Chinese Optics, Volume. 15, Issue 4, 825(2022)

Spectral irradiance degradation model of halogen tungsten lamps at wavelength from 400 nm to 1300 nm

Ya-chao ZHANG, Xin YE*, Zhi-wei XIA, Long SUI, and Wei FANG
Author Affiliations
  • Changchun Institute of Optics, Fine Mechanics and Physics, Chinese Academy of Sciences, Changchun 130033, China
  • show less

    Spectral irradiance degradation of a halogen tungsten lamp increases the spectral radiance uncertainty of an on-board lamp-diffuser calibrator that is composed of a lamp and a diffuser reflector. Therefore, it is necessary to investigate the degradation characteristics of the lamp to decrease the spectral radiance uncertainty. A spectral irradiance degradation model of a halogen tungsten lamp with an undetermined order at wavelengths from 400 nm to 1300 nm was proposed according to the blackbody radiation law and Weierstrass theorem. Then, the spectral irradiance degradation curve of the halogen tungsten lamp was experimentally measured, and it was fitted by different order models using the least squares method, respectively. The model order was determined as 2, which is the minimum order to satisfy the reconstruction accuracy required by the spectral radiance of the on-board lamp-diffuser calibrator. The reconstruction accuracy of the spectral irradiance degradation curve was better than 0.25% according to this two-order model, which lays a theoretical foundation to decrease the spectral radiance uncertainty of the on-board lamp-diffuser calibrator.

    Keywords

    1 Introduction

    In the 21st century, climate change has became a topic of global concern. Global climate change research has put forward unprecedented requirements for the radiation measurement uncertainty of remote sensing satellites. The ASIC3 (Achieving Satellite International Corporation for Climate Change) states that the measurement uncertainty of the remote sensing satellites has to remain better than 0.3% at the reflected solar wavebands to accurately predict global climate change[1].

    However, the radiometric calibration of the satellite optical remote sensors at reflected solar wavebands is usually performed by the standard halogen tungsten lamp, the solar-diffuser calibrator or vicarious calibration. At present, the minimum and maximum spectral irradiance uncertainty of standard halogen tungsten lamp is 0.47% at wavebands of 1600 nm (k=2) and 1.16% at wavebands of 400 nm (k=2) during its lifetime[2]. Therefore, standard halogen tungsten lamps cannot satisfy a measurement uncertainty of 1% (k=2) at whole reflected solar wavebands. Moreover, radiometric calibration uncertainty is only about 5% at reflected solar wavebands by solar-diffuser calibrator and vicarious calibration[3-7]. Therefore, the current radiometric calibration methods hardly satisfy the high-precision radiometric calibration requirements in the field of climate research.

    An alternative method to improve the accuracy of radiometric calibration is to trace the spectral radiance of the lamp-diffuser calibrator to the on-board space cryogenic absolute radiometer. However, it is impossible to do this at each operating waveband of the remote sensor in orbit. Therefore, it is necessary to reconstruct the spectral radiance of the entire operating waveband range from the spectral radiances of several operating wavebands. In addition, the reconstruction uncertainty must be better than 0.3% to realize the target measurement uncertainty of 1% of reflected solar spectral radiation.

    At present, the spectral radiance (irradiance) of the light source is reconstructed according to its spectral radiance (irradiance) model. For example, the spectral irradiance of a halogen tungsten lamp can be reconstructed with an uncertainty of approximately 0.25% according to its spectral irradiance model[8]. However, the spectral radiance of the lamp-diffuser calibrator cannot be described by a simple but precise model, as such a model has not yet been reconstructed for the diffuser reflector. Therefore, it is difficult to reconstruct the spectral radiance of the lamp-diffuser calibrator with an uncertainty of 0.3% using the present reconstruction method.

