterization, Chinese Academy of Sciences, Hefei, Anhui 230031, China
  • show less
    Figures & Tables(11)
    Structural schematics of the hollow planar waveguide infrared spectrometer. (a) Schematic of beam propagation in space; (b) scheme of optical path design
    Flow chart of laboratory wavelength calibration
    Sketch map of laboratory wavelength calibration
    Physical map of wavelength calibration device in laboratory
    Relation of pixel and response at different wavelengths
    Fitting curves of wavelength calibration
    Fitting residual of wavelength calibration
    • Table 1. Specific parameters of spectrometer

      View table

      Table 1. Specific parameters of spectrometer

      ParameterValue
      Wavelength range /μm8-12
      Spectral sampling interval /μm0.0625
      Numerical aperture0.22
      Detector array size64×1
      Detector pixel size /(mm×mm)0.5×1.5
    • Table 2. Parameters of PL2-M CO2 laser

      View table

      Table 2. Parameters of PL2-M CO2 laser

      ParameterValue
      Wavelength range /μm9.1-10.9
      Mode>90% TEM00, M2<1.25
      Amplitude stability±1% short term (unstabilized)
      Frequency stability /(kHz·s-1)200
      Beam diameter (1/e2)4.8 mm@output coupler
      Beam divergence /mrad6.5
    • Table 3. Results of wavelength calibration

      View table

      Table 3. Results of wavelength calibration

      Central wavelength /μmPixelSpectral resolution /nm
      9.26148.925139
      9.48845.925144
      9.62144.822137
      10.23337.558134
      10.47634.535129
      10.76430.828135
    • Table 4. Calculation results of two fitting methods

      View table

      Table 4. Calculation results of two fitting methods

      Central Wavelength /μmPixel1st order polynomial fitting3rd order polynomial fitting
      Fitted value /μmError /μmFitted value /μmError /μm
      8.5060.558.284-0.2168.461-0.039
      11.5213.6012.2280.70711.415-0.100
    Tools

    Get Citation

    Copy Citation Text

    Yue Wu, Jiaxiang Liu, Yonghua Fang, Min Yang. Wavelength Calibration of Infrared Spectrometer Based on Hollow Planar Waveguide[J]. Acta Optica Sinica, 2018, 38(11): 1130002