    It is well known that the degradation characteristics of the diffuser reflectors in the Moderate Resolution Imaging Spectroradiometer (MODIS) and Visible Infrared Imaging Radiometer Suite (VIIRS) can be described by an analytical model[9]. Therefore, it is reasonable to believe that the degradation characteristics of the other on-board diffuser reflectors can also be described in this way. The spectral radiance degradation characteristics of the lamp-diffuser calibrator can correspondingly be described by an analytical model if the spectral irradiance degradation characteristics of the halogen tungsten lamp can be described by an analytical model. In addition, the spectral radiance degradation curve of the lamp-diffuser calibrator can be reconstructed with high accuracy according to its spectral radiance degradation model. The high-precision spectral radiance of the lamp-diffuser calibrator can subsequently be calculated by the product of the reconstructed spectral radiance degradation curve and the spectral radiance calibrated before launch.

    In summary, it is necessary to research the high-precision spectral irradiance degradation model of a halogen tungsten lamp to reconstruct the spectral radiance of the on-board lamp-diffuser calibrator with an uncertainty better than 0.3%. In this paper, the following work was performed to construct the spectral irradiance degradation model of a halogen tungsten lamp from the Chinese Radiometric Benchmark of Reflected Solar Band project. First, the spectral irradiance degradation model of a halogen tungsten lamp with undetermined order was derived according to the blackbody radiation law and Weierstrass theorem in Section 2. Next, the spectral irradiance degradation curve was measured and the criterion to determine the model order was given, according to which the model order was determined in Section 3. The conclusion is presented in Section 4.

    2 Derivation of spectral irradiance degradation model

    Because the spectral irradiance degradation curve of a halogen tungsten lamp is the ratio of its spectral irradiances at different moments, the spectral irradiance degradation model is closely related to its spectral irradiance model. The physical model of the halogen tungsten lamp irradiance is expressed in Eq.(1):

    $ E\left( {\lambda ,T} \right) = B\left( T \right) \cdot {\varepsilon _{\text{W}}}\left( {\lambda ,T} \right) \cdot {\varepsilon _\Delta }\left( \lambda \right) \cdot \frac{{2\hbar{c^2}}}{{{\lambda ^5}\left( {\exp \left( {\dfrac{{\hbar c}}{{\lambda kT}}} \right) - 1} \right)}} , $ (1)

    where $ B\left( T \right) $ is a geometrical factor of the lamp filament that takes into account the measurement distance and the dimensions of the filament; $ \lambda $ is the wavelength in a vacuum; $ T $ is the temperature of the filament; $ {\varepsilon _{\text{W}}}\left( {\lambda ,T} \right) $ is the nominal spectral emissivity of tungsten; $ {\varepsilon _\Delta }\left( \lambda \right) $ is the spectral emissivity correction factor for the emissivity of the lamp; $ \hbar $ is the Planck constant; $ c $ is the velocity of light in a vacuum; and $ k $ is the Boltzmann constant[10]. The residual correction factor accounts for the effects of all factors in addition to the nominal spectral emissivity of tungsten. These include the transmittance of the quartz bulb, the transmittance of the filling gas, the difference in the properties of tungsten used in the lamp filament and in the nominal emissivity determination, and the light recycling effect in the coiled filament[11-12].

    In this paper, the product of the geometric factor $ B\left( T \right) $, the nominal spectral emissivity of tungsten $ {\varepsilon _{\text{W}}}\left( {\lambda ,T} \right) $, and the spectral emissivity correction factor $ {\varepsilon _\Delta }\left( \lambda \right) $ is referred to as the spectral emissivity of the halogen tungsten lamp $ \varepsilon \left( {\lambda ,T} \right) $. Therefore, the spectral irradiance of the halogen tungsten lamp can be expressed as the product of the Planck function $ H\left( {\lambda ,T} \right) $ and the spectral emissivity, as shown in Eq.(2):

    $ E\left( {\lambda ,T} \right) = \frac{{2\hbar {c^2}}}{{{\lambda ^5}\left( {\exp \left( {\dfrac{{\hbar c}}{{\lambda kT}}} \right) - 1} \right)}} \cdot \varepsilon \left( {\lambda ,T} \right)\quad. $ (2)

    If the increments of the Planck function and spectral emissivity are expressed as $ \Delta H\left( {\lambda ,T} \right) $ and $ \Delta \varepsilon \left( {\lambda ,T} \right) $, respectively, when the filament temperature changes from $T$ to $ T{\text{ + }}\Delta T $, the spectral irradiance degradation model of the halogen tungsten lamp can be expressed as Eq. (3):

    $ \zeta \left( {\lambda ,T} \right) = \frac{{H\left( {\lambda ,T} \right) + \Delta H\left( {\lambda ,T} \right)}}{{H\left( {\lambda ,T} \right)}} \cdot \frac{{\varepsilon \left( {\lambda ,T} \right){\text{ + }}\Delta \varepsilon \left( {\lambda ,T} \right)}}{{\varepsilon \left( {\lambda ,T} \right)}} . $ (3)

    When $ \lambda $ is less than 1300 nm, the Planck function $ H\left( {\lambda ,T} \right) $ is reduced to Eq. (4) as $ \exp \left( {\dfrac{{{C_2}}}{{\lambda T}}} \right) \gg 1 $ is satisfied:

    $ H\left( {\lambda ,T} \right) = \frac{{{C_1}}}{{{\lambda ^5}}} \cdot \exp \left( { - \frac{{{C_2}}}{{\lambda T}}} \right) \quad, $ (4)

    where ${C_1}$ and ${C_2}$ are the first and second radiation constants, respectively. The increment of the Planck function can be expressed as Eq. (5) derived from Eq. (4) by the differential operation.

    $ \begin{split} \Delta H\left( {\lambda ,T} \right) =& \frac{{{C_2} \cdot {\raise0.7ex\hbox{${\Delta T}$} \mathord{\left/ {\vphantom {{\Delta T} {{T^2}}}}\right.} \lower0.7ex\hbox{${{T^2}}$}}}}{\lambda } \cdot \frac{{{C_1}}}{{{\lambda ^5}}} \cdot \exp \left( { - \frac{{{C_2}}}{{\lambda T}}} \right)= \\ & \frac{{{C_2} \cdot {\raise0.7ex\hbox{${\Delta T}$} \mathord{\left/ {\vphantom {{\Delta T} {{T^2}}}}\right.} \lower0.7ex\hbox{${{T^2}}$}}}}{\lambda } \cdot H\left( {\lambda ,T} \right)\quad. \\ \end{split} $ (5)

    Because the spectral emissivity degradation model of a halogen tungsten lamp is affected by the spectral emissivity of tungsten, the spectral transmittance of the quartz lamp shell, the spectral transmittance of the halogen gas, and the filament morphology, it is difficult to achieve the analytical expression through theoretical derivation. However, any continuous function can be approximated by an algebraic polynomial function according to the Weierstrass theorem. Therefore, the spectral emissivity degradation model of a halogen tungsten lamp can be approximated by the polynomial function shown in Eq. (6):

    $ \frac{{\varepsilon \left( {\lambda ,T} \right){\text{ + }}\Delta \varepsilon \left( {\lambda ,T} \right)}}{{\varepsilon \left( {\lambda ,T} \right)}} = \sum\limits_{i = 0}^n {{a_i}{\lambda ^i}} \quad, $ (6)

    where $ {a_i} $ is an undetermined coefficient. Therefore, the spectral irradiance degradation model of the halogen tungsten lamp can be expressed as Eq. (7) according to Eqs.(4)−(6).

    $ \begin{split} &\zeta \left( {\lambda ,T} \right) =\\& \frac{{H\left( {\lambda ,T} \right) + \Delta H\left( {\lambda ,T} \right)}}{{H\left( {\lambda ,T} \right)}} \cdot \frac{{\varepsilon \left( {\lambda ,T} \right){\text{ + }}\Delta \varepsilon \left( {\lambda ,T} \right)}}{{\varepsilon \left( {\lambda ,T} \right)}} = \\ &\left( {1 + \frac{{{C_2} \cdot {{\Delta T}/{{T^2}}}}}{\lambda }} \right) \cdot \left( {{a_0} + {a_1}\lambda + {a_2}{\lambda ^2} + {a_3}{\lambda ^3} + \cdots } \right) \quad. \end{split} $ (7)

    Since the temperature $T$ and temperature difference $\Delta T$ were contained in the undetermined coefficients ${\xi _0},{\xi _1},{\eta _1},{\eta _2},{\eta _3} \cdots$, Eq.(7) can be simplified to be Eq.(8).

    $ \zeta \left( \lambda \right)\;{\text{ = }}{\xi _0} + {\xi _1}/\lambda + {\eta _1}\lambda + {\eta _2}{\lambda ^2} + {\eta _3}{\lambda ^3} + \cdots $ (8)

    There may be several different orders of the model that satisfy the reconstruction uncertainty required by the spectral radiance of the lamp-diffuser calibrator. However, the model order should be determined as the lowest order that satisfies the spectral radiance reconstruction uncertainty of the on-board lamp-diffuser calibrator, which is referred to as the model-order-determination criterion in this study. The lowest-order model requires the fewest wavebands to reconstruct from the spectral radiance of the on-board lamp-diffuser calibrator, which incurs the lowest cost to the on-board spectral radiometric calibration. Therefore, the model order of Eq.(7) should be determined experimentally.

    3 Model order determined by experiment method

    3.1 Measurement of the spectral irradiance degradation curve

    Because the spectral irradiance degradation characteristics of halogen tungsten lamps are closely related to their manufacturing process, it is reasonable to believe that halogen spectral irradiance degradation characteristics of the same type of the halogen tungsten lamp can be described by the same analytic model. In other words, the model order of one type of halogen tungsten lamp can be determined from the spectral irradiance degradation curves of samples. The spectral irradiance degradation curve of the halogen tungsten lamp is the ratio of its spectral irradiances at different moments, which can be expressed by Eq.(9) where $ {E_o}\left( \lambda \right) $ and $ {E_j}\left( \lambda \right) $ are the initial and the spectral irradiance after the jth aging interval, respectively.

    $ {\zeta _j}\left( \lambda \right) = \frac{{E_j}\left( \lambda \right)}{{E_o}\left( \lambda \right)}\quad. $ (9)

    Therefore, the spectral irradiance degradation curve can be measured by the scheme shown in Fig.1, which is composed of a halogen tungsten lamp, diffuser reflection plate, and spectrometer. Two Osram 64610HLX lamps have been measured to investigate the spectral irradiance degradation model of the halogen tungsten lamp, whose rated voltage and power are 12 V and 50 V, respectively. Therefore, the operating current at which the halogen tungsten lamp operates is set to a constant 4 amperes which is the same as the operating current of the halogen tungsten lamp for on-board calibration during the test.

    Measurement scheme of the spectral irradiance degradation curve of the halogen tungsten lamp

    Figure 1.Measurement scheme of the spectral irradiance degradation curve of the halogen tungsten lamp

    The diffuser plate was made of spectralon (one kind of PTFE supplied with Labsphere). The spectral reflectance degradation of the spectralon diffuser was caused by exposure to radiation at wavebands of 200 nm−400 nm according to J.E.Leland[13]. Fig.2 shows the spectral irradiance of the tested halogen tungsten lamp and the extra-atmospheric solar spectral irradiance calculated by MODTRAN[14], the ratio of which at wavebands of 250 nm to 400 nm was shown in Fig.3. According to Fig.3, the averaged ratio at wavebands of 250−400 nm was calculated to be about 0.86%. Moreover, the radiation of the halogen tungsten lamp at wavebands of 200−250 nm was so low that it can be neglected. Therefore, the diffuser equals to be exposure to the extra-atmospheric solar radiation for about one hour when it is irradiated by the halogen tungsten lamp for 160 hours. According to the technical guide of Labsphere, the reflectance of the diffuser decreases at wavelength of 250 nm to about 0.04%, when the irradiation is equivalent to one extra-hour of atmospheric solar irradiation. Moreover, the diffuser reflectance returned to near original values when it returned to atmospheric conditions, presumably due to oxidation and the loss of the surface contaminants that caused the discoloration[15]. Therefore, it was reasonable to infer that the reflectance degradation of the diffuser reflector was much smaller than 0.04%, which was much smaller than the degradation of the halogen tungsten lamp. That is to say that the reflectance degradation of the diffuser reflector can be neglected in researching the degradation of the halogen tungsten lamp.

    (a) Extra-atmospheric solar spectral irradiance and (b) spectral irradiance of tested halogen tungsten lamp

    Figure 2.(a) Extra-atmospheric solar spectral irradiance and (b) spectral irradiance of tested halogen tungsten lamp

    Ratio of the spectral irradiance of the tested halogen tungsten lamp to the sun

    Figure 3.Ratio of the spectral irradiance of the tested halogen tungsten lamp to the sun

    The spectrometer was an HR-1024i, supplied by SVC (Spectra Vista Corporation, US), that operates at wavebands of 350−2500 nm and has 1024 spectral channels. The detector assembly of the HR-1024i spectrometer was composed of three linear array detectors: a 512 CCD detector for wavebands from 350−990 nm, a 256 InGaAs detector for wavebands from 990−1900 nm and a 256 extended InGaAs detector for wavebands from 1900−2500 nm. The temperature stability of the laboratory and the three detectors in the spectrometer were kept within $ \pm {1^{ \circ \,}}{\text{C}} $ and $ \pm {0.1^{ \circ \,}}{\text{C}} $, respectively. Besides, the lamp, the spectrometer and the diffuser plate had never been moved or replaced during the 7-day measurement. Therefore, the responsivity of the HR-1024i spectrometer can be assumed to remain unchanged during the 7-day measurement.

    Additionally, the dark current had been removed from the measuring results during the measurement, which means that the spectral irradiance decay curve only contains the random noise. The method to remove dark current from the measuring results was shown in the following. First, the dark current is measured by closing the shutter, and then, the reflected light from the diffuser is measured by opening the shutter. The dark current is removed by reducing the closed shutter measurement results from the final opened shutter measurement results. To make the two unused new lamps remain stable, they were aged for 20 hours in the same experimental setting before the measurement.

    To sum up, the spectral irradiance degradation curve shown in Eq. (9) can be simplified as Eq.(10), where $ D{N_o}\left( \lambda \right) $ and $ D{N_j}\left( \lambda \right) $ are the initial and jth output of the spectrometer with the dark current removed, respectively.

    $ {\zeta _j}\left( \lambda \right) = \frac{D{N_j}\left( \lambda \right)}{D{N_o}\left( \lambda \right)} \quad. $ (10)

    According to Eq.(10), the measurement uncertainty of the spectral irradiance degradation curve is simplified to its measurement repeatability, when the measurement setup is not touched and the dark current is removed during the measurement. The calculation of the measurement uncertainty is shown in chapter 3.2 in this paper.

    To simulate the on-orbit operation, the lamp is turned on for one hour and then turned off for five minutes, and the above procedure is repeated 160 times during the measurement to ensure the lamp can satisfy the on-orbit calibration requirements. The measured spectral irradiance degradation curves of the two halogen tungsten lamps are shown in Fig.4 (Color online).

    Spectral irradiance degradation curves of (a) 1# and (b) 2# halogen tungsten lamps

    Figure 4.Spectral irradiance degradation curves of (a) 1# and (b) 2# halogen tungsten lamps

    3.2 Measurement uncertainty of the spectral irradiance degradation curve

    According to Eq.(10), the measurement uncertainty of the spectral irradiance degradation curve can be expressed by Eq. (11).

    $ \frac{{\Delta {\zeta _j}\left( \lambda \right)}}{{{\zeta _j}\left( \lambda \right)}} = \sqrt {{{\left( {\frac{{\Delta D{N_j}\left( \lambda \right)}}{{D{N_j}\left( \lambda \right)}}} \right)}^2} + {{\left( {\frac{{\Delta D{N_o}\left( \lambda \right)}}{{D{N_o}\left( \lambda \right)}}} \right)}^2}} \quad, $ (11)

    where $ \dfrac{{\Delta D{N_o}\left( \lambda \right)}}{{D{N_o}\left( \lambda \right)}} $ and $ \dfrac{{\Delta D{N_j}\left( \lambda \right)}}{{D{N_j}\left( \lambda \right)}} $ are the repeatability of the initial and jth spectral irradiance curves, respectively. During the measurement, the measurement repeatability is assumed to remain unchanged, and is thus expressed by Eq. (12) because the temperatures of the laboratory and the detector remain stable.

    $ \frac{{\Delta DN\left( \lambda \right)}}{{DN\left( \lambda \right)}} = \frac{{\Delta D{N_o}\left( \lambda \right)}}{{D{N_o}\left( \lambda \right)}} = \frac{{\Delta D{N_j}\left( \lambda \right)}}{{D{N_j}\left( \lambda \right)}} \quad. $ (12)

    Therefore, the measurement uncertainty of the spectral irradiance degradation curve is simplified to Eq. (13):

    $ \frac{{\Delta {\zeta _j}\left( \lambda \right)}}{{{\zeta _j}\left( \lambda \right)}} = \sqrt 2 \frac{{\Delta DN\left( \lambda \right)}}{{DN\left( \lambda \right)}} \quad. $ (13)

    Eq.(13) shows that the uncertainty of the spectral irradiance degradation curve can be determined by the measurement repeatability of DN that stands for the spectral irradiance. However, the repeatability of the DN is determined by the spectral irradiance of the light source and the signal to noise ratio of the HR-1024i spectrometer, which can be calculated according to the measured spectral irradiance expressed by DN. To summarize, the measurement uncertainty of the spectral irradiance degradation curve can be calculated by the measured spectral irradiance according to Eq.(13).

    The measurement uncertainty of the spectral irradiance degradation curve shown in Fig.4 was calculated and shown in Fig.5 (Color online) according to Eq.(13). The standard deviations of the measurement uncertainty of the two lamps were calculated to be approximately 0.038% according to the measurement uncertainty curves shown in Fig.5, both of which were significantly less than the spectral radiance reconstruction uncertainty of 0.3% required by the on-board lamp-diffuser calibrator. Therefore, it is reasonable to construct a spectral irradiance degradation model that satisfies the spectral radiance reconstruction uncertainty of the on-board lamp-diffuser calibrator according to the measured spectral irradiance degradation curve.

    Measurement uncertainty of the spectral irradiance degradation curve for (a) 1# and (b) 2# halogen tungsten lamps

    Figure 5.Measurement uncertainty of the spectral irradiance degradation curve for (a) 1# and (b) 2# halogen tungsten lamps

    3.3 Determination of the model order

    The spectral irradiance degradation curves of the two lamps were fitted by Eq. (7) with the model order ranging from one to four using the least-squares method. Fig.6 showed the measured spectral irradiance degradation curves and fitted results with the two-order model of the two halogen tungsten lamps. Because the fitting results of the models almost coincide, only the fitting results of the second-order model are shown in Fig.6 as an example. The relative standard deviations are expressed in Eq.(14):

    Measured spectral irradiance degradation curves and fitted results with the second-order model of (a) 1# and (b) 2# halogen tungsten lamps

    Figure 6.Measured spectral irradiance degradation curves and fitted results with the second-order model of (a) 1# and (b) 2# halogen tungsten lamps

    $ {\sigma _\nu } = \left[ \sum\limits_{i = 1}^N \frac{\left\{ \dfrac{\left[ {{\zeta _f}\left( {{\lambda _i}} \right) - {\zeta _m}\left( {{\lambda _i}} \right)} \right]}{{\zeta _m}\left( {{\lambda _i}} \right)} \right\}^2}{( {N - n - 1})} \right]^{1/2} , $ (14)

    where $ {\zeta _f} $, $ {\zeta _m} $, $N$ and $n + 1$ are the fitted data, measured data, number of measuring wavebands, and number of fitting parameters, respectively. The relative standard deviations were calculated and is shown in Fig.7 (Color online) according to Eq.(14).

    Relative standard deviations of the spectral irradiance degradation curves of (a) 1# and (b) 2# halogen tungsten lamps

    Figure 7.Relative standard deviations of the spectral irradiance degradation curves of (a) 1# and (b) 2# halogen tungsten lamps

    Fig.7(a) shows that the relative standard deviations of the first-order model increase from about 0.03% to about 0.11% and from about 0.03% to about 0.04% for the model orders from two to four during the burning time for 1# lamp. Fig.7(b) shows that the relative standard deviations of the first-order model increase from about 0.02% to about 0.16% and from about 0.02% to about 0.04% for the model orders from two to four during the burning time for 2# lamp. That is to say that the relative standard deviation of the first-order model increases much faster than the higher order model, which indicates that first-order model cannot describe the degradation characterization of the halogen tungsten lamp accurately enough during the burning life.

    Moreover, Fig.7(a) and Fig.7(b) show that the relative standard deviations are better than 0.05% and very similar for the model orders from two to four during the burning time for the two lamps. Therefore, the model order of the two halogen tungsten lamps are determined to be two since it obtains the least fitting parameters in this paper.

    The model precisions are estimated by relative errors between the fitted and measured spectral irradiance degradation curve. The relative errors of the two lamps are calculated to be better than 0.25% at wavebands from 400 nm to 1300 nm as shown in Fig.8(a) (Color online) and Fig.8(b) (Color online), respectively. Therefore, the proposed spectral irradiance degradation model of halogen tungsten lamps can satisfy the spectral radiance reconstruction uncertainty of the on-board lamp–diffuser calibrator.

    Relative errors between the fitted and measured spectral irradiance degradation curve of the two halogen tungsten lamps fitted by the second-order model. (a) 1# and (b) 2# halogen tungsten lamp

    Figure 8.Relative errors between the fitted and measured spectral irradiance degradation curve of the two halogen tungsten lamps fitted by the second-order model. (a) 1# and (b) 2# halogen tungsten lamp

    Hence, the spectral irradiance degradation model of the two halogen tungsten lamps can be expressed by Eq. (15):

    $ \zeta \left( \lambda \right) = {\xi _0} + {\xi _1}/\lambda + {\eta _1}\lambda + {\eta _2}{\lambda ^2} . $ (15)

    The proposed spectral irradiance degradation model may not be suitable for other types of lamps; however, the method used to construct the spectral irradiance degradation model is still valid for other types of halogen tungsten lamp.

    4 Conclusions

    In this paper, a hemi-empirical spectral irradiance degradation model of a halogen tungsten lamp with an undetermined model order at wavelengths from 400 nm to 1300 nm was derived based on the blackbody radiation law and the Weierstrass theorem, and the model order was subsequently determined to be two by the proposed model-order-determination criterion. The proposed spectral irradiance degradation model may not be suitable for other types of lamps. However, the method used to construct the spectral irradiance degradation model is still valid for other types of halogen tungsten lamps. The uncertainty of the proposed model is approximately 0.25%, which satisfies the spectral radiance reconstruction uncertainty of the on-board lamp-diffuser calibrator at wavelengths from 400 nm to 1300 nm and lays a theoretical foundation to realize a measurement uncertainty of 0.3% at reflected solar wavebands. The spectral irradiance degradation model of halogen tungsten lamps at wavelengths from 1300 nm to 2350 nm will be researched, subsequently.

    Acknowledgements

    The author thanks Dr. Lei kai-chao from Chang Chun Institute of Optics, Fine Mechanics and Physics, Chinese Academy of Sciences, for his valuable suggestions and selfless help that have helped improve this manuscript substantially.

    [1] LU N M, DING L, ZHENG X B, et al. Introduction of the radiometric benchmark satellite being developed in China for remote sensing[J]. Journal of Remote Sensing, 24, 672-680.(Chinese)(2020).

    [2] [2] YOON H W, GIBSON C E. NIST measurement services: spectral irradiance calibrations[R]. Gaithersburg: Optical Technology Division, Physical Measurement Labaty, National Institute of Stards Technology, 2011.

    [3] ESPOSITO J A, XIONG X X, WU A SH, et al. MODIS reflective solar bands uncertainty analysis[J]. Proceedings of SPIE, 5542, 448-458(2004).

    [4] KANG G, COSTE P, YOUN H, et al. An in-orbit radiometric calibration method of the geostationary ocean color imager[J]. IEEE Transactions on Geoscience and Remote Sensing, 48, 4322-4328(2010).

    [5] BAUDIN G, MATTHEWS S, BESSUDO R, et al. Medium-resolution imaging spectrometer (MERIS) calibration sequence[J]. Proceedings of SPIE, 2819, 141-150(1996).

    [6] CAO CH Y, XIONG J, BLONSKI S, et al. Suomi NPP VIIRS sensor data record verification, validation, and long-term performance monitoring[J]. Journal of Geophysical Research, 118, 11664-11678(2013).

    [7] CZAPLA-MYERS J, ANDERSON N, THOME K, et al. The absolute radiometric calibration of the Landsat 8 Operational Land Imager using the reflectance-based approach and the Radiometric Calibration Test Site (RadCaTS)[J]. Proceedings of SPIE, 9218, 921819(2014).

    [8] HUANG L K, CEBULA R P, HILSENRATH E. New procedure for interpolating NIST FEL lamp irradiances[J]. Metrologia, 35, 381-386(1998).

    [9] SHAO X, CAO CH Y, LIU T C. Spectral dependent degradation of the solar diffuser on Suomi-NPP VIIRS due to surface roughness-induced Rayleigh scattering[J]. Remote Sensing, 8, 254(2016).

    [10] OJANEN M, KÄRHÄ P, IKONEN E. Spectral irradiance model for tungsten halogen lamps in 340-850 nm wavelength range[J]. Applied Optics, 49, 880-886(2010).

    [11] FU L, LEUTZ R, RIES H. Physical modeling of filament light sources[J]. Journal of Applied Physics, 100, 103528(2006).

    [12] [12] FU L. Increasing the brightness of light sources[D]. Marburg: PhilippsUniversitat Marburg, 2006.

    [13] LELAND J E, ARECCHI A V. Phase 2 analysis of Spectralon material for use in on-board calibration systems for the medium-resolution imaging spectrometer (MERIS)[J]. Proceedings of SPIE, 2475, 384-392(1995).

    [14] [14] BERK A, ERSON G P, AYA P K, et al. . Modtran4 User’s Manual[M]. Hanscom AFB: Air Fce Research Labaty, 1999.

    [15] [15] Technical Guide: reflectance materials coatings[R]. Labsphere, 1997.

    Tools

    Get Citation

    Copy Citation Text

    Ya-chao ZHANG, Xin YE, Zhi-wei XIA, Long SUI, Wei FANG. Spectral irradiance degradation model of halogen tungsten lamps at wavelength from 400 nm to 1300 nm[J]. Chinese Optics, 2022, 15(4): 825

    Download Citation

    EndNote(RIS)BibTexPlain Text
    Save article for my favorites
    Paper Information

    Category: Original Article

    Received: Nov. 30, 2021

    Accepted: Feb. 22, 2022

    Published Online: Sep. 6, 2022

    The Author Email: Xin YE (yexin@ciomp.ac.cn)

    DOI:10.37188/CO.EN.2021-0011

    Topics